The Keysight (formerly Agilent / HP) E4418B is a high-performance single-channel RF and microwave power meter from the EPM Series. Designed for benchtop and automatic test equipment (ATE) applications, it provides fast, accurate, and repeatable power measurements. When paired with E-series sensors, it offers high-speed measurements up to 200 readings per second and a dynamic range of up to 90 dB without range-switching delays. It features full backward compatibility with older HP 437B and 436A command sets and supports both Keysight 8480 and E-series power sensors.

Specifications

General

ParameterValue
DisplayLCD with LED backlight

Physical

ParameterValue
Power Supply100 to 240 VAC at 50/60 Hz; 100 to 120 VAC at 400 Hz; 50 VA maximum
Operating Temperature0 to 55 °C
Dimensions (W×H×D)212.6 × 88.5 × 348.3 mm
Weight4.0 kg

Measurement Specifications

ParameterValue
Frequency Range9 kHz to 110 GHz (sensor dependent)
Power Range-70 dBm to +44 dBm (100 pW to 25 W) (sensor dependent)
Single Sensor Dynamic RangeUp to 90 dB (E-series sensors), up to 50 dB (8480 series sensors)
Measurement SpeedUp to 200 readings per second (over GPIB with E-series sensors)
Resolution (Logarithmic)1, 0.1, 0.01, or 0.001 dB
Absolute Accuracy (Logarithmic)±0.02 dB (meter only)
Absolute Accuracy (Linear)±0.5% (meter only)

Display and Front Panel

ParameterValue
Measurement Channels1

Power Reference Subsystem

ParameterValue
Reference Oscillator Frequency50 MHz
Reference Oscillator Power Output1.0 mW (0 dBm)
Reference Oscillator Connector TypeType-N (female), 50 Ω

Sensor Compatibility

ParameterValue
Supported Sensor FamiliesKeysight 8480 series and E-series sensors

Connectivity and Interfaces

ParameterValue
Programming Command SetSCPI, HP 437B, and HP 436A compatibility

Power Requirements

ParameterValue
Battery Operating TimeUp to 5.5 hours continuous use (with Option 001)
Battery TypeInternal rechargeable battery (Option 001)

Physical Specifications

ParameterValue
Rackmount Form FactorHalf-rack width, 2U height

Performance

ParameterValue
Relative Accuracy (Logarithmic)±0.04 dB
Relative Accuracy (Linear)±1.0%
Resolution (Linear)1.0, 0.1, 0.01, and 0.001% of full scale
Reference Oscillator Power Accuracy±1.2% (worst case, 1 year)
Reference Oscillator VSWR≤ 1.06
Warm-up Time30 minutes

Inputs & Outputs

ParameterValue
Analog Output0 to 1.0 V (proportional to power, BNC connector)
Analog Output Voltage Range0 to 1.0 V (corresponding to full scale)
Analog Output Impedance1 kΩ (nominal)
Remote I/O (TTL)TTL output for pass/fail limits; TTL input for remote zero/calibration

Power

ParameterValue
Power Consumption50 VA maximum
Line Frequency50 / 60 / 400 Hz
Battery Recharge Time< 2 hours (instrument off) / < 8 hours (instrument on) (Option 001 dependent)

Environmental

ParameterValue
Storage Temperature-20 °C to +70 °C
Operating HumidityUp to 95% relative humidity at 40 °C (non-condensing)
Operating AltitudeUp to 3,000 m (9,840 ft)
Non-Operating AltitudeUp to 15,240 m (50,000 ft)

Measurement

ParameterValue
Display UnitsAbsolute: Watts or dBm; Relative: Percent or dB
Display Offset Range±99.99 dB
Averaging (Filter)1 to 1024 (auto or manual)

Compliance

ParameterValue
EMC ComplianceCISPR 11 / EN 55011 Group 1, Class A; EN 50082-1
Safety ComplianceEN 61010-1, IEC 1010-1, CSA C22.2 No. 1010.1

Options & Configuration

ParameterValue
Rear Panel Sensor InputAvailable via Option 002 or Option 003

Technology

RF/Microwave Power Measurement

Applications

Benchtop & ATE Power Measurements