The Keysight N9962A FieldFox is a high-performance handheld microwave spectrum analyzer operating from 5 kHz up to 50 GHz. Designed for field testing, the N9962A is rugged, complies with MIL-PRF-28800F Class 2 standards, and features an IP53 rated chassis. It offers laboratory-grade spectrum analysis, including low displayed average noise levels (DANL), exceptional phase noise, and fast sweep speeds, with optional upgrades for real-time spectrum analysis (RTSA) and interference analysis.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | 2 x USB 2.0 Host, 1 x USB 2.0 Device, LAN (10/100Base-T), SD/SDHC slot |
| Display | 6.5-inch transflective color LCD, 640 x 480 resolution |
| Warm-up Time | 20 minutes |
| Internal Storage Capacity | 4 GB nominal |
Physical
| Parameter | Value |
|---|---|
| Power Supply | AC adapter (100 to 240 VAC, 50 to 60 Hz input; 15 VDC, 4 A output) and rechargeable Li-ion battery (10.8 V, 4.6 Ah) |
| Operating Temperature | -10 to +55 °C |
| Dimensions (W×H×D) | 292 mm x 188 mm x 72 mm |
| Weight | 3.2 kg (including battery) |
Core Signal & Electrical Performance
| Parameter | Value |
|---|---|
| Frequency Range | 9 kHz to 50 GHz |
| Frequency Resolution | 1 Hz |
| Frequency Span Range | 10 Hz to 50 GHz, plus 0 Hz (zero span) |
| Internal Timebase Accuracy (Standard) | ±0.7 ppm + aging |
| Internal Timebase Accuracy (with Option 010) | ±0.2 ppm + aging |
| Annual Aging Rate | ±1 ppm/year |
| Phase Noise at 1 GHz (10 kHz offset) | -106 dBc/Hz (typical -111 dBc/Hz) |
| Phase Noise at 1 GHz (100 kHz offset) | -115 dBc/Hz (typical -119 dBc/Hz) |
| DANL at 1 GHz (Preamplifier Off) | -137 dBm (typical -139 dBm) |
| DANL at 1 GHz (Preamplifier On, Option 235) | -154 dBm (typical -163 dBm) |
| DANL at 50 GHz (Preamplifier Off) | -110 dBm (typical -117 dBm) |
| DANL at 50 GHz (Preamplifier On, Option 235) | -124 dBm (typical -132 dBm) |
| Maximum Input Safe Power | +27 dBm (0.5 W) |
| Maximum Input DC Voltage | ±40 VDC |
| Input Impedance | 50 Ω (nominal) |
| Third-order Intermodulation Distortion (TOI) at 2.4 GHz | +15 dBm |
| Second Harmonic Distortion (SHI) at 1 GHz | +37 dBm (nominal) |
| LO Feedthrough at RF Input | <-80 dBm (nominal) |
| Amplitude Accuracy at 2 GHz | ±0.5 dB (typical ±0.2 dB) |
| Amplitude Accuracy at 50 GHz | ±2.1 dB (typical ±0.5 dB) |
Connectivity, Ports, & Interfaces
| Parameter | Value |
|---|---|
| RF Input Connector | 2.4 mm female (50 Ω) |
| External Reference Input Connector | BNC female (50 Ω), 10 MHz |
| External Reference Output Connector | BNC female (50 Ω), 10 MHz |
| External Trigger Input | BNC female, 10 kΩ impedance |
| GPS Antenna Port | SMA female (with Option 307) |
Measurement Capabilities & Timing
| Parameter | Value |
|---|---|
| Measurement Points | 101 to 10001 (user-selectable) |
| Trigger Types | Free Run, Video, External, RF Burst |
| RTSA Maximum Bandwidth | 10 MHz (with Option 350) |
| RTSA Minimum POI | 122 µs for full amplitude accuracy |
User Interface & Software Control
| Parameter | Value |
|---|---|
| Trace Detectors | Normal, Positive Peak, Negative Peak, Sample, Average (RMS) |
| Trace States | Clear/Write, Max Hold, Min Hold, Average, View, Blank |
| Programming Language Compatibility | SCPI |
Power & Electrical
| Parameter | Value |
|---|---|
| Power Consumption | 14 W (typical, spectrum analyzer mode) |
| Battery Operating Time | ~4 hours (typical) |
| Battery Recharge Time | 4 hours (typical, instrument off) |
Physical & Mechanical
| Parameter | Value |
|---|---|
| Form Factor | Handheld |
Environmental & Compliance
| Parameter | Value |
|---|---|
| Storage Temperature Range | -51 to +71 °C (without battery) |
| Maximum Operating Altitude | 9,144 m (30,000 ft) |
| Maximum Storage Altitude | 15,240 m (50,000 ft) |
| Operating Humidity Range | Up to 95% relative humidity (non-condensing) |
| IP Rating / Environmental Protection | IP53 / MIL-PRF-28800F Class 2 compliant |
| Calibration Interval | 1 year (recommended) |
| Safety Standards | Complies with IEC/EN 61010-1 |
| EMC/EMI Compliance | Complies with European EMC Directive 2004/108/EC, IEC/EN 61326-1 |
Configurations, Options, & Accessories
| Parameter | Value |
|---|---|
| Preamplifier Option | Option 235 (full band preamplifier support) |
| Real-time Spectrum Analyzer Option | Option 350 |
| GPS Receiver Option | Option 307 |
| Interference Analyzer Option | Option 236 |
| Standard Included Accessories | AC/DC adapter, localized power cord, rechargeable Li-ion battery, soft carrying case with backpack and shoulder straps, Quick Reference Guide |
Performance
| Parameter | Value |
|---|---|
| Sweep Time (Span ≥ 10 Hz) | 1 ms to 4000 s |
| Sweep Time (Zero Span) | 1 µs to 6000 s |
Receiver
| Parameter | Value |
|---|---|
| RF Attenuator Range | 0 to 40 dB |
| RF Attenuator Steps | 5 dB |
| Preamplifier Frequency Range | 100 kHz to 50 GHz (Option 235) |
Amplitude
| Parameter | Value |
|---|---|
| Reference Level Range | -150 dBm to +30 dBm |
| Reference Level Steps | 0.1 dB |
Frequency
| Parameter | Value |
|---|---|
| External Reference Input Level | -15 dBm to +10 dBm |
| External Reference Input Lock Range | ±1 ppm (nominal) |
| External Reference Output Level | 0 dBm ± 4 dB into 50 Ω (nominal) |
Power
| Parameter | Value |
|---|---|
| Battery Type | Lithium-ion, 10.8 V, 4.6 Ah |
| Battery Capacity | 6.2 Ah (67 Wh) |
Environmental
| Parameter | Value |
|---|---|
| Shock & Vibration | Meets MIL-PRF-28800F Class 2 |
| Transit Drop | Meets MIL-PRF-28800F Class 2 |
Options
| Parameter | Value |
|---|---|
| Analog Demodulation Option | Option 355 |
| Noise Figure Measurement Option | Option 356 |
| I/Q Analyzer Option | Option 351 |
| Built-in Power Meter Option | Option 310 |
| USB Power Sensor Support Option | Option 208 |
| Channel Scanner Option | Option 312 |
| Reflection Measurements Option | Option 320 |
| Extended Range Transmission Analysis Option | Option 209 |
RF Performance
| Parameter | Value |
|---|---|
| RF Input VSWR | 1.5 (10 MHz to 2.7 GHz), 1.6 (2.7 GHz to 7.5 GHz), 1.9 (7.5 GHz to 26.5 GHz), 2.1 (26.5 GHz to 40 GHz), 2.5 (40 GHz to 50 GHz) |
Measurement Features
| Parameter | Value |
|---|---|
| Markers | 6 markers per trace, delta markers, noise markers, band power |
Technology
RF & Microwave Spectrum Analysis
Applications
Field spectrum analysis, interference hunting, transmitter testing, real-time spectrum analysis