The Keysight N4433A is a 4-port Electronic Calibration (ECal) module operating from 300 kHz to 20 GHz. Designed for fast, highly accurate, and repeatable calibrations of vector network analyzers (VNAs), it features 3.5 mm RF connectors and is controlled directly via a USB connection, minimizing the risk of operator error and connector wear compared to manual calibration kits.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | USB 2.0 (Type-B port) |
| Display | None (LED status indicators) |
| Warm-up Time | 20 minutes |
Physical
| Parameter | Value |
|---|---|
| Power Supply | USB powered (+5 VDC) |
| Operating Temperature | 15 °C to 40 °C |
| Dimensions (W×H×D) | 121.9 mm × 111.4 mm × 41.2 mm |
| Weight | 1.6 kg |
| Recommended Connector Torque | 0.9 N-m (8 in-lb) |
RF, Electrical & Calibration Performance
| Parameter | Value |
|---|---|
| Frequency Range | 300 kHz to 20 GHz |
| Nominal Impedance | 50 Ω |
| Number of Ports | 4 |
| Maximum Input Power | +20 dBm |
Connectivity & Host Interfaces
| Parameter | Value |
|---|---|
| Host Interface Connection | USB (direct control from VNA or via external PC) |
Software & Programming
| Parameter | Value |
|---|---|
| Supported Network Analyzers | Keysight PNA Series, ENA Series |
Environmental Specifications
| Parameter | Value |
|---|---|
| Storage Temperature Range | -40 °C to +70 °C |
| Operating Humidity | Up to 95% RH (non-condensing) |
Compliance & Certifications
| Parameter | Value |
|---|---|
| Recommended Calibration Interval | 1 year |
Connector Options
| Parameter | Value |
|---|---|
| Option 010 Configuration | Four female 3.5 mm connectors |
| Option 020 Configuration | Four male 3.5 mm connectors |
| Option 120 Configuration | Two male and two female 3.5 mm connectors |
Port Characteristics
| Parameter | Value |
|---|---|
| Connector Type | 3.5 mm |
Electrical
| Parameter | Value |
|---|---|
| Maximum DC Voltage | 0 V |
Features
| Parameter | Value |
|---|---|
| Calibration Data Storage | Internal non-volatile EEPROM |
| Temperature Compensation | Internal temperature sensor with automatic compensation |
Environmental
| Parameter | Value |
|---|---|
| Performance Temperature Range | +20 °C to +30 °C |
Technology
Electronic Calibration (ECal)
Applications