The Keysight Agilent 8714ES is an RF S-parameter Vector Network Analyzer operating from 300 kHz to 3 GHz. It features a built-in S-parameter test set with full two-port vector-error correction, a synthesized source, narrowband and broadband receivers, and a standard 60-dB source attenuator, making it a high-performance, cost-effective solution for active and passive RF component characterization.

Specifications

General

ParameterValue
InterfacesGPIB, LAN, RS-232, Parallel, VGA-out, Keyboard (6-pin mini-DIN)
Display9-inch diagonal monochrome CRT

Physical

ParameterValue
Operating Temperature0 to 55 °C

Stimulus (Source)

ParameterValue
Frequency Range300 kHz to 3.0 GHz
Frequency Resolution1 Hz
Frequency Accuracy±5 ppm (at 25 °C ±5 °C)

Receiver

ParameterValue
Receiver Frequency Range300 kHz to 3.0 GHz
Input Impedance50 Ω (standard) / 75 Ω (Option 1EC)

Test Port Characteristics

ParameterValue
Number of Ports2
Connector TypeType-N female

Measurement Functions & Features

ParameterValue
Measurement Channels2 independent channels
Measurement ParametersS11, S21, S12, S22 (ES models)
Display FormatsLog magnitude, Phase, Group delay, Smith chart, Polar, Linear magnitude, SWR, Real, Imaginary
Calibration Types SupportedFull 2-port, 1-port, Response, Response/Isolation
Internal Storage3.5-inch floppy drive (LIF and MS-DOS formats)
Programming LanguageInstrument BASIC (IBASIC)
Limit Line TestingUpper, lower, flat, sloping, and single point limit lines

Connectivity & Interfaces

ParameterValue
GPIB InterfaceIEEE-488.2, SCPI compatible
LAN InterfaceRJ-45 (10Base-T)
Parallel Port25-pin D-sub, Centronics-compatible
Serial Port9-pin D-sub, RS-232
External VGA Port15-pin D-sub (VGA compatible)

Hardware Options & Accessories

ParameterValue
Option 1EC75 Ohm impedance version
Option 1D5High stability frequency reference
Option 100Fault location and structural return loss measurement

System Features

ParameterValue
Built-in Disk Drive3.5-inch floppy drive, MS-DOS and LIF formats, 1.44 MB
Real-Time ClockTime and date stamps for saved files

Options

ParameterValue
Option 1E1Adds a 60 dB step attenuator to extend power range down to -85 dBm (50 Ω) or -80 dBm (75 Ω)
Option 1C2Adds IBASIC (Instrument BASIC) capability
Option 1CMRack mount kit without handles
Option 1CPRack mount kit with handles

Storage

ParameterValue
Floppy Disk Format3.5-inch 1.44 MB (MS-DOS and LIF formats)

Display Capabilities

ParameterValue
Magnitude Display Resolution0.01 dB/division
Phase Display Resolution0.01°/division

Technology

Vector Network Analysis

Applications

RF component characterization, S-parameter testing, filter tuning, amplifier test, CATV tap testing