The Keysight N9928A FieldFox is a handheld microwave vector network analyzer operating from 30 kHz to 26.5 GHz. It provides high benchtop-level accuracy in a rugged, portable form factor, designed to withstand harsh field environments. Featuring a 4-receiver architecture, it supports a wide variety of calibrations and capabilities including full 2-port S-parameter measurements, time domain analysis, distance-to-fault measurements, and integrated power monitoring.

Specifications

General

ParameterValue
InterfacesLAN, USB Host (2 ports), USB Device (Mini-B), External Reference Input (BNC), External Trigger (BNC), MicroSD slot, Headphone jack, GPS antenna (SMA)
Display6.5-inch transflective color LCD, 640 x 480 resolution
Kensington Lock SlotStandard Kensington lock slot

Physical

ParameterValue
Operating Temperature-10 to 55 °C
Dimensions (W×H×D)292 x 188 x 72 mm
Weight3.0 kg (with battery)

Frequency & Stimulus Specifications

ParameterValue
Frequency Range30 kHz to 26.5 GHz
Frequency Resolution1 Hz
Frequency Aging Rate±1 ppm/year
Number of Test Ports2
Test Port Impedance50 Ω (nominal), 75 Ω (with appropriate adapter/cal kit)
Output Power Resolution / Step Size1 dB
Output Power RangeHigh, low, and manual (Low power is -45 dBm nominal)
Test Port Output Power (30 kHz to 300 kHz)-11 dBm (typical)
Test Port Output Power (>300 kHz to 2 MHz)-3 dBm (typical)
Test Port Output Power (>2 MHz to 625 MHz)-2 dBm (typical)
Test Port Output Power (>625 MHz to 3 GHz)+1 dBm (typical)
Test Port Output Power (3 to 6.5 GHz)-1 dBm (typical)
Test Port Output Power (6.5 to 9 GHz)-2 dBm (typical)
Test Port Output Power (9 to 14 GHz)-4 dBm (typical)
Test Port Output Power (14 to 18 GHz)-6 dBm (typical)
Test Port Output Power (18 to 23 GHz)-10 dBm (typical)
Test Port Output Power (23 to 26.5 GHz)-12 dBm (typical)

Receiver & Dynamic Range Specifications

ParameterValue
System Dynamic Range (>300 kHz to 9 GHz)100 dB (typical, 300 Hz IFBW)
System Dynamic Range (9 to 14 GHz)97 dB (typical, 300 Hz IFBW)
System Dynamic Range (14 to 18 GHz)94 dB (typical, 300 Hz IFBW)
System Dynamic Range (18 to 20 GHz)90 dB (typical, 300 Hz IFBW)
System Dynamic Range (20 to 25 GHz)79 dB (typical, 300 Hz IFBW)
System Dynamic Range (25 to 26.5 GHz)70 dB (typical, 300 Hz IFBW)
Trace Noise Magnitude (>300 kHz to 10 GHz)±0.002 dB rms (300 Hz IFBW)
Trace Noise Phase (>300 kHz to 10 GHz)±0.014 degrees rms (300 Hz IFBW)
Trace Noise Magnitude (10 to 20 GHz)±0.004 dB rms (300 Hz IFBW)
Trace Noise Phase (10 to 20 GHz)±0.027 degrees rms (300 Hz IFBW)
Trace Noise Magnitude (20 to 26.5 GHz)±0.010 dB rms (300 Hz IFBW)
Trace Noise Phase (20 to 26.5 GHz)±0.066 degrees rms (300 Hz IFBW)
Maximum RF Input Damage Level+27 dBm
Maximum DC Input Damage Level±40 VDC

Measurement Speed & Settings

ParameterValue
IF Bandwidths10 Hz, 30 Hz, 100 Hz, 300 Hz, 1 kHz, 3 kHz, 10 kHz, 30 kHz, 100 kHz
Number of Measurement Points101, 201, 401, 601, 801, 1001, 1601, 4001, 10001 (arbitrary points settable through SCPI)

Measurement Capabilities & Calibration

ParameterValue
Measured Parameters (T/R)S11, S21
Measured Parameters (Full 2-Port)S11, S21, S22, S12 (requires Option 211)
Calibration TypesCalReady, 1-port QuickCal, 1-port SOL, 1-port Frequency Response, Enhanced Response, 2-port QuickCal, 2-port SOLT/Offset Short, 2-port SOLT, 2-port Unknown Thru
Supported Calibration KitsType-N 50 Ω, Type-N 75 Ω, 7/16, TNC, 3.5 mm, 2.4 mm, Waveguide (WR-90, WR-62, WR-42), custom coaxial/waveguide
Time Domain OptionSupported (Option 010; includes gating)
Cable and Antenna Analyzer OptionSupported (Option 305; Distance-to-fault, return loss, cable loss)
Vector Voltmeter (VVM) OptionSupported (Option 308)
Built-in Power Meter OptionSupported (Option 310)
Built-in GPS Receiver OptionSupported (Option 307)

Hardware Interfaces & Connections

ParameterValue
Test Port Connector Type3.5 mm male (50 Ω nominal)
External Reference Input Frequency10 MHz
External Reference Input Level-3 to +10 dBm
External Trigger Input/OutputBNC female, TTL compatible
USB Ports2 x USB 2.0 Host, 1 x Mini-USB Device (Type-B)
LAN Interface100Base-T (RJ-45)
GPS Antenna InputSMA female

System, Controller & Storage

ParameterValue
Data Storage Formats.s1p, .s2p, .csv, .txt, .png, .state
Remote Control LanguageSCPI
Storage MediaInternal Flash, SD card, or USB Flash Drive

Physical, Environmental & Power

ParameterValue
Storage Temperature Range-51 °C to 71 °C (without battery)
Environmental ResistanceMeets MIL-PRF-28800F Class 2 requirements
Ingress Protection RatingIP53
Battery Life4 hours (typical)
Battery Charging Time4 hours (typical, analyzer off)
Warranty3-year standard

Performance

ParameterValue
Sweep Speed (201 Points, 1-Port Cal)210 ms (100 kHz IFBW)
Sweep Speed (201 Points, 2-Port Cal)400 ms (100 kHz IFBW)

Inputs & Outputs

ParameterValue
External Reference Output Frequency10 MHz
External Reference Output Level0 to 10 dBm (nominal)
Headphone Jack3.5 mm miniature jack
SD Card SlotSupports SD/SDHC card (up to 16 GB)
Internal SpeakerBuilt-in speaker

Power

ParameterValue
Battery Type10.8 V, 4.6 Ah, 50 Wh Lithium-Ion
AC/DC Adapter Output15 VDC, 4.0 A
Power Consumption (Typical)14 W (battery charging off)
External DC Input15 to 19 VDC, 4 A

Environmental

ParameterValue
Maximum Operating Altitude9,144 m (30,000 ft)
Maximum Non-Operating Altitude15,240 m (50,000 ft)
Operating Humidity95% relative humidity at 40 °C (non-condensing)
Battery Charging Temperature Range0 to 45 °C

Compliance & Standards

ParameterValue
Vibration StandardMeets MIL-PRF-28800F Class 2, MIL-STD-810G, Method 514.6, Procedure I
Explosive Atmosphere StandardMIL-STD-810G, Method 511.5, Procedure I

Options

ParameterValue
USB Power Sensor Support (Option 302)Enables connection to Keysight USB power sensors
Remote Control Capability (Option 030)Enables remote control via LAN using SCPI or FieldFox iOS application

Measurement Capabilities

ParameterValue
Group Delay Aperture0.25% to 20% of span
Mixed-Mode S-Parameters (Option 212)Supported
TDR Cable Measurements (Option 330)Supported

Compliance

ParameterValue
EMC Standard ComplianceIEC/EN 61326-1, CISPR 11 Class A, Group 1
Safety Standard ComplianceIEC/EN 61010-1

Calibration

ParameterValue
QuickCal (Option 112)Supported

Display

ParameterValue
Display Size6.5 in (165 mm) diagonal
Display Resolution640 x 480 VGA

Technology

Vector Network Analysis (VNA)

Applications

Microwave antenna, cable, and transmitter/receiver installation, maintenance, and troubleshooting in industrial, aerospace, defense, and field environments.