The Keysight (Agilent/HP) E1437A is a high-performance, single-slot, C-size VXI modular digitizer. It features 23 bits of resolution at a 20 MSa/s sampling rate, designed for advanced digital signal processing (DSP) applications such as spectral analysis, filtering, and demodulation. It includes integrated analog conditioning, an 8 MHz analog anti-alias filter, a digital local oscillator (LO), and 24 decimation filters with a large onboard FIFO memory.

Specifications

General

ParameterValue
InterfacesVXIbus, VXI Local Bus, Front Panel SMB (Input, External Trigger, External Clock)
Warm-up Time30 minutes

Physical

ParameterValue
Power SupplyVXI Backplane powered (+5 V, +12 V, -12 V, -5.2 V, -2 V)
Operating Temperature0 to 55 °C
Dimensions (W×H×D)262 mm × 30.5 mm × 353 mm (VXI C-size, 1-slot)

Analog Input

ParameterValue
Number of Input Channels1
Input Impedance50 Ω
Analog Bandwidth8 MHz
Input Ranges-30 dBm to +26 dBm (in 2 dB steps)
Input CouplingDC

Analog-to-Digital Conversion (ADC)

ParameterValue
ADC Resolution14 bits
Maximum Sampling Rate20 MSa/s
Alias-Free Bandwidth8 MHz (at 20 MSa/s)

Digital Signal Processing (DSP)

ParameterValue
Digital Local Oscillator (LO) Range0 Hz to 10 MHz
Digital LO Resolution32-bit
Number of Decimation Filters24
Decimation Range1 to 8,388,608 (2^0 to 2^23)

Onboard Memory

ParameterValue
Onboard Memory TypeFIFO
FIFO Memory Capacity2 MSamples

Triggering & Synchronization

ParameterValue
Trigger SourcesImmediate, External, TTLTRG (VXI), Software, Local Bus
Trigger ModesFree run, Level, External
External Trigger InputFront panel SMB

Clocking & Timebase

ParameterValue
External Clock InputFront panel SMB
External Clock Frequency20 MHz (nominal)

Bus & System Interface

ParameterValue
Module Form FactorVXI C-size
Number of Slots Occupied1 slot
VXI Device TypeRegister-based slave
VXI Local Bus SupportYes (Left & Right local bus)

Software & Compliance

ParameterValue
Driver Software SupportVXIplug&play
Application Software CompatibilityVisual Basic, Agilent VEE, LabVIEW, C/C++
I/O Library CompatibilityAgilent IO Libraries / Keysight Connection Expert

Dynamic Performance

ParameterValue
Spurious-Free Dynamic Range (SFDR)> 90 dBfs (dc to 8 MHz)
Signal-to-Noise Ratio (SNR)> 74 dB (dc to 8 MHz)
Noise Spectral Density<-142 dBfs/Hz (dc to 8 MHz)
Amplitude Flatness±0.1 dB (dc to 8 MHz)
Phase Linearity±2.0 degrees (dc to 8 MHz)

Filter Performance

ParameterValue
Passband Ripple< ±0.04 dB
Stopband Attenuation> 110 dB

Input Specifications

ParameterValue
Maximum Overvoltage Protection±15 V peak

Environmental

ParameterValue
Storage Temperature Range-40 °C to +75 °C
Operating HumidityUp to 95% RH (non-condensing)
Operating AltitudeUp to 4,570 m (15,000 ft)
Cooling Requirements4.5 L/s @ 4.4 Pa (0.45 mm H2O) pressure drop

Inputs & Outputs

ParameterValue
Input ConnectorSingle BNC (female)
External Clock, Trigger, and Sync ConnectorsSMB (male)

VXI Interface

ParameterValue
VXI Device ClassRegister-based (A16/A32, D16/D32/D64)
VXIbus Specification ComplianceRevision 1.4

Technology

VXI Digitizer

Applications

High-performance signal analysis, digital signal processing (DSP), spectral analysis, and communications testing