The Keysight Agilent E5052A is a high-performance Signal Source Analyzer offering comprehensive phase noise, AM noise, transient, and spectrum measurement capabilities from 10 MHz to 7 GHz. It is widely recognized for its high measurement speed, cross-correlation sensitivity enhancement, and integrated DC bias sources.

Specifications

General

ParameterValue
InterfacesGPIB, LAN (10/100 Base-T), USB 2.0, VGA, Parallel, Serial, PS/2
Display10.4-inch color TFT LCD with touch screen, XGA (1024 x 768)

Physical

ParameterValue
Power Supply90 V to 132 V / 198 V to 264 V AC, 47 Hz to 63 Hz, 300 VA max
Operating Temperature10 to 40
Dimensions (W×H×D)426 mm x 222 mm x 373 mm
Weight19

RF Input Specifications

ParameterValue
RF Input Frequency Range10 MHz to 7 GHz
RF Input Power Level Range-15 dBm to +20 dBm
Maximum Input RF Power+30 dBm (1 W)
Input VSWR< 1.6 (10 MHz to 3 GHz, typical)
RF Input Impedance50 ohms (nominal)
RF Input Connector TypeType-N (female)

Phase Noise Measurement

ParameterValue
Offset Frequency Range1 Hz to 100 MHz
Measurement Carrier Frequency Range10 MHz to 7 GHz
Phase Noise Measurement Relative Accuracy±1.5 dB
Cross-Correlation TechniqueYes
Maximum Number of Cross-Correlations10,000

AM Noise Measurement

ParameterValue
AM Noise Offset Frequency Range10 Hz to 10 MHz
AM Noise Carrier Frequency Range10 MHz to 7 GHz

Baseband Noise Measurement

ParameterValue
Baseband Noise Measurement Range1 Hz to 100 MHz
Baseband Input Impedance50 ohms (nominal)
Baseband Input ConnectorBNC (female)

Transient Measurements

ParameterValue
Wide Band Measurement Span50 MHz to 3 GHz
Narrow Band Measurement Span100 Hz to 25 MHz
Transient Time Resolution10 ns (maximum resolution in wideband mode)
Transient Trigger SourcesRF Level, Frequency, External, Video, GPIB/Manual

Frequency & Power Measurements

ParameterValue
Frequency Measurement Range10 MHz to 7 GHz
RF Power Measurement Range-15 dBm to +20 dBm
RF Power Measurement Resolution0.01 dB

Internal DC Bias Sources

ParameterValue
DC Tuning Voltage Output Range-15 V to +35 V
DC Tuning Voltage Current CapacityMax 20 mA
DC Tuning Voltage Resolution1 mV
DC Supply Voltage Output Range0 V to +16 V
DC Supply Voltage Current CapacityMax 80 mA
DC Supply Voltage Resolution1 mV
DC Current Measurement Range0 to 80 mA
DC Current Measurement Accuracy±(0.2% of reading + 50 µA)

Internal Reference Oscillator

ParameterValue
Reference Frequency10 MHz
External Reference Input Frequency10 MHz ± 10 Hz
External Reference Input Level-3 dBm to +10 dBm
Internal Reference Output Level0 dBm ± 3 dB

Spectrum Monitor Mode

ParameterValue
Frequency Span Range15 Hz to 15 MHz

System Interfaces & Connectivity

ParameterValue
GPIB Interface Port24-pin IEEE-488
LAN Interface PortRJ-45, 10/100 Base-T
USB Interface PortsUSB 2.0 (front and rear panels)
External Monitor Output15-pin D-Sub (VGA)
Reference Input Connector TypeBNC (female), 50 ohms nominal
Reference Output Connector TypeBNC (female), 50 ohms nominal
Trigger Input Connector TypeBNC (female)
DC Control Output TerminalsBNC (female) on front panel (Tuning & Supply)
Keyboard and Mouse InterfacePS/2 or USB

Display & User Interface

ParameterValue
Display Size and Type10.4-inch color TFT LCD
Display Resolution1024 x 768 (XGA)

Physical & Power Specifications

ParameterValue
AC Supply Line Voltage Range90 V to 132 V / 198 V to 264 V (autoswitching)
AC Supply Line Frequency47 Hz to 63 Hz
Maximum Power Consumption300 VA
Operating Temperature Range10 °C to 40 °C
Storage Temperature Range-20 °C to +60 °C
Operating Humidity Range20% to 80% RH (non-condensing)
Dimensions (W x H x D)426 mm x 222 mm x 373 mm

Ordering & Configurable Options

ParameterValue
Frequency Range ExpansionUp to 26.5 GHz with E5053A Microwave Downconverter
Millimeter-wave SupportUp to 110 GHz with E5053A and external mixers
Rackmount Kit OptionOption 1CP compatible

RF Input

ParameterValue
RF Input Attenuator Range0 to 35 dB
RF Input Attenuator Step Size5 dB

Measurement Capabilities

ParameterValue
Measurement ModesPhase Noise, Spectrum Monitor, Frequency/Power/DC Current Transients, Baseband Noise, Frequency/Power/DC Current Sweep

Reference Frequency

ParameterValue
High Stability Reference Aging Rate (Option 1D5)±5 x 10^-10 per day, ±1 x 10^-7 per year
High Stability Reference Temperature Stability (Option 1D5)±1 x 10^-8 (0 °C to 40 °C, referenced to 23 °C)

Compliance

ParameterValue
Safety ComplianceIEC 61010-1:2001 / EN 61010-1:2001, CAN/CSA C22.2 No. 61010-1-04
EMC ComplianceCISPR 11:1997 / EN 55011:1998 Group 1 Class A, EN 61326:1997 + A1:1998 + A2:2001

Rear Panel Connectivity

ParameterValue
Keyboard ConnectorPS/2 (6-pin mini-DIN)
Mouse ConnectorPS/2 (6-pin mini-DIN)
Audio Output Connector Type3.5 mm monaural phone jack
VGA Monitor Connector Type15-pin mini D-sub (female)
LAN Interface Connector TypeRJ-45
GPIB Interface Connector Type24-pin IEEE-488
Rear Panel Reference Input Impedance50 ohms (nominal)
Rear Panel Reference Output Impedance50 ohms (nominal)

System Features

ParameterValue
Measurement Receiver ArchitectureDual-channel cross-correlation with independent internal local oscillators

General Features

ParameterValue
Instrument Warm-up Time30 minutes

Technology

Signal Source Analysis / Phase Noise Measurement

Applications

Phase noise measurement, VCO characterization, transient analysis, clock jitter evaluation