The Keysight (formerly Agilent) E4419A is a high-performance, dual-channel, programmable power meter designed for benchtop and automatic test equipment (ATE) environments. Part of the EPM Series, it delivers fast, accurate, and repeatable RF power measurements. With dual-channel capability, it provides two simultaneous power readings on its high-resolution backlit LCD, supporting compatible E-series and 8480-series power sensors.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | GPIB (IEEE-488), RS-232, RS-422 |
| Display | High-resolution Liquid Crystal Display (LCD) with LED backlighting |
| Chassis Ground Terminal | Rear panel binding post |
| Warm-Up Time | 30 minutes |
| Calibration Factor Range | 1.0% to 150.0% |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 85 to 264 VAC, 50 to 440 Hz, automatic selection, < 70 VA (70 VA max, 25 W typical) |
| Operating Temperature | 0 to 55 |
| Dimensions (W×H×D) | 212.6 mm × 88.5 mm × 348.3 mm |
| Weight | 4.1 |
| Rackmount Configuration Options | Single or dual rackmount kits available |
Measurement
| Parameter | Value |
|---|---|
| Number of Measurement Channels | 2 |
| Frequency Range | 100 kHz to 110 GHz (sensor dependent) |
| Power Range | -70 to +44 dBm (sensor dependent) |
| Display Resolution | Selectable 1.0, 0.1, 0.01, and 0.001 dB (logarithmic), or 1 to 4 significant digits (linear) |
| Instrumentation Accuracy | ±0.5% (single channel) or ±1.0% (difference/ratio) |
| Single-Sensor Dynamic Range | 90 dB max (E-series sensors), 50 dB max (8480 series sensors) |
| Measurement Speed (Normal Mode) | 20 readings per second (with E-series sensors) |
| Measurement Speed (Fast Mode) | 100 readings per second (with E-series sensors over GPIB) |
Reference Calibrator
| Parameter | Value |
|---|---|
| Calibrator Power Output | 1.00 mW (0.00 dBm) |
| Calibrator Frequency | 50 MHz |
| Calibrator Power Accuracy (at 25 ± 10 °C) | ±0.9% |
| Calibrator Power Accuracy (0 to 55 °C) | ±1.2% |
| Calibrator VSWR | 1.06 max |
| Calibrator Connector Type | Precision Type-N (female) |
Sensors & Calibration
| Parameter | Value |
|---|---|
| Compatible Sensor Families | Agilent/HP 8480 Series and E-Series (except E9320A family) |
| Calibration Factor Entry Method | Keyboard entry, download via GPIB, or automated read from E-series EEPROM |
| Sensor Calibration Table Storage | Up to 20 tables of 80 frequency points each (for 8480 series) |
| Offset Range | ±100 dB in 0.001 dB increments |
Triggering & Synchronization
| Parameter | Value |
|---|---|
| Trigger Source | Internal, External, GPIB/Software |
| Trigger Delay Range | 0 to 1,000 ms |
| Trigger Delay Resolution | 1 ms |
| External Trigger Input Level | TTL compatible (active low) |
| Recorder Output Voltage Range | 0 to 1 V |
| Recorder Output Impedance | 1 kΩ |
User Interface
| Parameter | Value |
|---|---|
| Display Units | Absolute: Watts, dBm; Relative: %, dB |
| Programming Language | SCPI (Standard Commands for Programmable Instruments) |
| Code Emulation Mode | HP 438A |
| Save/Recall Registers | 10 instrument states |
Connectivity
| Parameter | Value |
|---|---|
| Sensor Input Connectors Location | Front panel (Option 002: rear panel) |
| External Trigger Input Connector | BNC (rear panel) |
| Recorder Output Connector | BNC (rear panel) |
| Calibrator Output Connector Location | Front panel (Option 002: rear panel) |
| GPIB Conformance | IEEE 488.2 |
Power Requirements
| Parameter | Value |
|---|---|
| Battery Option Availability | Option 001 (rechargeable NiMH battery, up to 3 hours of continuous operation) |
| Mains Voltage Range | 85 to 264 VAC (automatic selection) |
Environmental
| Parameter | Value |
|---|---|
| Storage Temperature Range | -20 °C to +70 °C |
| Operating Humidity | 95% at 40 °C (non-condensing) |
| Operating Altitude | Up to 4,600 meters (15,000 feet) |
| Storage Altitude | Up to 15,240 meters (50,000 feet) |
| Safety Standards Compliance | Conforms to CSA C22.2 No. 1010.1, EN 61010-1 / IEC 1010-1 |
| Electromagnetic Compatibility (EMC) | Conforms to EN 55011 / CISPR 11 Group 1 Class A |
Power & Electrical
| Parameter | Value |
|---|---|
| Power Consumption | 70 VA maximum (40 W) |
| Line Frequency Range | 50 to 440 Hz (100 to 120 VAC), 50 to 60 Hz (220 to 240 VAC) |
System Capabilities
| Parameter | Value |
|---|---|
| Maximum Power Sensor Cable Length | 121 m (400 ft) |
Performance
| Parameter | Value |
|---|---|
| Channel-to-Channel Isolation | >90 dB (at 1 GHz), >60 dB (at 40 GHz) |
Calibrator
| Parameter | Value |
|---|---|
| Calibrator Output Impedance | 50 Ω (nominal) |
Display
| Parameter | Value |
|---|---|
| Display Type | Liquid Crystal Display (LCD) |
| Display Backlight | LED |
Battery Option (Option 001)
| Parameter | Value |
|---|---|
| Battery Type | Nickel Metal Hydride (NiMH) (Option 001) |
| Battery Operating Time | Up to 3 hours (backlight on), up to 4 hours (backlight off) (Option 001) |
| Battery Charge Time | 2 hours to 90% capacity, 3 hours to 100% capacity (Option 001) |
Interfaces
| Parameter | Value |
|---|---|
| Serial Interface Connector | 6-pin RJ-11 |
| GPIB Subset Compatibility | SH1, AH1, T6, TE0, L4, LE0, SR1, RL1, PP1, DC1, DT1, C0 |
Technology
RF / Microwave Power Measurement