The Keysight InfiniiVision MSO6032A is a 300 MHz, 2+16 channel mixed signal oscilloscope (MSO) featuring MegaZoom III deep memory technology, a high-definition color display with 256 levels of intensity, and an industry-leading update rate of up to 100,000 waveforms/sec. Combining analog and digital channels with advanced serial triggering options, this instrument provides deep insights into complex system-level designs.
Specifications
General
| Parameter | Value |
|---|---|
| Display | 6.3-inch high-definition color LCD with 256 levels of intensity |
| Internal Storage Capacity | 10 setups and 10 waveforms |
| Calibration Cycle | 1 year |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 100-240 VAC, 50/60 Hz, 120 W maximum |
| Operating Temperature | 0 to +55 °C |
| Dimensions (W×H×D) | 354 mm x 188 mm x 174 mm |
| Weight | 4.9 kg |
Analog Vertical System
| Parameter | Value |
|---|---|
| Analog Bandwidth | 300 MHz |
| Number of Analog Channels | 2 |
| Calculated Rise Time | 1.17 ns |
| Input Coupling | AC, DC |
| Input Impedance | 1 MΩ ±1% (14 pF), 50 Ω ±1.5% |
| Input Sensitivity Range | 2 mV/div to 5 V/div (1 MΩ), 2 mV/div to 1 V/div (50 Ω) |
| Vertical Resolution | 8 bits |
| DC Gain Accuracy | ±2.0% of full scale |
| Maximum Input Voltage | CAT I 400 Vpk (1 MΩ), 5 Vrms (50 Ω) |
| Hardware Bandwidth Limits | 20 MHz (selectable) |
Digital Vertical System
| Parameter | Value |
|---|---|
| Number of Digital Channels | 16 |
| Input Impedance (Digital) | 100 kΩ ±2% (~8 pF) at probe tip |
| Minimum Voltage Swing | 500 mV pp |
| Threshold Selections | TTL, CMOS, ECL, user-defined |
| Threshold Accuracy | ±(100 mV + 3% of threshold setting) |
| Maximum Input Voltage (Digital) | ±40 V peak |
| Threshold Range | ±8.0 V in 10 mV steps |
Horizontal System & Acquisition
| Parameter | Value |
|---|---|
| Maximum Real-Time Sample Rate (Analog) | 2 GSa/s (half-channel), 1 GSa/s (each channel) |
| Maximum Real-Time Sample Rate (Digital) | 2 GSa/s (half-pod), 1 GSa/s (each channel) |
| Maximum Memory Depth | 8 Mpts (half-channel / half-pod), 4 Mpts (each channel / pod) |
| Timebase Range | 1 ns/div to 50 s/div |
| Timebase Accuracy | ±15 ppm (0.0015%) |
| Acquisition Modes | Normal, Peak Detect, Average, High Resolution |
| Peak Detection Resolution | 500 ps |
| Waveform Update Rate | 100,000 waveforms/sec |
Trigger System
| Parameter | Value |
|---|---|
| Trigger Sources | Analog channels, digital channels, external, line |
| Trigger Modes | Auto, Normal, Single |
| Trigger Coupling | AC, DC, LF Reject, HF Reject, Noise Reject |
| Trigger Holdoff Range | 60 ns to 10 s |
| Trigger Sensitivity | 0.6 div or 2 mV (DC to 100 MHz), 1.0 div or 3 mV (100 MHz to 300 MHz) |
| Edge Triggering | Rising, falling, or alternating edge on any source |
| Pulse Width Triggering | Trigger on pulse width <, >, =, or ≠ a specific time (2 ns to 10 s on analog channels, 5 ns to 10 s on digital channels) |
| Pattern / State Triggering | Pattern of high, low, or don't care levels, or state on any input |
| TV / Video Triggering | NTSC, PAL, SECAM, HDTV (480i, 480p, 720p, 1080i, 1080p) |
| Serial Protocol Triggering Options | I2C, SPI, CAN, LIN, UART/RS-232, FlexRay, MIL-STD-1553 |
| Pulse Width Trigger Range | 2 ns to 10 s |
| Supported Video Standards | NTSC, PAL, SECAM, HDTV (480i, 480p, 720p, 1080i, 1080p) |
Waveform Measurements & Analysis
| Parameter | Value |
|---|---|
| Automatic Voltage Measurements | Peak-to-peak, maximum, minimum, average, amplitude, top, base, overshoot, preshoot, RMS |
| Automatic Time Measurements | Period, frequency, rise time, fall time, + width, - width, duty cycle |
| Measurement Cursors | Manually placed or track automatic measurements, X and Y axes |
| Waveform Math Functions | Addition, subtraction, multiplication, FFT |
| FFT Window Types | Hanning, Flat Top, Rectangular |
Connectivity & Interfaces
| Parameter | Value |
|---|---|
| USB Host Ports | 1 x front panel, 1 x rear panel |
| USB Device Port | 1 x rear panel (supports USBTMC) |
| LAN Port | RJ-45 (10/100Base-T), LXI class C compliant |
| GPIB Interface | IEEE 488.2 connector on rear panel |
| External Trigger Input | 1 MΩ (±1%), 300 Vrms (400 Vpk) max or 50 Ω (±1.5%), 5 Vrms max |
| Trigger Output Port | BNC on rear panel |
| VGA Video Output | 15-pin D-sub (XGA resolution) |
Power Specifications
| Parameter | Value |
|---|---|
| Line Voltage Range | 100 V to 240 V AC |
| Line Frequency Range | 50/60 Hz (100-240 V), 400 Hz (100-120 V) |
| Power Consumption | 120 W maximum |
Environmental & Safety
| Parameter | Value |
|---|---|
| Storage Temperature Range | -40 °C to +70 °C |
| Operating Humidity Range | 95% RH at +40 °C |
| Operating Altitude | Up to 4,600 meters (15,000 feet) |
| Safety Standards Met | UL61010-1, CAN/CSA-C22.2 No. 61010-1, EN61010-1 |
| Electromagnetic Compatibility | IEC 61326-1, EN 61326-1 |
Physical Characteristics
| Parameter | Value |
|---|---|
| Kensington Lock Slot | Available on rear panel |
Accessories
| Parameter | Value |
|---|---|
| Standard Passive Probes | 10073C 10:1 passive probe (1 per analog channel) |
| Digital Channel Cable Kit | 54620-61801 16-channel cable kit (standard) |
Vertical System (Analog)
| Parameter | Value |
|---|---|
| Channel-to-Channel Isolation | > 40 dB from DC to 300 MHz |
| Channel-to-Channel Deskew Range | ±100 ns |
Vertical System (Digital)
| Parameter | Value |
|---|---|
| Digital-to-Analog Channel Deskew | 2 ns typical, 3 ns maximum |
Acquisition
| Parameter | Value |
|---|---|
| Segmented Memory Max Segments | Up to 1000 (with Option SGM) |
| High Resolution Acquisition Mode | Up to 12 bits of resolution (real-time boxcar averaging) |
Environmental
| Parameter | Value |
|---|---|
| Operating Vibration | 0.21 g rms, 5 Hz to 500 Hz, 10 minutes per axis |
| Non-Operating Vibration | 2.09 g rms, 5 Hz to 500 Hz, 10 minutes per axis |
| Operating Shock | 30 g, half-sine, 11 ms duration |
| Non-Operating Altitude | 15,300 m (50,000 ft) |
| Acoustic Noise | 30 dBA (typical) at 25 °C |
Digital Channels
| Parameter | Value |
|---|---|
| Digital Input Capacitance | 8 pF (approximate) |
| Digital Input Dynamic Range | ±10 V about threshold |
| Digital Channel-to-Channel Skew | 2 ns typical, 3 ns maximum |
| Digital Minimum Toggle Rate | 250 MHz |
Analog Channels
| Parameter | Value |
|---|---|
| Maximum Input Voltage (50 Ω) | 5 Vrms |
| Maximum Input Voltage (1 MΩ) | 400 V (DC + peak AC) CAT I, 300 V (DC + peak AC) CAT II |
| Probe Interface | AutoProbe |
Display & Controls
| Parameter | Value |
|---|---|
| Display Resolution | 640 columns × 480 rows (VGA) |
| Display Size | 6.3-inch (160 mm) diagonal |
| Display Type | Color TFT LCD |
Timebase & Triggering
| Parameter | Value |
|---|---|
| Trigger Jitter | < 100 ps RMS (typical) |
| Timebase Delay Jitter | < 1 ppm |
Built-In Frequency Counter
| Parameter | Value |
|---|---|
| Built-In Frequency Counter Resolution | 5 digits |
| Built-In Frequency Counter Max Frequency | 300 MHz |
| Built-In Frequency Counter Sources | Any analog channel or external trigger source |
Connectivity & Storage
| Parameter | Value |
|---|---|
| LXI Compliance | LXI Class C compliant |
| File Export Formats | CSV, BMP, PNG, ASCII XY, Binary, Setup (.set), Waveform (.wfm) |
Technology
Mixed Signal Oscilloscope (MSO)
Applications
Embedded system design, debugging analog, digital, and serial communication buses (I2C, SPI, CAN, LIN, USB, UART)