The Keysight (Agilent/HP) 53132A Option 030 is a high-performance Universal Frequency Counter/Timer. This specific configuration features a 225 MHz two-channel base unit enhanced with the Option 030 3 GHz Channel 3 input, delivering 12-digit-per-second frequency resolution and 150 ps single-shot time interval resolution.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | GPIB (IEEE-488.2, SCPI compatible), RS-232 (Talk-Only) |
| Display | 12-digit Vacuum Fluorescent Display (VFD) |
| Save/Recall Instrument Setups | 20 setups |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 100-120 VAC ±10% (50/60/400 Hz), 220-240 VAC ±10% (50/60 Hz), 100 VA maximum |
| Operating Temperature | 0 °C to 55 °C |
| Dimensions (W×H×D) | 212.6 mm × 88.5 mm × 348.3 mm |
| Weight | 3.5 kg |
Input Channels 1 & 2 Characteristics
| Parameter | Value |
|---|---|
| Frequency Range (DC Coupled) - Ch 1 & 2 | 0.1 Hz to 225 MHz |
| Frequency Range (AC Coupled) - Ch 1 & 2 | 30 Hz to 225 MHz (1 MΩ), 1 MHz to 225 MHz (50 Ω) |
| Input Impedance - Ch 1 & 2 | 1 MΩ (parallel with 30 pF) or 50 Ω |
| Input Coupling - Ch 1 & 2 | AC or DC |
| Input Attenuation - Ch 1 & 2 | x1 or x10 |
| Low-Pass Filter Selectability - Ch 1 & 2 | 100 kHz cutoff, switchable |
| Trigger Level Range - Ch 1 & 2 | -5.125 V to +5.125 V (x1), -51.25 V to +51.25 V (x10) |
| Trigger Level Resolution - Ch 1 & 2 | 2.5 mV (x1), 25 mV (x10) |
| Input Sensitivity (DC - 100 MHz) - Ch 1 & 2 | 20 mVrms |
| Input Sensitivity (100 MHz - 225 MHz) - Ch 1 & 2 | 40 mVrms |
| Maximum Input Damage Level (1 MΩ) - Ch 1 & 2 | 350 V (DC + AC peak) to 10 kHz, derating to 5 Vrms at 100 kHz |
| Maximum Input Damage Level (50 Ω) - Ch 1 & 2 | 5 Vrms |
Optional Channel 3 (High-Frequency) Characteristics
| Parameter | Value |
|---|---|
| Frequency Range - Ch 3 (Option 030) | 100 MHz to 3 GHz |
| Input Impedance - Ch 3 (Option 030) | 50 Ω nominal |
| Input Coupling - Ch 3 (Option 030) | AC |
| Connector Type - Ch 3 (Option 030) | BNC female (front panel) |
| Input Sensitivity (100 MHz - 2.7 GHz) - Ch 3 | 40 mVrms |
| Input Sensitivity (2.7 GHz - 3.0 GHz) - Ch 3 | 75 mVrms |
| Maximum Input Damage Level - Ch 3 (Option 030) | 5 Vrms (+27 dBm) |
Measurement Capabilities & Modes
| Parameter | Value |
|---|---|
| Frequency Resolution | 12 digits/second |
| Single-Shot Time Interval Resolution | 150 ps |
| Time Interval Range | -1 ns to 10^5 s |
| Time Interval Delay Range | 100 ns to 10 s |
| Rise/Fall Time Measurement Range | 5 ns to 10 μs |
| Pulse Width Measurement Range | 5 ns to 10^5 s |
| Duty Cycle Measurement Range | 0% to 100% (5 ns minimum pulse width) |
| Phase Measurement Range | -180° to +360° |
| Frequency Ratio Measurement Channels | Ch 1/Ch 2, Ch 1/Ch 3, Ch 2/Ch 1, Ch 3/Ch 1 |
| Voltage Peak-to-Peak Measurement Range | 100 Hz to 20 MHz, 20 mV to 10 Vp-p (Ch 1 & 2) |
Timebase & Reference Oscillator Options
| Parameter | Value |
|---|---|
| Standard Timebase Aging Rate | < 3.7 x 10^-7 per day |
| Standard Timebase Temperature Stability | < 5 x 10^-6 (0 °C to 50 °C) |
| External Reference Input Frequencies | 1, 5, or 10 MHz |
| External Reference Input Amplitude Range | 100 mVrms to 5 Vrms |
| External Reference Output Frequency | 10 MHz |
| External Reference Output Amplitude | > 1 Vp-p into 50 Ω |
Gating, Triggering & Math Functions
| Parameter | Value |
|---|---|
| Gate Time Range | 1 ms to 10.00 s |
| Gate Time Resolution | 1 ms |
| Statistics Functions | Mean, Min, Max, Standard Deviation |
| Allan Deviation Measurement | Calculated from statistical measurements |
| Math Functions | Scale and Offset: (Measure - Offset) / Scale |
| Limit Testing Capabilities | High/Low limits with visual fail/pass indication |
Connectivity & System Interfaces
| Parameter | Value |
|---|---|
| SCPI Command Compatibility Mode | SCPI-1992 compliant |
| Backward Compatibility Emulation Modes | HP 5335A |
Power & Environmental Specifications
| Parameter | Value |
|---|---|
| Storage Temperature Range | -40 °C to +70 °C |
| Calibration Interval | 1 year |
Physical Characteristics & Compliance
| Parameter | Value |
|---|---|
| Safety Compliance Certifications | EN 61010-1, CSA C22.2 No. 1010.1 |
| EMC/EMI Compliance Standard | EN 55011, EN 50082-1 |
Optional Timebases
| Parameter | Value |
|---|---|
| Option 001 Medium Stability Oven Timebase Aging Rate | < 2 x 10^-7 per month |
| Option 001 Medium Stability Oven Timebase Temperature Stability (0 to 50 °C) | < 1 x 10^-6 |
| Option 010 High Stability Oven Timebase Aging Rate | < 5 x 10^-10 per day, < 1.5 x 10^-8 per month |
| Option 010 High Stability Oven Timebase Temperature Stability (0 to 50 °C) | < 2.5 x 10^-9 |
| Option 012 Ultra-High Stability Oven Timebase Aging Rate | < 1 x 10^-10 per day, < 3 x 10^-9 per month |
| Option 012 Ultra-High Stability Oven Timebase Temperature Stability (0 to 50 °C) | < 2.5 x 10^-9 |
Channel 3 (Option 030)
| Parameter | Value |
|---|---|
| Channel 3 VSWR (Option 030) | < 2.5:1 (typical) |
Measurement Speed
| Parameter | Value |
|---|---|
| GPIB Measurement Transfer Rate | Up to 200 measurements/s |
Triggering
| Parameter | Value |
|---|---|
| Auto Trigger Frequency Range (Ch 1 & 2) | 100 Hz to 225 MHz |
| Auto Trigger Minimum Amplitude (Ch 1 & 2) | > 100 mVpp |
| Trigger Holdoff Range | Not supported |
Power Requirements
| Parameter | Value |
|---|---|
| AC Voltage Input Range | 90 VAC to 265 VAC |
| AC Frequency Input Range | 47 Hz to 63 Hz (for 90-265 VAC), 360 Hz to 440 Hz (for 90-132 VAC) |
| Power Consumption | 170 VA max (30 W typical) |
Physical Specifications
| Parameter | Value |
|---|---|
| Rackmount Form Factor | 2U, half-rack width |
Interfaces & Connectivity
| Parameter | Value |
|---|---|
| GPIB Interface Function Codes | SH1, AH1, T5, TE0, L4, LE0, SR1, RL1, PP0, DC1, DT1, C0, E2 |
Input Characteristics - Ch 1 & 2
| Parameter | Value |
|---|---|
| Low-Pass Filter Cutoff Frequency | 100 kHz (nominal) |
Timebase
| Parameter | Value |
|---|---|
| Internal Timebase Frequency | 10 MHz |
Inputs & Outputs
| Parameter | Value |
|---|---|
| External Gate Input Impedance | 1 kΩ (nominal) |
| External Gate Input Threshold | TTL compatible |
| External Gate Minimum Pulse Width | 50 ns |
Environmental
| Parameter | Value |
|---|---|
| Maximum Operating Altitude | 4,600 m |
| Operating Humidity | Up to 95% RH at 40 °C (non-condensing) |
Interfaces
| Parameter | Value |
|---|---|
| RS-232 Baud Rates | 300, 1200, 2400, 4800, 9600, 19200 bps |
Technology
Universal Frequency Counter/Timer
Applications
Precision time and frequency measurement