The Keysight (formerly Agilent) N5224A is a high-performance microwave vector network analyzer from the industry-leading PNA Series. Delivering a frequency range of 10 MHz to 43.5 GHz, it is optimized for passive and active device testing. Available in 2-port or 4-port configurations with up to two built-in sources, the N5224A provides exceptional measurement accuracy, featuring a system dynamic range of 127 dB, a 38 dB power sweep range, and advanced linear receivers.

Specifications

General

ParameterValue
InterfacesLAN, USB, GPIB (IEEE-488)
Display10.4-inch color LCD touchscreen
Warm-up Time90 minutes (for full specifications)

Physical

ParameterValue
Dimensions (W×H×D)426 × 266 × 533

General System Specifications

ParameterValue
Frequency Range10 MHz - 43.5 GHz
Frequency Resolution1 Hz
System Impedance50 Ω
Number of Ports2 or 4 (configuration dependent)
Test Port Connector Type2.4 mm (male)
Number of Points per TraceUp to 100,001

Measurement Capabilities & Options

ParameterValue
Standard MeasurementsFull S-parameters (S11, S21, S12, S22, etc.)
Electronic Calibration (ECal) SupportYes
Time Domain Analysis OptionOption 010

Front Panel Interfaces

ParameterValue
Display TypeColor LCD Touchscreen
Display Size10.4 inch

Software & Processing

ParameterValue
Internal Operating SystemWindows
Supported Data FormatsTouchstone (.sNp)
Programming InterfacesSCPI, COM, DCOM

Test Ports

ParameterValue
Test Port Damage Level (DC)40 VDC

Inputs & Outputs

ParameterValue
External Reference Input Frequency10 MHz
Rear Panel Trigger Inputs/OutputsBNC (female)

Sweep & Measurements

ParameterValue
Sweep TypesLinear frequency, Log frequency, CW time, Power sweep, Segment sweep

Source

ParameterValue
Source Power Resolution0.01 dB

Power

ParameterValue
Maximum Power Consumption450 W

Software & Measurement Options

ParameterValue
Frequency Offset OptionOption 080
Scalar Mixer/Converter (SMC) OptionOption 082
Vector Mixer/Converter (VMC) OptionOption 083
Gain Compression Application OptionOption 086
Intermodulation Distortion (IMD) OptionOption 087
Source Phase Control OptionOption 088
Spectrum Analysis OptionOption 090
True-Mode Stimulus OptionOption 460 (requires 4-port configuration)

Hardware Configurations

ParameterValue
Standard 2-Port ConfigurationOption 200 (base) / Option 201 (configurable test set)
Advanced 2-Port ConfigurationOption 219 (configurable test set, extended power range, bias tees)
Standard 4-Port ConfigurationOption 400 (base) / Option 401 (configurable test set)
Advanced 4-Port ConfigurationOption 419 (configurable test set, extended power range, bias tees)

Compliance

ParameterValue
EMC ComplianceIEC/EN 61326-1, CISPR 11
Safety ComplianceIEC/EN 61010-1

Technology

Vector Network Analysis

Applications

Passive and active device testing, filter and amplifier characterization, frequency converters, millimeter-wave analysis, signal integrity, and materials measurements.