The Keysight DAQ970A is a high-performance Data Acquisition System and Data Logger. Equipped with a built-in 6.5-digit digital multimeter (DMM) and a 3-slot mainframe, it is designed for highly accurate and versatile multiplexing, switching, and scanning applications.

Specifications

General

ParameterValue
InterfacesUSB 2.0 Host (front panel), USB 2.0 Device (rear panel), LAN (10/100 Base-T, LXI-C compliant)
Display4.3-inch color TFT graphical display, 480 x 272 resolution with built-in trend charting and histograms

Physical

ParameterValue
Power Supply100 V to 240 V AC ±10%, 50/60 Hz ±10%, 400 Hz ±10%, 150 VA maximum power consumption
Operating Temperature0 °C to 55 °C
Dimensions (W×H×D)261.2 mm x 104.0 mm x 374.0 mm (with bumpers and feet)
Weight4.2 kg

System Performance

ParameterValue
Internal DMM Resolution6.5 digits (22-bit)
Maximum Scan Rate450 channels/second
Maximum Reading Speed to Memory50,000 readings/second
Internal Reading Memory Capacity1,000,000 readings with timestamps
Number of Module Card Slots3
Maximum Channels per Mainframe120 channels (using 3x DAQM908A 40-channel multiplexers)
A/D Converter Integration Times0.02 PLC, 0.2 PLC, 1 PLC, 10 PLC, 100 PLC

DC Voltage Measurements

ParameterValue
DC Voltage Ranges100 mV, 1 V, 10 V, 100 V, 300 V
DC Voltage Maximum Resolution100 nV
DCV 1-Year Basic Accuracy0.0035% of reading + 0.0005% of range (10 V range)
DCV Input Impedance (100mV to 10V ranges)> 10 GΩ (selectable) or 10 MΩ ± 1%
DCV Input Impedance (100V and 300V ranges)10 MΩ ± 1%
Maximum Input Voltage300 V on any input terminal

AC Voltage Measurements

ParameterValue
AC Voltage Measurement TypeTrue RMS AC-coupled
AC Voltage Ranges100 mV, 1 V, 10 V, 100 V, 300 V
AC Voltage Frequency Range3 Hz to 300 kHz
ACV 1-Year Basic Accuracy0.06% of reading + 0.03% of range (10 Hz to 20 kHz, 10 V range)
ACV Crest Factor3:1 at full scale

Resistance Measurements

ParameterValue
Resistance Measurement Configurations2-wire and 4-wire
Resistance Ranges100 Ω, 1 kΩ, 10 kΩ, 100 kΩ, 1 MΩ, 10 MΩ, 100 MΩ, 1 GΩ
Resistance Maximum Resolution100 μΩ
Resistance 1-Year Basic Accuracy0.01% of reading + 0.001% of range (10 kΩ range)

DC Current Measurements

ParameterValue
DC Current Ranges1 μA, 10 μA, 100 μA, 1 mA, 10 mA, 100 mA, 1 A
DC Current Maximum Resolution1 pA
DCI 1-Year Basic Accuracy0.05% of reading + 0.005% of range (10 mA range)

AC Current Measurements

ParameterValue
AC Current Ranges100 μA, 1 mA, 10 mA, 100 mA, 1 A
AC Current Measurement TypeTrue RMS AC-coupled
AC Current Frequency Range3 Hz to 10 kHz
ACI 1-Year Basic Accuracy0.1% of reading + 0.04% of range (10 mA range, 10 Hz to 5 kHz)

Frequency and Period Measurements

ParameterValue
Frequency Measurement Range3 Hz to 300 kHz
Period Measurement Range3.3 μs to 333 ms
Frequency 1-Year Basic Accuracy0.01% of reading

Capacitance Measurements

ParameterValue
Capacitance Measurement Ranges1 nF, 10 nF, 100 nF, 1 μF, 10 μF, 100 μF
Capacitance 1-Year Basic Accuracy1% of reading + 0.1% of range

Temperature Measurements

ParameterValue
Thermocouple Types SupportedB, E, J, K, N, R, S, T
RTD Types SupportedPT100, PT1000 (2-wire and 4-wire)
Thermistor Types Supported2.2 kΩ, 5 kΩ, 10 kΩ
Internal CJC Accuracy±1.0 °C (typical)

Triggering and Alarms

ParameterValue
Scan Trigger SourcesInternal Timer, External, Software, Front Panel (Manual), Bus
Alarm Limits per Channel2 limits (High, Low)
Alarm Outputs4 hardware alarm outputs (TTL)
External Trigger InputTTL compatible, selectable edge
Trigger Output PulseTTL compatible, selectable polarity
Built-in Math FunctionsScaling (mX+B), dB, dBm, Limit Testing, Min/Max/Average

Supported Modules

ParameterValue
Supported Multiplexer ModulesDAQM900A, DAQM901A, DAQM902A, DAQM908A
Supported General Purpose & Matrix ModulesDAQM903A, DAQM904A
Supported RF ModulesDAQM905A
Supported Multifunction ModulesDAQM907A
Supported Digitizer ModulesDAQM909A (4-channel simultaneous digitizer)

Connectivity and Control

ParameterValue
Programming Language SupportSCPI, Agilent 34970A/34972A compatibility modes
Built-in Web Server FeaturesConfiguration, monitoring, data logging control, SCPI console
Companion Desktop SoftwareKeysight PathWave BenchVue DAQ App
Instrument Drivers SupportedIVI-C, IVI-COM, LabVIEW

Physical and General

ParameterValue
Form FactorBenchtop or 2U half-rack
Kensington Lock SlotAvailable on rear panel
Line Frequency45 Hz to 66 Hz, and 360 Hz to 440 Hz
Power-On Configuration OptionsLast State, Factory Default, or User Preset
Storage Temperature Range-40 °C to 70 °C
Operating Humidity RangeUp to 80% RH at 40 °C (non-condensing)
Maximum Operating Altitude3,000 meters (9,842 feet)
Warm-up Time60 minutes
Calibration Interval1 year or 2 years

Compliance and Certifications

ParameterValue
Safety StandardsIEC 61010-1, EN 61010-1, UL 61010-1, CAN/CSA-C22.2 No. 61010-1
EMC StandardsIEC 61326-1, EN 61326-1, ICES-001, AS/NZS CISPR 11
Environmental ComplianceRoHS, WEEE

Power Requirements

ParameterValue
Maximum Power Consumption75 VA
AC Power Input Voltage Range100 V to 240 V (±10%)
AC Power Input Frequency Range50 Hz to 60 Hz (±10%), 400 Hz (±10% at 100 V to 120 V)

System Characteristics

ParameterValue
Non-volatile State Storage Capacity5 instrument states

Connectivity

ParameterValue
USB Host Port LocationFront panel
USB Host Port CompatibilityUSB 2.0 (supports FAT32 format USB flash drives)
USB Device Port LocationRear panel
USB Device Port ProtocolUSBTMC (USB 2.0)
LAN Interface Speed and Standard10BASE-T/100BASE-TX, LXI Core 2011 Device Specification Version 1.4

Environmental Compliance

ParameterValue
Vibration and Shock ComplianceMIL-PRF-28800F, Class 3
Pollution Degree RatingPollution Degree 2

Measurement Parameters

ParameterValue
Continuity Test Current Source1 mA
Diode Test Current Source1 mA

Measurement Performance

ParameterValue
DC Common Mode Rejection (CMRR)140 dB (with 1 kΩ unbalance in LO)
AC Common Mode Rejection (CMRR)70 dB (at 50/60 Hz, 1 kΩ unbalance in LO)
Normal Mode Rejection (NMRR)60 dB (at 50/60 Hz ± 0.1%, for integration times ≥ 1 PLC)
DC Input Bias Current< 30 pA at 25 °C
A/D Converter Linearity0.0002% of reading + 0.0001% of range
Overrange Capability120% of range (except max voltage and current ranges)
Internal DMM Current Fuse Protection3.15 A, 250 V fast-acting fuse
Frequency/Period Gate Times1 ms, 10 ms, 100 ms, 1 s
Continuity ThresholdFixed at 10 Ω

Interfaces and Software

ParameterValue
LXI ComplianceLXI Device Specification 2011, Class C / Core

Technology

Data Acquisition System (DAQ) / Data Logger

Applications

Automated testing, temperature logging, environmental monitoring, and multi-channel sensor characterization