The Keysight (formerly Agilent / Hewlett-Packard) E4419B is a high-performance true dual-channel RF and microwave power meter belonging to the EPM Series. Designed for both benchtop and automatic test equipment (ATE) applications, it provides fast, accurate, and simultaneous dual-channel power measurements when paired with compatible 8480 or E-series power sensors. The meter achieves high measurement speeds of up to 100 readings per second over GPIB and features built-in backward code compatibility with the legacy HP 438A.

Specifications

General

ParameterValue
DisplayLCD with LED backlight
Save/Recall Registers10 instrument states

Physical

ParameterValue
Power Supply100 to 240 VAC, 50/60/400 Hz
Operating Temperature0 to 55
Dimensions (W×H×D)212.6 × 88.5 × 348.3
Weight4.1

Basic Characteristics

ParameterValue
Number of Channels2 (Simultaneous)
Frequency Range9 kHz to 110 GHz (sensor dependent)
Power Range–70 dBm to +44 dBm / 100 pW to 25 W (sensor dependent)
Dynamic Range90 dB maximum (E-series sensors), 50 dB maximum (8480 series sensors)
Display UnitsdBm, dB, Watts, %
Display ResolutionSelectable 1.0, 0.1, 0.01, 0.001 dB (logarithmic) or 1 to 4 digits (linear)

Measurement Accuracy & Performance

ParameterValue
Absolute Accuracy±0.02 dB (logarithmic) or ±0.5% (linear)
Measurement SpeedUp to 200 readings per second (over GPIB, with E-series sensors)

Power Reference Oscillator

ParameterValue
Reference Frequency50 MHz
Reference Power Output1.00 mW (0.0 dBm)
Reference Connector TypeType N (female), 50 Ω

Sensor Compatibility & Interfaces

ParameterValue
Supported Sensor FamiliesKeysight/Agilent/HP 8480 series and E-series

Connectivity & Remote Control

ParameterValue
Remote Command SetSCPI
Legacy Emulation ModesHP 438A code compatibility

Available Options & Configurations

ParameterValue
Internal Battery OptionRechargeable battery (Option 001)
Battery Operating TimeUp to 3 hours continuous use

Environmental & Regulatory

ParameterValue
Warm-up Time30 minutes

Analog & Recorder Outputs

ParameterValue
Analog Recorder Output0 to 1 V (one analog output per channel)

Performance

ParameterValue
Reference Power Accuracy±1.2% (1 year, 0 to 55 °C)
Math FunctionsA/B, B/A, A-B, B-A
Instrumentation Accuracy (Logarithmic)±0.02 dB
Instrumentation Accuracy (Linear)±0.5%

Signal Processing

ParameterValue
Averaging (Filter Length)1 to 1024 (auto or manual)
Display Offset Range-99.99 to +99.99 dB

Environmental

ParameterValue
Storage Temperature-20 °C to +70 °C
Operating HumidityUp to 95% relative humidity at 40 °C (non-condensing)
Operating AltitudeUp to 3,000 meters (9,840 ft)
Non-operating AltitudeUp to 15,240 meters (50,000 ft)

Configuration

ParameterValue
Hardware Input OptionsOption 002 (Rear-panel sensor inputs), Option 003 (Rear-panel inputs & reference oscillator)

Compliance

ParameterValue
EMC ComplianceIEC 61326-1, CISPR 11 Group 1 Class A
Safety ComplianceIEC 61010-1

Technology

RF / Microwave Power Measurement

Applications

Benchtop & ATE Systems