The Keysight Technologies E4402B ESA-E Series Spectrum Analyzer offers a frequency range of 100 Hz to 3.0 GHz, making it an ideal mid-performance instrument for general-purpose RF and microwave measurements. It features 0.4 dB overall amplitude accuracy, a +16 dBm Third Order Intercept (TOI), and a low -166 dBm Displayed Average Noise Level (DANL) with an internal preamplifier.

Specifications

General

ParameterValue
InterfacesGPIB, LAN (10Base-T), RS-232, Parallel, VGA
Display6.5-inch Color TFT LCD
Recommended Calibration Cycle1 year

Physical

ParameterValue
Power Supply90-132 VAC / 198-264 VAC, 47-66 Hz, 250 VA max
Operating Temperature0 to 55 °C
Dimensions (W×H×D)350 mm (W) × 222 mm (H) × 410 mm (D)
Weight15.5 kg
Rack Mount KitAvailable (Agilent P/N E4400-60002)

Frequency Characteristics

ParameterValue
Frequency Range100 Hz to 3.0 GHz
Frequency Reference Aging Rate±1 x 10^-7 / year
Frequency Readout Accuracy±( (frequency x reference uncertainty) + 1% x span + 10% x RBW + 10 Hz)
Span Accuracy10 Hz to 5 MHz (10% steps), 9 kHz, 120 kHz (standard)
Resolution Bandwidth (RBW) Range1 Hz to 5 MHz (1-3-10 sequence), 9 kHz, 120 kHz, 1 Hz (optional)
RBW Accuracy±10%
Video Bandwidth (VBW) Range1 Hz to 5 MHz (1-3-10 sequence)
Sweep Time Range1 ms to 1000 s
Phase Noise @ 10 kHz Offset-98 dBc/Hz (1 GHz carrier)
Minimum Settable Frequency0 Hz (Zero Span)
Maximum Settable Frequency3.0 GHz
Frequency Counter Resolution1 Hz, 10 Hz, 100 Hz, 1 kHz selectable

Amplitude Characteristics

ParameterValue
Maximum Input Level (Average)+30 dBm
Maximum Input Level (Peak, with attenuator)+50 dBm
Maximum DC Input0 VDC
Input Attenuator Range0 to 70 dB (10 dB steps)
Input Attenuator Accuracy±0.5 dB
Displayed Average Noise Level (DANL)-155 dBm/Hz (with internal preamp, Option 1DP, 10 MHz to 2.5 GHz)
Third Order Intercept (TOI)+16 dBm
Second Harmonic Distortion (SHD)-90 dBc
Input VSWR<1.5:1 (10 dB attenuation, preamplifier off)
Absolute Amplitude Accuracy±0.4 dB
Frequency Response Flatness±0.5 dB
Amplitude Measurement UnitsdBm, dBmV, dBµV, V, W
Log Scale Fidelity±0.2 dB (0 to -70 dB from reference level)

Sweep and Measurement

ParameterValue
Number of Display Points2 to 8192 (variable)
Analysis Bandwidth10 MHz (with digitizer options)
Trace DetectorsPeak, Sample, Normal, Negative Peak, Average (RMS, Power Avg, Voltage Avg); Quasi-Peak (Optional)
Trace FunctionsClear/Write, Max Hold, Min Hold, View, Blank, Trace Math
Measurement FunctionsChannel Power, Occupied Bandwidth (OBW), Adjacent Channel Power Ratio (ACPR), xdB Bandwidth, Harmonic Distortion, Spurious Emissions
Triggering ModesFree Run, Video, External, Line
Segmented SweepUp to 32 discontinuous spans in one sweep

Input/Output & Interfaces

ParameterValue
RF Input Connector TypeN-type (female)
RF Input Impedance50 Ohm
External Reference Input10 MHz, BNC (female)
External Reference Output10 MHz, BNC (female)
VGA OutputDB-15 connector
External Trigger InputBNC (female)
Remote Control CapabilityGPIB, LAN, RS-232

Display

ParameterValue
Display Resolution640 x 480 pixels
Number of TracesUp to 4
Display ScalesLinear, Log (10, 5, 2, 1 dB/div)

General Specifications

ParameterValue
Storage Temperature Range-40 to +75 °C
Humidity Range95% RH at 40 °C (non-condensing)
Warm-up Time5 minutes (to guaranteed measurement accuracy)

Options and Accessories

ParameterValue
Tracking Generator OptionOption 1D5
Preamplifier OptionOption 1DP (internal preamplifier)
EMI Measurement CapabilityOption 1DS (includes 9 kHz, 120 kHz RBW and quasi-peak detector)
Digital Demodulation OptionOption AYA
Narrow Resolution Bandwidth OptionOption 1DR (1 Hz to 1 kHz)

Interfaces

ParameterValue
GPIB InterfaceIEEE-488.2
LAN Interface10/100Base-T Ethernet (Option 1FP)
USB Ports2 (rear panel)
Parallel PortDB-25 (printer port)
RS-232 Serial PortDB-9
Composite Video OutputBNC
IF Output21.4 MHz (Option AYZ)

Resolution Bandwidth

ParameterValue
RBW Filter Shape Factor (60 dB/3 dB)<5:1 (standard), <15:1 (Option 1D6)

Frequency Reference

ParameterValue
External Reference Output Level10 MHz, +10 dBm nominal into 50 Ohm
External Reference Input Level10 MHz, -10 dBm to +10 dBm into 50 Ohm

RF Input Characteristics

ParameterValue
Input Return Loss (RF Input)>15 dB (preamp off, 10 MHz to 3 GHz)

Spectral Purity

ParameterValue
Residual Response (non-harmonic spurious)<-90 dBm (with input terminated, 0 dB attenuation, preamplifier off, >10 MHz)

Power Requirements

ParameterValue
Power Consumption250 VA max (150 VA typical)

Environmental

ParameterValue
Altitude (Operating)Up to 4,600 m (15,000 ft)
Altitude (Non-operating)Up to 15,300 m (50,000 ft)

Compliance

ParameterValue
Safety ComplianceCSA, UL, IEC, EN
EMC ComplianceIEC/EN 61326-1, CISPR 11 (Group 1, Class B)

Amplitude

ParameterValue
Amplitude Display Range100 dB (log scale)

Connectivity

ParameterValue
External Reference Input Impedance50 Ohm
External Reference Output Impedance50 Ohm
Keyboard/Mouse SupportUSB

Bandwidth

ParameterValue
VBW Filter TypeDigital

Audio

ParameterValue
Internal SpeakerYes

Power

ParameterValue
AC Input Voltage Range90 - 264 VAC
AC Input Frequency Range47 - 63 Hz

Technology

Spectrum Analyzer

Applications

RF and Microwave Measurement