The Keysight DSO6104L (originally Agilent) is a 1 GHz, 4-channel digital storage oscilloscope designed for automated and manufacturing test systems. Packing standard 100% software-compatibility with portable InfiniiVision 6000A Series models, it is optimized for high-density applications, fitting into a space-efficient, low-profile 1U rackmount chassis.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | USB 2.0 (host & device), LAN (10/100Base-T), GPIB (IEEE-488), XGA video output |
| Display | None (requires external monitor via rear XGA out or remote web browser graphical interface) |
| Warm-Up Time | 30 minutes |
| Calibration Interval | 1 year |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 100 to 240 VAC, 50 to 60 Hz (automatic selection), 80 W maximum |
| Operating Temperature | 0 °C to 50 °C |
| Dimensions (W×H×D) | 437 mm × 43.6 mm × 272 mm |
| Weight | 4.5 kg |
Vertical System
| Parameter | Value |
|---|---|
| Analog Bandwidth | 1 GHz |
| Number of Analog Channels | 4 |
| Vertical Resolution | 8 bits |
| Input Impedance | 1 MΩ ±1% or 50 Ω ±1% (user selectable) |
| Input Coupling | AC, DC, GND |
| Maximum Input Voltage (1 MΩ) | 300 Vrms CAT I, 100 Vrms CAT II |
| Maximum Input Voltage (50 Ω) | 5 Vrms |
| Hardware Bandwidth Limit | 20 MHz (selectable) |
| Input Sensitivity Range (1 MΩ) | 2 mV/div to 5 V/div |
| Input Sensitivity Range (50 Ω) | 2 mV/div to 1 V/div |
| DC Gain Accuracy | ±2% of full scale |
| DC Offset Range (2 mV to 200 mV/div) | ±5 V on ranges < 10 mV/div; ±20 V on ranges 10 mV/div to 200 mV/div |
| DC Offset Range (201 mV to 5 V/div) | ±75 V on ranges > 200 mV/div to 5 V/div |
| Dynamic Range | ±8 divisions from screen center |
| Channel-to-Channel Isolation | ≥ 40 dB from DC to 1 GHz |
| Input Capacitance (1 MΩ) | 13 pF |
Horizontal System
| Parameter | Value |
|---|---|
| Real-Time Sample Rate (Half Channel) | 4 GSa/s |
| Real-Time Sample Rate (All Channels) | 2 GSa/s |
| Standard Memory Depth (Half Channel) | 8 Mpts |
| Standard Memory Depth (All Channels) | 4 Mpts |
| Time Base Range | 500 ps/div to 50 s/div |
| Time Base Accuracy | ±15 ppm |
| Time Base Delay Range (Pre-trigger) | Greater of 1 screen width or 250 µs |
| Time Base Delay Range (Post-trigger) | 1 s to 500 s |
| Channel-to-Channel Deskew Range | ±100 ns |
Trigger System
| Parameter | Value |
|---|---|
| Trigger Modes | Auto, Normal, Single |
| Trigger Sources | Channel 1, Channel 2, Channel 3, Channel 4, External, Line |
| Trigger Coupling | AC, DC, LF Reject, HF Reject, Noise Reject |
| Trigger Holdoff Range | 60 ns to 10 s |
| Trigger Sensitivity (Internal, <10 MHz) | 0.3 div or 3.5 mV |
| Trigger Sensitivity (Internal, 10 MHz to 1 GHz) | 1 div or 10 mV |
| Trigger Sensitivity (External, DC to 100 MHz) | 100 mV |
| Trigger Sensitivity (External, 100 MHz to 1 GHz) | 200 mV |
| Edge Triggering | Rising, Falling, Alternating |
| Pulse Width Triggering | Trigger on pulse width <, >, within, or outside 2 ns to 10 s |
| Pattern Triggering | Trigger on logical AND, OR, NAND, NOR of channel states |
| TV/Video Triggering | NTSC, PAL, SECAM, HDTV (480i, 480p, 720p, 1080i, 1080p) |
| Sequence Triggering | Arm on A, Trigger on B with delay or event count |
Acquisition System
| Parameter | Value |
|---|---|
| Acquisition Modes | Normal, Peak Detect, Averaging, High Resolution |
| Peak Detect Performance | 250 ps (1 GHz model) |
| Averaging Factors | Selectable from 2 to 16384 |
| High-Resolution Mode Resolution | Up to 12 bits of resolution when bandwidth is limited |
| Waveform Update Rate | Up to 100,000 waveforms/sec |
Waveform Measurements and Analysis
| Parameter | Value |
|---|---|
| Automatic Voltage Measurements | Vp-p, Vmax, Vmin, Vtop, Vbase, Vamp, Vavg, Vrms, Overshoot, Preshoot |
| Automatic Time Measurements | Period, Frequency, Rise Time, Fall Time, +Width, -Width, Duty Cycle |
| Waveform Math Operators | Addition, Subtraction, Multiplication, FFT |
| FFT Window Functions | Hanning, Flat Top, Rectangular |
| Serial Bus Triggering Options | I2C, SPI, CAN, LIN standard / option dependent |
System Connectivity & I/O Interfaces
| Parameter | Value |
|---|---|
| Software Compatibility | 100% compatible with 6000A Series portable oscilloscopes |
| LXI Compliance Class | LXI Class C |
| LAN Port Speed | 10/100Base-T |
| GPIB Interface | Built-in (standard IEEE-488) |
| USB Host Ports | 1x Front, 1x Rear (USB 2.0 High Speed) |
| USB Device Port | 1x Rear (USB 2.0 High Speed, USBTMC compliant) |
| External Trigger Input | BNC, front panel |
| External Trigger Output | BNC, rear panel |
| Reference Input | 10 MHz BNC, rear panel |
| Reference Output | 10 MHz BNC, rear panel |
| Video Display Output Port | XGA (1024 x 768) 15-pin D-sub, rear panel |
| Web Browser Remote Control Support | Built-in graphical Web server (Java-based) |
Physical & Rackmount Characteristics
| Parameter | Value |
|---|---|
| Rack Unit (U) Height | 1U |
| Rackmount Kit | Standard built-in 19-inch rackmount ears |
Power Specifications
| Parameter | Value |
|---|---|
| Line Voltage Range | 100 to 240 VAC |
| Line Frequency Range | 50 to 60 Hz |
| Power Consumption (Max) | 80 W |
Environmental and Compliance
| Parameter | Value |
|---|---|
| Storage Temperature Range | -40 °C to +70 °C |
| Operating Humidity Range | Up to 95% RH at 40 °C (non-condensing) |
| Operating Altitude | Up to 4,600 m (15,000 ft) |
| Safety Certifications | UL61010-1, CAN/CSA-C22.2 No. 61010-1, EN61010-1 |
| Electromagnetic Compatibility (EMC) | IEC 61326-1 / EN 61326-1 |
Measurements
| Parameter | Value |
|---|---|
| Hardware Frequency Counter Resolution | 5 digits |
| Hardware Frequency Counter Max Frequency | 1 GHz |
| Cursor Measurements | Time, Amplitude, dX, dY, 1/dX |
Display & Video
| Parameter | Value |
|---|---|
| External Monitor Resolution | 1024 x 768 (XGA) |
Storage
| Parameter | Value |
|---|---|
| Internal Storage Capacity | 10 setups and 10 waveforms |
Environmental
| Parameter | Value |
|---|---|
| Environmental Vibration Resistance | MIL-PRF-28800F, Class 3 (random) |
Performance
| Parameter | Value |
|---|---|
| XY Mode Phase Error | ±1.8° at 1 MHz |
| Segmented Memory Max Segments | 1000 |
Connectivity & Software
| Parameter | Value |
|---|---|
| Waveform Storage Formats | BMP, PNG, CSV, ASCII XY, BIN |
Environmental & Compliance
| Parameter | Value |
|---|---|
| Operating Vibration | 0.21 g rms, 5 Hz to 500 Hz |
| Non-operating Vibration | 2.09 g rms, 5 Hz to 500 Hz |
| Pollution Degree | 2 |
| Measurement Category | CAT I |
Technology
Digital Storage Oscilloscope (DSO)
Applications
Automated Test Systems, Manufacturing Test, Design Verification