The Teledyne LeCroy MS-500 is a high-performance Mixed Signal Oscilloscope (MSO) module that adds 36 digital channels to compatible LeCroy oscilloscopes, including the WaveRunner Xi and WaveSurfer Xs series. Designed to capture long records of fast digital signals, it offers a maximum input frequency of 500 MHz, up to 2 GS/s sampling, and a deep memory of up to 50 Mpts/Ch. This module enables embedded system designers to view and analyze analog, digital, and serial data signals simultaneously with precise time correlation.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | L-BUS, USB 2.0 |
| Display | Uses host oscilloscope display (typically 10.4" color touch screen) |
Physical
| Parameter | Value |
|---|---|
| Power Supply | Powered via host oscilloscope L-BUS/USB port, or via external AC adapter (100 - 240 V AC) |
| Operating Temperature | 5 °C to 40 °C |
| Dimensions (W×H×D) | 213 mm × 38 mm × 211 mm |
| Weight | 0.9 kg |
Digital Input Channels
| Parameter | Value |
|---|---|
| Number of Digital Channels | 36 |
| Input Impedance (Resistance) | 100 kΩ |
| Input Capacitance | 5 pF |
| Maximum Input Voltage | ±30 V non-destruct |
| Threshold Grouping | 4 groups (9 channels per group) |
| User-Defined Threshold Range | -10 V to +10 V |
| Threshold Accuracy | ±(100 mV + 3% of threshold setting) |
| Minimum Input Voltage Swing | 500 mV p-p |
Performance
| Parameter | Value |
|---|---|
| Channel-to-Channel Skew | < 1 ns (typical) |
| Digital-to-Analog Time Correlation Skew | < 2 ns |
| Maximum Input Frequency | 250 MHz |
Acquisition & Memory
| Parameter | Value |
|---|---|
| Maximum Sampling Rate | 2 GS/s (on 18 channels), 1 GS/s (on 36 channels) |
| Maximum Memory Depth | 50 Mpts/Ch (on 18 channels), 25 Mpts/Ch (on 36 channels) |
| Minimum Detectable Pulse Width | 2 ns |
Triggering
| Parameter | Value |
|---|---|
| Trigger Sources | Any digital or analog channel |
| Trigger Types | Edge, Pattern, State, Pulse Width, Glitch, Setup and Hold |
| Edge Triggering Modes | Rising edge, falling edge, or either edge |
| Pattern Triggering Setup | Pattern of HIGH, LOW, or DON'T CARE on 36 channels |
| State Triggering Setup | Pattern clocked by edge on another channel |
| Pulse Width Trigger Range | 2 ns to 20 s |
| Glitch Trigger Minimum | 2 ns |
| Setup and Hold Triggering | Setup/hold time violations on 36 channels relative to a clock edge |
| Analog/Digital Crosspattern Trigger | Yes |
Integration & Software
| Parameter | Value |
|---|---|
| Bus Decoding Protocol Support | I2C, SPI, UART, RS-232, CAN, LIN (software-dependent) |
| Mixed Signal Display Modes | Bus display (hex, binary, octal, decimal), individual digital lines |
| Search Engine Integration | WaveScan Search and Find |
| Compatible Oscilloscope Series | WaveRunner Xi/Xi-A, WaveSurfer Xs/Xs-A, WaveRunner 6000A, WavePro 7 Zi/Zi-A, HRO 12-bit |
Measurements
| Parameter | Value |
|---|---|
| Timing Measurements | single channel, digital-to-digital, or analog-to-digital |
| Cursor Readout | Hexadecimal, binary, octal, decimal bus values |
Accessories
| Parameter | Value |
|---|---|
| Lead Set Cable Length | 1.0 m |
| Micro-gripper Quantity Included | 40 |
Environmental
| Parameter | Value |
|---|---|
| Storage Temperature Range | -20 °C to +60 °C |
| Operating Humidity | 5% to 80% RH (non-condensing) at 30 °C |
| Storage Humidity | 5% to 95% RH |
| Operating Altitude | Up to 3,000 m (10,000 ft) |
| Storage Altitude | Up to 12,192 m (40,000 ft) |
Compliance
| Parameter | Value |
|---|---|
| Safety Certifications | EN 61010-1 |
| EMC Compliance | EN 61326-1 |
| RoHS Compliance | Yes |
| WEEE Directive Compliance | Yes |
Thresholds
| Parameter | Value |
|---|---|
| Logic Threshold Presets | TTL, ECL, CMOS (2.5 V, 3.3 V, 5.0 V), PECL, LVDS |
| Threshold Resolution | 10 mV |
Digital Inputs
| Parameter | Value |
|---|---|
| Digital Input Bandwidth | 500 MHz |
| Digital Channel Grouping | 4 banks of 9 channels |
| Input Dynamic Range | -10 V to +10 V |
Connectivity
| Parameter | Value |
|---|---|
| Host Interface | Proprietary LBUS |
| LBUS Cable Length | 1.0 m |
Power Requirements
| Parameter | Value |
|---|---|
| Power Source | Powered directly from host oscilloscope via LBUS interface |
Probing
| Parameter | Value |
|---|---|
| Digital Lead Connector | Standard 0.1-inch (2.54 mm) square-pin socket |
Technology
Mixed Signal Oscilloscope (MSO) Module
Applications
Debugging and validation of embedded systems, microcontrollers, DSPs, FPGAs, ADCs, and DACs