The Agilent InfiniiVision DSOX3034A is a 4-channel, 350 MHz digital storage oscilloscope designed to offer high performance, flexibility, and extensive upgradability. Featuring an 8.5-inch WVGA display and a fast waveform update rate of up to 1,000,000 waveforms per second, this model is built for efficient serial measurement and signal analysis.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | Two USB 2.0 high-speed host ports (one front, one rear), one USB 2.0 high-speed device port (rear); optional 10/100Base-T LAN and Video out (via DSOXLAN module); optional GPIB (via DSOXGPIB module) |
| Display | 8.5-inch WVGA with 64 levels of intensity grading |
| Acoustic Noise | 30 dBA sound pressure level (at operator position) |
| Recommended Calibration Cycle | 2 years |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 100-120 V (50/60/400 Hz), 100-240 V (50/60 Hz), 100 W max |
| Operating Temperature | 0 to +55 °C |
| Dimensions (W×H×D) | 380.6 mm × 204.4 mm × 141.5 mm |
| Weight | 3.85 kg |
Vertical System (Analog Channels)
| Parameter | Value |
|---|---|
| Analog Bandwidth (-3 dB) | 350 MHz |
| Number of Input Channels | 4 |
| Calculated Rise Time (10% to 90%) | ≤ 1 ns |
| Input Coupling | AC, DC |
| Input Impedance | Selectable 1 MΩ ± 1% or 50 Ω ± 1.5% |
| Vertical Sensitivity Range | 1 mV/div to 5 V/div (1 MΩ), 1 mV/div to 1 V/div (50 Ω) |
| Vertical Resolution (A/D Converter) | 8 bits |
| Channel-to-Channel Isolation | > 100:1 from DC to 350 MHz (measured with same V/div and coupling on channels) |
| Offset Range | ±2 V (1 mV/div to 200 mV/div), ±50 V (> 200 mV/div to 5 V/div) |
| Hardware Bandwidth Limiters | 20 MHz selectable |
Horizontal System & Acquisition
| Parameter | Value |
|---|---|
| Maximum Real-time Sample Rate (Single-Channel) | 4 GSa/s (half-channel interleaved) |
| Maximum Real-time Sample Rate (Multi-Channel) | 2 GSa/s per channel |
| Maximum Record Length (Memory Depth) | Standard: 1 Mpts per channel (2 Mpts half-channel interleaved); Optional: 2 Mpts per channel (4 Mpts half-channel interleaved) |
| Time Base Range | 1 ns/div to 50 s/div |
| Channel-to-Channel Deskew Range | ± 100 ns |
Trigger System
| Parameter | Value |
|---|---|
| Trigger Modes | Normal (triggered), Auto, Single, Force |
| Internal Trigger Level Range | ± 6 div from center screen |
| External Trigger Level Range | N/A (external trigger input is only available on 2-channel models) |
| Trigger Coupling | AC, DC, LF reject, HF reject, noise reject |
| Trigger Holdoff Range | 40 ns to 10 s |
Acquisition Modes & Signal Processing
| Parameter | Value |
|---|---|
| Acquisition Modes | Normal, Peak detect, Averaging, High resolution, Segmented (optional) |
| Peak Detect Glitch Capture | Narrow as 250 ps at all timebase settings |
| High Resolution Mode | 12 bits of resolution when ≥ 10 μs/div at 4 GSa/s or ≥ 20 μs/div at 2 GSa/s |
| Segmented Memory Re-arm Time | 1 μs (minimum time between trigger events) |
| Waveform Update Rate | 1,000,000 waveforms/s |
| Waveform Math Operators | 7 functions, including FFT |
Measurements & Analysis
| Parameter | Value |
|---|---|
| Total Automatic Measurements | 33 automated measurements |
| Voltage Measurements | Peak-to-peak, maximum, minimum, amplitude, top, base, overshoot, pre-shoot, average- N cycles, average- full screen, DC RMS- N cycles, DC RMS- full screen, AC RMS- N cycles, AC RMS- full screen (standard deviation), ratio (RMS1/RMS2) |
| Time Measurements | Period, frequency, counter, + width, - width, burst width, duty cycle, rise time, fall time, delay, phase, X at min Y, X at Max Y |
| Count Measurements | Positive pulse count, negative pulse count, rising edge count, falling edge count |
| Mixed Measurements | Area- N cycles, area- full screen |
| Measurement Statistics | Continuously updated with statistics; cursors track last selected measurement (up to four active measurements) |
| Hardware-based Mask Testing | Up to 280,000 tests per second |
| Reference Waveforms | Store up to 2 waveforms in non-volatile reference memory |
Display & User Interface
| Parameter | Value |
|---|---|
| Display Type and Size | 8.5-inch WVGA |
| Display Intensity Levels | 64 levels of intensity grading |
| Autoscale | Autoscale button for automatic setting of vertical, horizontal, and trigger controls (can be disabled or enabled) |
| Localized Interface | Localized GUI and built-in help |
Input/Output Interfaces
| Parameter | Value |
|---|---|
| USB Host Ports | Two USB 2.0 hi-speed host ports (one front, one rear); supports memory devices and printers |
| USB Device Port | One USB 2.0 hi-speed device port on rear panel |
| LAN Port | 10/100Base-T (requires optional DSOXLAN module) |
| Video Out Port | Allows external monitor or projector connection (requires optional DSOXLAN module) |
| GPIB Port | Migration capability for legacy test systems (requires optional DSOXGPIB module) |
| Probe Compensator Output | Square wave: 2.5 Vpp, 1 kHz |
| WaveGen Output | Front-panel BNC Connector (requires WaveGen option) |
| Security Lock Slot | Rear-panel security slot compatible with Kensington-style locks |
Accessories & Ordering Options
| Parameter | Value |
|---|---|
| Logic Timing Analyzer Upgrade | Upgradable to MSO functionality with 16 digital channels via DSOX3MSO option |
| Function Generator Option | Built-in 20-MHz function generator WaveGen |
| Optional Serial Protocol Analysis | I²C, SPI, RS232, UART, CAN, LIN, and I²S |
| Post-Purchase Upgradability | Upgradable bandwidth, digital channels, or WaveGen function generator after purchase |
Vertical System
| Parameter | Value |
|---|---|
| DC Vertical Gain Accuracy | ±2.0% full scale |
| DC Vertical Offset Accuracy | ±0.1 div ± 2 mV ± 1% of offset setting |
| Maximum Input Voltage (1 MΩ) | 300 Vrms, 400 Vpk (CAT I and CAT II) |
| Maximum Input Voltage (50 Ω) | 5 Vrms (CAT I) |
| Dynamic Range | ±8 divisions from screen center |
Horizontal System
| Parameter | Value |
|---|---|
| Time Base Accuracy | ±25 ppm + 2 ppm per year aging |
Environmental
| Parameter | Value |
|---|---|
| Non-operating Temperature | -40 to +70 °C |
| Operating Humidity | Up to 90% RH at 40 °C (non-condensing) |
| Non-operating Humidity | Up to 90% RH at 65 °C (non-condensing) |
| Operating Altitude | Up to 3,000 m (10,000 ft) |
| Non-operating Altitude | Up to 15,240 m (50,000 ft) |
Waveform Math & Analysis
| Parameter | Value |
|---|---|
| FFT Window Functions | Hanning, Flat Top, Rectangular, Blackman-Harris |
Built-in Instruments
| Parameter | Value |
|---|---|
| Integrated Digital Voltmeter | 3-digit (characterizes AC RMS, DC, and DC RMS) |
| Integrated Frequency Counter | 5-digit (standard) |
Standards & Safety
| Parameter | Value |
|---|---|
| Overvoltage Category | CAT I 300 Vrms, CAT II 300 Vrms (1 M̤) |
| Electromagnetic Compatibility (EMC) | Meets IEC 61326-1:2005 / EN 61326-1:2006 Group 1 Class A |
Technology
Digital Storage Oscilloscope (DSO)
Applications
Electronic hardware troubleshooting, serial protocol analysis, logic timing measurement, education, and R&D testing