The Keysight N4692A is a high-performance, 2-port microwave Electronic Calibration (ECal) module operating from 10 MHz to 40 GHz with 2.92 mm connectors. It provides fast, highly accurate, and repeatable vector network analyzer calibrations, greatly reducing the time and potential errors associated with traditional mechanical calibration kits.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | USB 2.0 |
Physical
| Parameter | Value |
|---|---|
| Power Supply | USB-powered (+5 VDC) |
| Operating Temperature | 15 °C to 35 °C |
| Dimensions (W×H×D) | 60.1 mm × 21.3 mm × 111.8 mm |
| Weight | 0.29 |
| Recommended Connector Torque | 90 N-cm (8 lb-in) |
RF & Port Specifications
| Parameter | Value |
|---|---|
| Frequency Range | 10 MHz to 40 GHz |
| Number of Ports | 2 |
| Port Connector Type | 2.92 mm |
| Nominal Impedance | 50 Ω |
| Connector Gender Configurations | Option 00A (male-female), Option 00F (female-female), Option 00M (male-male) |
Power & Electrical Limits
| Parameter | Value |
|---|---|
| Maximum RF Input Power (Damage Level) | +20 dBm |
| Maximum DC Voltage Limit | 0 V |
Compatibility & Integration
| Parameter | Value |
|---|---|
| Supported Network Analyzers | Keysight / Agilent PNA and ENA Series Vector Network Analyzers |
Environmental & Compliance
| Parameter | Value |
|---|---|
| Error-Corrected Operating Temperature Range | 20 °C to 30 °C |
| Storage Temperature Range | -40 °C to +70 °C |
Accessories, Calibration & Support
| Parameter | Value |
|---|---|
| Standard Calibration Interval | 1 year |
General Specifications
| Parameter | Value |
|---|---|
| Warm-up Time | 15 minutes |
Error-Corrected Performance
| Parameter | Value |
|---|---|
| Directivity (10 MHz to 2 GHz) | 45 dB |
| Source Match (10 MHz to 2 GHz) | 36 dB |
| Load Match (10 MHz to 2 GHz) | 41 dB |
| Directivity (2 to 20 GHz) | 42 dB |
| Source Match (2 to 20 GHz) | 35 dB |
| Load Match (2 to 20 GHz) | 38 dB |
| Directivity (20 to 40 GHz) | 38 dB |
| Source Match (20 to 40 GHz) | 32 dB |
| Load Match (20 to 40 GHz) | 34 dB |
Ordering Information
| Parameter | Value |
|---|---|
| Calibration Certificate Options | Option 1A7 (ISO 17025 compliant), Option A6J (ANSI Z540 compliant) |
Performance
| Parameter | Value |
|---|---|
| Reflection Tracking (10 MHz to 2 GHz) | ±0.08 dB |
| Reflection Tracking (2 to 20 GHz) | ±0.08 dB |
| Reflection Tracking (20 to 40 GHz) | ±0.14 dB |
| Transmission Tracking (10 MHz to 2 GHz) | ±0.08 dB |
| Transmission Tracking (2 to 20 GHz) | ±0.12 dB |
| Transmission Tracking (20 to 40 GHz) | ±0.20 dB |
Technology
Electronic Calibration (ECal)
Applications
Vector Network Analyzer Calibration