The Keysight (Agilent) N5172B EXG X-Series RF Vector Signal Generator is optimized for manufacturing test, providing accurate and reliable signals for basic parametric testing of components and functional verification of receivers. It delivers industry-leading adjacent channel power ratio (ACPR), error vector magnitude (EVM), and output power in a compact two rack-unit (2U) size.

Specifications

Frequency

ParameterValue
Minimum Frequency9 kHz
Maximum Frequency6 GHz (option dependent)
Frequency Resolution0.01 Hz

Output

ParameterValue
Phase Noise-122 dBc/Hz (at 1 GHz, 20 kHz offset)

Modulation

ParameterValue
ModulationAM, FM, Phase (ØM), Narrow Pulse, I/Q (ASK, FSK, MSK, PSK, QAM, custom I/Q)
Sweep ModesStep, List
Pulse Modulation On/Off Ratio> 80 dB
AM Depth0 to 100%

General

ParameterValue
Interfaces1000BaseT LAN, LXI, USB 2.0, GPIB
Recommended Calibration Cycle3 years

Physical

ParameterValue
Power Supply100-120 VAC (50/60/400 Hz), 220-240 VAC (50/60 Hz)

Frequency Specifications

ParameterValue
Frequency Range9 kHz to 3 GHz or 6 GHz (configuration dependent)

Spectral Purity

ParameterValue
SSB Phase Noise (20 kHz offset)-122 dBc/Hz (at 1 GHz)
Harmonics-35 dBc (at 1 GHz)
Non-harmonics (Spurious)-72 dBc (at 1 GHz)

Internal Baseband Generator

ParameterValue
Baseband Generator Bandwidth120 MHz
Baseband Generator ModesWaveform playback and real-time

Vector / I/Q Modulation

ParameterValue
Adjacent Channel Power Ratio (ACPR)≤ -73 dBc (W-CDMA 64 DPCH)
Error Vector Magnitude (EVM)< 0.4% (80 MHz 802.11ac)

Analog and Pulse Modulation

ParameterValue
AM RateDC to 50 kHz
Pulse Width (Minimum)20 ns (narrow pulse)

Connectivity and Interfaces

ParameterValue
LAN (LXI) Port1000BaseT LAN, LXI
USB PortsUSB 2.0 (compatible with Agilent USB power sensors)
GPIB (IEEE-488) PortStandard

Software and Automation

ParameterValue
Real-time Baseband Generation Supported FormatsLTE, HSPA+, WLAN, GNSS, DVB, and more
Command Compatibility (SCPI)SCPI, backward compatible with ESG, MXG, PSG, and 8648x signal generators

Physical and Compliance

ParameterValue
Rack Unit (RU) Size2U (two rack-unit height)

Configuration Options

ParameterValue
Frequency OptionsOption 503 (9 kHz to 3 GHz), Option 506 (9 kHz to 6 GHz)

Output Protection

ParameterValue
Maximum DC Voltage at RF Output50 VDC

Baseband Generator

ParameterValue
Internal Baseband Generator DAC Resolution16-bit

Environmental

ParameterValue
Operating Temperature0 to 55 °C
Storage Temperature-40 to 70 °C

Output Level

ParameterValue
Amplitude Resolution0.01 dB

Frequency Reference

ParameterValue
External Reference Input Frequency10 MHz (standard); 1 to 50 MHz (with Option 1ER)
Reference Output Frequency10 MHz

Technology

RF Vector Signal Generation

Applications

Manufacturing test, receiver functional verification, component parametric testing