The Keysight (Agilent) 8990A Peak Power Analyzer is a high-performance instrument designed for complete and accurate characterization of complex pulsed RF and microwave signals. Operating with compatible peak power sensors, it offers dual-channel power measurements alongside dual-channel oscilloscope functionality, facilitating precise time-delay measurements between modulating video signals and RF envelopes.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | GPIB (HP-IB) |
| Display | 9-inch monochrome CRT |
| Warm-up Time | 30 minutes |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 115 / 230 VAC (±10%), 48 - 66 Hz |
| Operating Temperature | 0 to +55 °C |
| Dimensions (W×H×D) | 425 mm × 194 mm × 356 mm |
| Weight | 11 kg |
Frequency and Power Specifications
| Parameter | Value |
|---|---|
| Frequency Range | 50 MHz to 40 GHz (sensor dependent) |
| Power Measurement Range | -32 dBm to +20 dBm (sensor dependent) |
| Rise Time Capability | < 5 ns |
| Minimum Measurable Pulse Width | 10 ns |
| Power Resolution | 0.01 dB |
Dynamic and Timebase Specifications
| Parameter | Value |
|---|---|
| Timebase Range | 2 ns/div to 5 s/div |
Triggering Specifications
| Parameter | Value |
|---|---|
| Trigger Sources | Channel 1, Channel 2, Channel 3, Channel 4 |
| Trigger Modes | Auto, Triggered, Single |
Measurement and Analysis Capabilities
| Parameter | Value |
|---|---|
| Automatic Timing Measurements | 7 (Rise time, Fall time, Pulse width, PRI, PRF, Duty cycle, Delay) |
| Automatic Power Measurements | 5 (Peak power, Average power, Pulse top amplitude, Pulse base amplitude, Overshoot) |
| Marker Functions | Amplitude and time markers |
| Waveform Math Functions | Subtraction (Ch1 - Ch4, etc.), scalar pulse parameters |
Input and Channel Characteristics
| Parameter | Value |
|---|---|
| Number of RF Channels | 2 (Channels 1 and 4) |
| Number of Oscilloscope Channels | 2 (Channels 2 and 3) |
| Compatible Sensor Models | HP 84812A, HP 84813A, HP 84814A, HP 84815A |
| Oscilloscope Channel Input Impedance | 1 MΩ or 50 Ω (selectable) |
| Oscilloscope Channel Bandwidth | 100 MHz |
Internal Calibration Source
| Parameter | Value |
|---|---|
| Calibrator Output Frequency | 1.00 GHz |
| Calibrator Output Power Level | 10.0 mW (+10 dBm) |
| Calibrator Output Impedance | 50 Ω |
| Calibrator Output Connector | Type-N (female) |
System and Interface Connectivity
| Parameter | Value |
|---|---|
| GPIB Interface Functions | Standard HP-IB remote control and data transfer |
Power and Environmental Requirements
| Parameter | Value |
|---|---|
| Storage Temperature Range | -40 °C to +75 °C |
Oscilloscope Channels
| Parameter | Value |
|---|---|
| Oscilloscope Vertical Sensitivity | 2 mV/div to 5 V/div |
| Oscilloscope Vertical Resolution | 8 bits |
| Oscilloscope Input Coupling | ac, dc |
| Oscilloscope Maximum Input Voltage | 1 MΩ: 250 V (dc + peak ac); 50 Ω: 5 V rms |
Horizontal System
| Parameter | Value |
|---|---|
| Timebase Resolution | 100 ps |
Calibrator
| Parameter | Value |
|---|---|
| Calibrator Output SWR | 1.10 max |
| Calibrator Output Power Accuracy | ±1.5% (at 25 °C ± 10 °C), ±2.0% (0 °C to 55 °C) |
Environmental
| Parameter | Value |
|---|---|
| Operating Humidity | up to 95% RH at 40 °C |
| Maximum Operating Altitude | 4570 m (15000 ft) |
| Maximum Storage Altitude | 15240 m (50000 ft) |
| Vibration Resistance | Meets MIL-T-28800D Type III, Class 5 |
Performance
| Parameter | Value |
|---|---|
| Instrumentation Accuracy | ±0.04 dB (single channel), ±0.08 dB (dual-channel ratio) |
| Power Linearity | ±0.13 dB (sensor input power ≥ -20 dBm), ±0.20 dB (sensor input power < -20 dBm) |
| Timebase Accuracy | ±0.005% |
| Oscilloscope Channel Rise Time | < 3.5 ns |
| RF Channel Isolation | Performance — Oscilloscope Channel Isolation = > 50 dB (at equal scale factors) |
| Display Resolution | 0.01 dB (power measurements) |
Data Storage
| Parameter | Value |
|---|---|
| Setup Registers | 10 non-volatile setups |
| Waveform Memories | 4 non-volatile waveform memories |
Power Requirements
| Parameter | Value |
|---|---|
| Line Voltage | 90 to 132 VAC / 198 to 264 VAC (switchable) |
| Line Frequency | 48 to 66 Hz |
| Power Consumption | 250 VA maximum |
Technology
RF Peak Power Analysis
Applications
Pulsed RF and Microwave Signal Characterization