The Keysight Technologies (formerly Agilent) InfiniiVision MSO6012A is a high-performance 100 MHz Mixed Signal Oscilloscope featuring 2 analog channels and 16 digital channels. Built with MegaZoom III technology, it delivers a deep 8 Mpts standard memory, an ultra-fast update rate of up to 100,000 waveforms/second, and a high-definition 6.3-inch XGA display with 256 intensity levels. These features make it highly effective for debugging complex analog, digital, and serial bus designs.

Specifications

General

ParameterValue
InterfacesUSB 2.0 host (2 ports), USB 2.0 device (1 port) standard; LAN (10/100Base-T), GPIB, XGA Video Output optional
Display6.3-inch (160 mm) diagonal color flat-panel TFT-LCD, 320 × 240 pixels (QVGA) with 256 levels of intensity

Physical

ParameterValue
Power Supply100-240 VAC, 50/60 Hz; 100-120 VAC at 400 Hz; Auto-ranging; 120 W maximum
Operating Temperature0 to +55 °C
Dimensions (W×H×D)354 mm × 188 mm × 174 mm
Weight4.9

Analog Vertical System

ParameterValue
Analog Input Channels2
Analog Bandwidth (-3 dB)100 MHz
Calculated Rise Time (10% to 90%)3.5 ns
Input CouplingAC, DC
Input Impedance1 MΩ ± 1% || ~14 pF, or 50 Ω ± 1.5%
Input Sensitivity Range2 mV/div to 5 V/div (1 MΩ), 2 mV/div to 1 V/div (50 Ω)
Vertical Resolution8 bits
Max Input Voltage (1 MOhm)CAT I 400 V (DC + peak AC), CAT II 300 V (DC + peak AC)
Max Input Voltage (50 Ohm)5 Vrms
DC Gain Accuracy±2.0% of full scale
Offset Range1 MΩ: ±5 V (<10 mV/div), ±25 V (10 mV/div to 198 mV/div), ±100 V (200 mV/div to 1.98 V/div), ±400 V (2 to 5 V/div); 50 Ω: ±5 V (<10 mV/div), ±25 V (10 to 198 mV/div), ±100 V (200 mV to 1 V/div)

Digital Vertical System

ParameterValue
Digital Channels16
Threshold GroupingPod 1 (D7 - D0), Pod 2 (D15 - D8)
Threshold SelectionsTTL, CMOS, ECL, user-defined
Threshold Voltage Range±8.0 V in 10 mV steps
Threshold Accuracy±(100 mV + 3% of threshold setting)
Input Impedance (Digital Probe)100 kΩ ± 2% || ~8 pF (with flying-lead probe)
Minimum Voltage Swing500 mVpp
Maximum Toggle Rate250 MHz

Horizontal System

ParameterValue
Maximum Real-time Sample Rate (Analog)2 GSa/s (interleaved), 1 GSa/s (per channel)
Maximum Sample Rate (Digital)1 GSa/s
Maximum Record Length (Analog Channels)8 Mpts (interleaved), 4 Mpts (per channel)
Maximum Record Length (Digital Channels)8 Mpts (interleaved), 4 Mpts (per channel)
Time Base Range5 ns/div to 50 s/div
Time Base Accuracy±15 ppm (0.0015%)
Channel-to-Channel Deskew Range±100 ns
Timebase ModesMain, Zoom, Roll, XY
XY Phase Difference±1° (from DC to 100 kHz)

Trigger System

ParameterValue
Trigger ModesNormal, Auto, Single
Trigger CouplingAC, DC, LF Reject, HF Reject, Noise Reject
Trigger Holdoff Range60 ns to 10 s
Edge TriggeringRising, falling, alternating, or either edge on any channel
Pulse Width TriggeringPositive or negative pulse from 15 ns to 10 s
Runt TriggeringNot supported
Setup and Hold TriggeringTrigger on setup/hold violations
Pattern TriggeringTrigger on logical state combination across any analog/digital channel

Acquisition Modes

ParameterValue
Sample ModeSuccessive samples are registered
Peak Detect ModeCaptures high-frequency glitches down to 2 ns
Averaging ModeSelectable from 2 to 65536 averages
High Resolution ModeReal-time boxcar filtering reduces random noise and yields up to 12 bits of resolution
Segmented Memory ModeSupported, partition acquisition memory into up to 1,000 segments

Waveform Analysis & Measurements

ParameterValue
Automatic MeasurementsVoltage: Peak-to-peak, Max, Min, Amplitude, Top, Base, Overshoot, Preshoot, Average, RMS. Time: Frequency, Period, Width, Duty Cycle, Rise, Fall, Phase, Delay
Math FunctionsAdd, Subtract, Multiply, FFT
FFT Window TypesHanning, Flat Top, Rectangular

Display & User Interface

ParameterValue
Waveform Update Rate100,000 waveforms/second

Environmental & Compliance

ParameterValue
Operating Humidity Range95% relative humidity at 40 °C (non-condensing)
Operating AltitudeUp to 4,600 m (15,000 ft)
Non-Operating AltitudeUp to 15,300 m (50,000 ft)

External Trigger Input

ParameterValue
External Trigger Max Input Voltage400 V (DC + peak AC) CAT I / 300 Vrms CAT II (1 MΩ), 5 Vrms (50 Ω)
External Trigger CouplingDC, AC, LF Reject, HF Reject

Trigger Sensitivity

ParameterValue
Internal Trigger SensitivityGreater of 1 div or 5 mV (< 10 mV/div), 0.6 div (≥ 10 mV/div)
External Trigger Sensitivity40 mV (DC to 100 MHz)

Environmental

ParameterValue
Storage Temperature Range-40 °C to +70 °C
Operating Vibration0.21 g rms (5 Hz to 500 Hz)
Non-Operating Vibration2.09 g rms (5 Hz to 500 Hz)
Operating Shock30 g (half-sine, 11 ms duration)
Non-Operating Shock50 g (half-sine, 11 ms duration)

Electrical

ParameterValue
Maximum Power Consumption120 W

Probe Compensation

ParameterValue
Probe Compensation Output Amplitudeapproximately 2.5 Vpp
Probe Compensation Output Frequencyapproximately 1.2 kHz

Hardware Frequency Counter

ParameterValue
Frequency Counter Resolution5 digits
Frequency Counter Max Frequency100 MHz

Compliance

ParameterValue
Safety Standards ComplianceUL61010-1, CSA C22.2 No. 61010-1, IEC 61010-1
EMC Standards ComplianceIEC 61326-1, EN 61326-1, CISPR 11/EN 55011 Group 1 Class A

Analog Input

ParameterValue
Channel-to-Channel Isolation≥ 40 dB (DC to 100 MHz)
Bandwidth Limit Filter20 MHz (selectable)

Triggering

ParameterValue
Trigger Jitter< 15 ps RMS (typical)
Video Triggering FormatsNTSC, PAL, SECAM, HDTV (480i, 480p, 576i, 720p, 1080i, 1080p)

Digital Channels

ParameterValue
Minimum Glitch Detection Width2 ns

Accessories

ParameterValue
Included Probes (Analog)2 × 10074C 10:1 passive probes (150 MHz)
Included Probes (Digital)1 × 54620-68701 16-channel digital probe kit

Connectivity

ParameterValue
Trigger Out Output0 to 5 V into open circuit, 0 to 2.5 V into 50 Ω

Technology

Mixed Signal Oscilloscope

Applications

Embedded system design, mixed-signal debugging, serial bus testing, automotive and industrial electronics repair