The Keysight (Agilent) InfiniiVision MSO7054A is a high-performance 500 MHz mixed signal oscilloscope featuring 4 analog channels and 16 digital timing channels. Designed with the MegaZoom III deep memory architecture, it delivers a standard 8 Mpts record length and an uncompromised waveform update rate of up to 100,000 waveforms per second, all displayed on a spacious 12.1-inch XGA LCD display. This oscilloscope provides exceptional signal visibility and comprehensive analytical tools for complex embedded system debugging.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | USB Host (2 front, 1 rear), USB Device (1 rear), LAN (10/100Base-T), XGA video output |
| Display | 12.1-inch diagonal liquid crystal display (LCD) with XGA (1024 x 768) resolution, 256 intensity levels |
| Built-in Help Languages | English, French, German, Italian, Japanese, Korean, Portuguese, Russian, Simplified Chinese, Spanish, Traditional Chinese |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 100-240 V AC, 50/60 Hz; 100-120 V AC, 400 Hz; 120 W maximum power consumption |
| Operating Temperature | 0 °C to +55 °C |
| Dimensions (W×H×D) | 454 mm × 298 mm × 220 mm |
| Weight | 5.9 kg |
Analog Vertical System
| Parameter | Value |
|---|---|
| Analog Input Channels | 4 |
| Analog Bandwidth (-3 dB) | 500 MHz |
| Calculated Rise Time (10% to 90%) | 700 ps |
| Input Coupling | AC, DC |
| Input Impedance | 1 MΩ ± 1% || 15 pF, 50 Ω ± 1.5% |
| Input Sensitivity Range | 1 mV/div to 5 V/div (1 MΩ), 1 mV/div to 1 V/div (50 Ω) |
| Vertical Resolution | 8 bits |
| Max Input Voltage (1 MΩ) | CAT I 300 Vrms, 400 Vpk; CAT II 150 Vrms |
| Max Input Voltage (50 Ω) | 5 Vrms |
| DC Gain Accuracy | ±2.0% of full scale |
| Offset Range | ±5 V (1 mV/div to 200 mV/div), ±50 V (200 mV/div to 5 V/div) for 1 MΩ |
Digital Vertical System
| Parameter | Value |
|---|---|
| Digital Channels | 16 |
| Threshold Grouping | Pod 1 (D7 - D0), Pod 2 (D15 - D8) |
| Threshold Selections | TTL, CMOS, ECL, User-defined |
| Threshold Voltage Range | ±8.0 V in 10 mV increments |
| Threshold Accuracy | ±(100 mV + 3% of threshold setting) |
| Input Impedance (Digital Probe) | 100 kΩ ± 2% || ~8 pF |
| Max Input Voltage (Digital Probe) | ±40 V peak |
| Minimum Voltage Swing | 500 mV peak-to-peak |
| Minimum Detectable Pulse Width | 2 ns |
| Maximum Toggle Rate | 250 MHz |
Horizontal System
| Parameter | Value |
|---|---|
| Maximum Real-time Sample Rate (Analog) | 4 GSa/s (half-channel interleaved), 2 GSa/s (all channels) |
| Maximum Sample Rate (Digital) | 2 GSa/s (1 pod), 1 GSa/s (2 pods) |
| Maximum Record Length (Analog Channels) | 8 Mpts (half-channel interleaved), 4 Mpts (all channels) |
| Maximum Record Length (Digital Channels) | 8 Mpts (1 pod), 4 Mpts (2 pods) |
| Time Base Range | 500 ps/div to 50 s/div |
| Time Base Accuracy | 15 ppm (0.0015%) |
| Channel-to-Channel Deskew Range | ±100 ns |
Trigger System
| Parameter | Value |
|---|---|
| Trigger Modes | Normal, Auto, Single, Force |
| Trigger Coupling | AC, DC, LF Reject, HF Reject, Noise Reject |
| Trigger Holdoff Range | 60 ns to 10 s |
| Trigger Sensitivity | 0.5 div or 2 mV (DC to 100 MHz), 1.0 div or 4 mV (100 MHz to 500 MHz) |
| Edge Triggering | Rising, falling, alternating, or either edge on any channel |
| Pulse Width Triggering | Trigger on pulse width <, >, =, or ≠ a specific time (15 ns to 10 s) |
| Logic Triggering | Trigger on AND, OR, NAND, NOR, or state pattern across analog and digital channels |
| Auxiliary Trigger Input Impedance | 1 MΩ ± 3% (~8 pF) |
| Auxiliary Trigger Max Input Voltage | CAT I 300 Vrms, CAT II 100 Vrms, 400 Vpk (at 1 MΩ) |
Acquisition Modes
| Parameter | Value |
|---|---|
| Sample Mode | Normal Mode = Acquire sample values only |
| Peak Detect Mode | Capture glitches as narrow as 250 ps (analog channels) |
| Averaging Mode | Selectable from 2, 4, 8, 16, ... up to 65536 averages |
| High Resolution Mode | Real-time boxcar averaging reduces random noise and increases resolution up to 12 bits |
| Segmented Memory Mode | Up to 2000 segments |
Waveform Analysis & Measurements
| Parameter | Value |
|---|---|
| Automatic Voltage Measurements | Peak-to-peak, Maximum, Minimum, Amplitude, Top, Base, Overshoot, Preshoot, Average, RMS |
| Automatic Time/Frequency Measurements | Period, Frequency, Rise time, Fall time, +Width, -Width, Duty cycle, Delay, Phase |
| Math Functions | Addition, subtraction, multiplication, FFT, differentiation, integration, square root |
| FFT Window Types | Hanning, Flat Top, Rectangular |
Display & User Interface
| Parameter | Value |
|---|---|
| Display Type | 12.1-inch color TFT LCD |
| Display Size (Diagonal) | 12.1 in (307 mm) |
| Display Resolution | 1024 × 768 pixels (XGA) |
| Waveform Update Rate | Up to 100,000 waveforms/sec |
| Graticules | 8 x 10 grid |
Connectivity and Interfaces
| Parameter | Value |
|---|---|
| USB Host Ports (Front/Rear) | 1 front port, 1 rear port (USB 2.0 compatible) |
| USB Device Port | 1 rear port (USB 2.0 High Speed) |
| LAN Port (Ethernet) | RJ-45 10/100Base-TX |
| External Video Output Port | XGA out (15-pin D-sub) |
| Auxiliary Output | BNC connector (triggers on trigger out, source out, or mask test) |
| LXI Compliance | LXI Class C |
Environmental & Compliance
| Parameter | Value |
|---|---|
| Operating Temperature Range | 0 °C to +50 °C |
| Non-Operating Temperature Range | -40 °C to +70 °C |
| Operating Humidity Range | Up to 95% RH at +40 °C |
| Operating Altitude | Up to 4,600 m (15,000 ft) |
| Safety Certifications | UL61010-1, CAN/CSA-C22.2 No. 61010-1, EN61010-1 |
Physical Characteristics
| Parameter | Value |
|---|---|
| Width | 454 mm |
| Height | 298 mm |
| Depth | 220 mm |
| Weight (Net) | 5.9 kg (13 lbs) |
Vertical System - Analog Channels
| Parameter | Value |
|---|---|
| Analog Bandwidth Limit | 20 MHz (selectable) |
| Channel-to-Channel Isolation | ≥ 100:1 from DC to 500 MHz |
Interfaces & Connectivity
| Parameter | Value |
|---|---|
| Optional GPIB Interface | Supported via optional N5428A module |
Power Requirements
| Parameter | Value |
|---|---|
| AC Power Line Frequency Range | 50/60/400 Hz (at 100 - 120 V), 50/60 Hz (at 100 - 240 V) |
| Power Consumption | 120 W maximum |
Environmental & Physical
| Parameter | Value |
|---|---|
| Ventilation Clearance | 100 mm (4 inches) on left, right, and rear sides |
Software & Options
| Parameter | Value |
|---|---|
| Supported Serial Protocols (Optional) | I²C, SPI, CAN, LIN, RS-232/UART, I²S, MIL-STD 1553, FlexRay |
Acquisition
| Parameter | Value |
|---|---|
| Glitch Capture | As narrow as 250 ps (using hardware peak detect) |
Triggering
| Parameter | Value |
|---|---|
| Video/TV Triggering | NTSC, PAL, SECAM, HDTV (480i, 480p, 576i, 576p, 720p, 1080i, 1080p) |
| Pattern Triggering | Trigger on high, low, or don't care levels across all analog and digital channels |
| Sequence Triggering | Arm on event A, trigger on event B with delay or reset options |
Standard Accessories
| Parameter | Value |
|---|---|
| Standard Analog Probes Included | Four 10073C 10:1 passive probes (500 MHz) |
| Standard Digital Cable Kit Included | 54620-68701 16-channel digital probe kit |
Storage
| Parameter | Value |
|---|---|
| Internal Non-volatile Storage | Stores up to 10 setups and 10 waveforms |
| File Export Formats | BMP, PNG, CSV, ASCII XY, BIN, Setup (.set), Waveform (.wfm) |
Measurements & Analysis
| Parameter | Value |
|---|---|
| Measurement Statistics | Mean, minimum, maximum, standard deviation, and count displayed for up to 4 measurements simultaneously |
Environmental & Standards
| Parameter | Value |
|---|---|
| Environmental Vibration Resistance | Meets MIL-PRF-28800F Class 3 random vibration |
| Environmental Shock Resistance | Meets MIL-PRF-28800F Class 3 half-sine shock |
Technology
Mixed Signal Oscilloscope
Applications
Embedded systems debug, mixed-signal circuit design, digital design verification, and serial protocol analysis