The Keysight N7952A is an Advanced Power System (APS) N7900 Series dynamic DC power supply, offering 1000 W of clean, high-speed power. Designed for applications requiring high-speed dynamic sourcing and measurement, it features 18-bit digitizers, fast transient response, and versatile power-sinking capabilities up to 100% of its rating with an optional power dissipator.

Specifications

General

ParameterValue
InterfacesGPIB, USB 2.0, LAN (LXI Class C compliant), Digital I/O
DisplayColor TFT LCD display showing voltage, current, and status
Command LanguageSCPI

Physical

ParameterValue
Power Supply100 VAC to 240 VAC (±10%), 50/60 Hz, single phase
Operating Temperature0 to 55 °C
Dimensions (W×H×D)425.6 mm × 88.1 mm × 558.3 mm
Weight14.1

Core Signal and Electrical Performance

ParameterValue
Output Voltage Rating0 to 40 V
Output Current Rating0 to 25 A
Output Power Rating1000 W
Voltage Programming Accuracy0.03% + 15 mV
Current Programming Accuracy0.04% + 15 mA
Voltage Load Regulation< 1.5 mV
Current Load Regulation< 2 mA
Voltage Line Regulation< 0.5 mV
Current Line Regulation< 1 mA
Voltage Ripple & Noise (RMS, 20 Hz to 20 MHz)1 mV
Voltage Ripple & Noise (Peak-to-Peak, 20 Hz to 20 MHz)9 mV
Current Ripple & Noise (RMS, 20 Hz to 20 MHz)15 mA
Standard Sinking Current (Continuous)2.5 A (10% of rated current)
Maximum Sinking Current (with N7909A)25 A (100% of rated current with external N7909A Power Dissipator)
Overvoltage Protection (OVP) Range0.5 V to 44 V
Overcurrent Protection (OCP) Range0.5 A to 27.5 A

Measurement Capabilities and Timing

ParameterValue
Voltage Measurement Accuracy0.03% + 15 mV
Current Measurement Accuracy0.04% + 15 mA
Transient Voltage Recovery Time< 100 µs (for voltage recovery to within 100 mV after a 50% to 100% load change)
Digitizer A/D Converter Resolution18-bit
Maximum Digitizing Sampling Rate200 kHz (5 µs interval)
Digitizer Buffer Size512k samples per measurement trace
Command Processing Time< 2 ms

Power and Electrical Specifications

ParameterValue
Power Consumption (Maximum)2250 VA

Physical and Mechanical Attributes

ParameterValue
Form FactorBenchtop / Rackmount (2U height)
Rackmount EIA Unit Height2 U
Cooling MethodFan cooled, variable speed, front-to-rear airflow

Environmental and Compliance Standards

ParameterValue
Storage Temperature Range-20 to 70 °C
Relative Humidity LimitsUp to 95% non-condensing
EMC ComplianceConforms to European EMC Directive, IEC/EN 61326-1
Safety StandardsComplies with European Low Voltage Directive, UL 61010-1, CSA C22.2 No. 61010-1

Programming Resolution

ParameterValue
Voltage Programming Resolution0.8 mV
Current Programming Resolution0.12 mA

Measurement Resolution

ParameterValue
Voltage Measurement Resolution0.17 mV
Current Measurement Resolution0.06 mA

Transient Performance

ParameterValue
Voltage Upward Transition Time≤ 0.5 ms (10% to 90% of full scale with resistive load)
Voltage Downward Transition Time≤ 0.35 ms (90% to 10% of full scale with resistive load)

Isolation

ParameterValue
Maximum Float Voltage±240 VDC (from chassis ground)

Analog & Digital I/O

ParameterValue
Digital I/O Port15-pin connector

Environmental

ParameterValue
Maximum Operating Altitude2000 m

General Features

ParameterValue
LXI ComplianceLXI Class C / LXI Core 2011
Parallel OperationUp to 4 units in parallel (using active current sharing)
Series OperationNot supported

Output Specifications

ParameterValue
Programmable Output Resistance Range0 to 1.6 Ω
Common Mode RF Current (RMS)< 1 mA (20 Hz to 20 MHz)
Common Mode RF Current (Peak-to-Peak)< 15 mA (20 Hz to 20 MHz)

Sinking Specifications

ParameterValue
Active Downprogramming Peak Sinking Power1000 W (without N7909A)
Active Downprogramming Continuous Sinking Power100 W (without N7909A)
Active Downprogramming Continuous Sinking Power with N7909A1000 W

AC Input

ParameterValue
AC Input Voltage Range85 VAC to 264 VAC
Nominal AC Input Voltage100 VAC to 240 VAC
AC Input Frequency Range47 Hz to 63 Hz
AC Input Power Factor0.99 (at nominal input and full load)

Digitizer / Measurement

ParameterValue
Digitizer Sampling Interval Range5 µs to 10 s

Transient Response

ParameterValue
Transient Voltage Recovery Settling Band100 mV (for 50% to 100% load change)

Technology

Switching DC Power Supply with Active Down-programmer

Applications

Dynamic current profile testing, battery emulation, high-speed ATE power sourcing, and aerospace/defense test systems