The Keysight (Agilent) N9344C Handheld Spectrum Analyzer (HSA) delivers field-ready performance with a tuning range of 9 kHz up to 20 GHz. Designed for demanding field applications, it features a rugged, fanless chassis, automatic LCD brightness control, and automated testing via the unique Task Planner. Its benchtop-grade capabilities are backed by an internal preamplifier, a built-in tracking generator, and integrated GPS.

Specifications

General

ParameterValue
Recommended Calibration Cycle1 year

Physical

ParameterValue
Operating Temperature-10 to +50 °C
Dimensions (W×H×D)318 mm × 207 mm × 69 mm
Weight3.6 kg

Frequency Specifications

ParameterValue
Frequency Range1 MHz to 20 GHz (tunable to 9 kHz)

Bandwidth (RBW/VBW)

ParameterValue
Resolution Bandwidth (RBW) Range10 Hz to 3 MHz
Video Bandwidth (VBW) Range1 Hz to 3 MHz

Amplitude Specifications

ParameterValue
Preamplifier RangeInternal preamplifier up to 20 GHz

Measurement Capabilities & Options

ParameterValue
Standard MeasurementsAdjacent Channel Power (ACP), Occupied Bandwidth (OBW), Channel Power, Spectrum Emission Mask (SEM), Carrier-to-Noise Ratio (C/N)
Modulation AnalysisAM/FM Demodulation Analysis, ASK/FSK Demodulation Analysis
Time-Gated SweepSupported (Time-gated spectrum analysis)
GPS ReceiverBuilt-in GPS receiver and GPS antenna
Task PlannerAutomates routine measurement tasks (up to 20 different tests)
USB Power Sensor SupportAgilent U2000 Series USB power sensors
Spectrum MonitorSpectrogram display, record, and playback
Security FeaturesUser data erase

Connectivity & Interfaces

ParameterValue
RF Input ConnectorType-N female, 50 Ω nominal

Environmental & Reliability

ParameterValue
Enclosure DesignRugged and fanless
Storage Temperature Range-40 °C to +70 °C

Physical Characteristics

ParameterValue
Weight (Without Battery)3.2 kg

Power & Battery System

ParameterValue
Battery TypeRechargeable Lithium-ion

Frequency

ParameterValue
Frequency Reference Aging Rate±1 ppm/year

Amplitude

ParameterValue
Maximum DC Input Voltage±50 VDC

Sweep

ParameterValue
Trigger SourcesFree run, Video, External
Number of Sweep Points461

Traces & Detection

ParameterValue
Trace DetectorsNormal, Positive Peak, Sample, Negative Peak, Average (RMS/Video)
Number of Traces4

Power & Battery

ParameterValue
Battery Operating Time3.5 hours (typical)
Battery Charge Time4 hours (typical, analyzer off)

General & Environmental

ParameterValue
Internal Storage64 MB (nominal)
Environmental ComplianceMIL-PRF-28800F Class 2

Phase Noise

ParameterValue
SSB Phase Noise at 30 kHz Offset<-86 dBc/Hz (at 1 GHz, normalized to 1 Hz)
SSB Phase Noise at 100 kHz Offset<-97 dBc/Hz (at 1 GHz, normalized to 1 Hz)
SSB Phase Noise at 1 MHz Offset<-115 dBc/Hz (at 1 GHz, normalized to 1 Hz)

Tracking Generator

ParameterValue
Tracking Generator Frequency Range5 MHz to 7 GHz (Option TG7 dependent)

Resolution Bandwidth (RBW)

ParameterValue
RBW Shape Factor (Selectivity)< 5:1 (nominal, 60 dB/3 dB bandwidth ratio)

Environmental

ParameterValue
Maximum Operating Altitude4,572 m (15,000 ft)
Maximum Non-operating Altitude15,240 m (50,000 ft)
Vibration and Shock StandardMeets MIL-PRF-28800F Class 2

Technology

Heterodyne Spectrum Analysis

Applications

Field RF Testing, Interference Analysis, Spectrum Monitoring