The Keysight Technologies (formerly Agilent) MSO8064A is a high-performance 4+16 channel Mixed Signal Oscilloscope in the Infiniium 8000 Series. Featuring 600 MHz of analog bandwidth and a real-time sample rate of 4 GSa/s, it seamlessly integrates 4 analog channels and 16 digital channels for comprehensive mixed-signal debugging. Powered by MegaZoom III technology, the MSO8064A offers up to 128 Mpts of deep memory on a responsive 10.4-inch XGA touchscreen display, operating on an open Windows XP Pro environment.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | USB 2.0 Host (front & rear), USB Device, LAN (10/100Base-T), GPIB, VGA, RS-232 |
| Display | 10.4-inch color XGA active-matrix TFT-LCD with touch screen (1024 x 768 resolution, 256 color intensity grades) |
| Operating System | Windows XP Professional |
| Calibration Interval | 1 year |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 100 V AC to 240 V AC (±10%) at 47 Hz to 63 Hz, maximum 440 W power consumption |
| Operating Temperature | 5 °C to +40 °C |
| Dimensions (W×H×D) | 437 mm × 216 mm × 440 mm |
| Weight | 13.9 |
Analog Vertical System
| Parameter | Value |
|---|---|
| Analog Input Channels | 4 |
| Analog Bandwidth (-3 dB) | 600 MHz |
| Calculated Rise Time (10% to 90%) | 583 ps |
| Input Impedance | Selectable 1 MΩ ±1% (~13 pF) or 50 Ω ±1% |
| Input Coupling | AC, DC |
| Vertical Resolution | 8 bits (up to 12 bits with high-resolution mode) |
| Input Sensitivity Range | 1 mV/div to 5 V/div (1 MΩ), 1 mV/div to 1 V/div (50 Ω) |
| Maximum Input Voltage (1 MΩ) | 150 Vrms (CAT I or CAT II) |
| Maximum Input Voltage (50 Ω) | 5 Vrms |
| DC Gain Accuracy | ±2% of full scale |
Digital Vertical System
| Parameter | Value |
|---|---|
| Digital Channels | 16 |
| Threshold Grouping | Pod 1 (D7 - D0) and Pod 2 (D15 - D8) thresholds can be set independently |
| Threshold Selections | TTL, CMOS, ECL, PECL, User-defined |
| Threshold Voltage Range | -8.0 V to +8.0 V (in 10 mV increments) |
| Threshold Accuracy | ±(100 mV + 3% of threshold setting) |
| Input Impedance (Digital Probe) | 100 kΩ ±2% (~8 pF) at probe tip |
| Maximum Input Voltage (Digital Probe) | ±40 V peak |
| Minimum Voltage Swing | 500 mV peak-to-peak |
| Minimum Detectable Pulse Width | 2.5 ns |
| Maximum Toggle Rate | 250 MHz |
Horizontal System
| Parameter | Value |
|---|---|
| Maximum Real-time Sample Rate (Analog) | 4 GSa/s (2 channels interleaved), 2 GSa/s (all 4 channels) |
| Maximum Sample Rate (Digital) | 1 GSa/s |
| Maximum Record Length (Digital) | Up to 32 Mpts (configuration dependent; 1 Mpts standard) |
| Time Base Range | 100 ps/div to 20 s/div |
| Time Base Accuracy | ±0.4 ppm (annual aging rate) |
Trigger System
| Parameter | Value |
|---|---|
| Trigger Modes | Edge, Glitch, Pattern, State, Setup and Hold, Transition, Runt, Timeout, Window, Video, Serial Bus |
| Trigger Coupling | DC, AC, LF Reject, HF Reject, Noise Reject |
| Trigger Holdoff Range | 60 ns to 10 s |
| Trigger Sensitivity | 0.6 div (DC to 600 MHz) |
| Serial Bus Triggering Options | I2C, SPI, CAN, USB (software/hardware options dependent) |
| Trigger Jitter | < 8 ps rms |
Acquisition Modes
| Parameter | Value |
|---|---|
| Acquisition Modes | Sample, Peak Detect, Averaging, High Resolution, Segmented Memory (optional) |
Waveform Analysis & Measurements
| Parameter | Value |
|---|---|
| Automatic Measurements | Up to 38 automatic measurements |
| Waveform Histograms | Vertical and horizontal with math statistics |
| Math Functions | Add, Subtract, Multiply, Divide, FFT, Differentiate, Integrate, Square Root, Absolute Value |
| FFT Window Types | Hanning, Flat Top, Rectangular, Blackman-Harris |
Display & User Interface
| Parameter | Value |
|---|---|
| Waveform Display Formats | YT, XY |
Environmental & Compliance
| Parameter | Value |
|---|---|
| Operating Humidity Range | Up to 95% relative humidity (non-condensing) at +40 °C |
| Operating Altitude | Up to 4,600 m (15,000 ft) |
| Non-Operating Altitude | Up to 15,300 m (50,000 ft) |
PC System
| Parameter | Value |
|---|---|
| CPU | Intel Celeron 2.0 GHz |
| PC System Memory | 512 MB (standard), expandable to 2 GB |
| Hard Drive | 40 GB (or larger) |
Environmental
| Parameter | Value |
|---|---|
| Vibration (Operating) | 0.3 g (rms), 5 Hz to 500 Hz |
| Vibration (Non-Operating) | 2.41 g (rms), 5 Hz to 500 Hz |
| Shock (Operating) | 30 g, half-sine, 11 ms duration |
| Shock (Non-Operating) | 50 g, half-sine, 11 ms duration |
Inputs & Outputs
| Parameter | Value |
|---|---|
| Probe Interface | Agilent AutoProbe interface |
Analog Channel Characteristics
| Parameter | Value |
|---|---|
| Channel-to-Channel Isolation | ≥ 40 dB from DC to 600 MHz |
Connectivity & I/O
| Parameter | Value |
|---|---|
| USB Ports | 5 × USB 2.0 (2 front, 3 rear) |
| LXI Compliance | LXI Class C |
Compliance
| Parameter | Value |
|---|---|
| EMC | IEC 61326-1, CISPR 11/EN 55011 Group 1 Class A |
| Safety Standards | IEC 61010-1, CAN/CSA-C22.2 No. 61010-1, UL 61010-1 |
Technology
Mixed Signal Oscilloscope (MSO)
Applications
Mixed-signal embedded system design, serial bus troubleshooting, digital and analog debug