WaveRunner 64MXi-A

LeCroy (Teledyne LeCroy)DSO

The Teledyne LeCroy WaveRunner 64MXi-A is a 600 MHz, 4-channel Digital Storage Oscilloscope (DSO) featuring high-performance specifications, active analysis tools, and a large display in a space-saving 15 cm deep form factor. Introduced in February 2009 as an update to the original 64MXi, the 'MXi-A' designation represents a standard package that includes upgraded deep acquisition memory (12.5 Mpts/ch), advanced math capabilities, customized VBScript/MATLAB/Excel scripting, and serial trigger options, providing deep troubleshooting capability for mixed-signal design, serial data, and power applications.

Specifications

General

ParameterValue
InterfacesUSB Host (6 ports: 2 front, 4 rear), USB Device (1 port), LXI-compliant Gigabit LAN (10/100/1000Base-T), External VGA (DB-15), GPIB (optional)
Display10.4" color TFT-LCD touchscreen, SVGA (800 x 600 pixels) with integrated stylus
Calibration Interval1 year
Standard Probes4 × PP009 (500 MHz, 10:1, 10 MΩ) passive probes

Physical

ParameterValue
Power Supply100 - 240 VAC (±10%) at 50/60 Hz (±10%), 100 - 120 VAC at 400 Hz; max power consumption 340 VA
Operating Temperature5 °C to 40 °C
Dimensions (W×H×D)397 mm × 263 mm × 152 mm
Weight7.26 kg

Vertical System (Analog Channels)

ParameterValue
Analog Bandwidth (-3 dB)600 MHz
Calculated Rise Time (10% to 90%)583 ps (typical)
Number of Input Channels4
Input CouplingAC, DC, GND (GND, 50 Ω, 1 MΩ)
Input Impedance1 MΩ ±1% || 16 pF, or 50 Ω ±1%
Vertical Resolution (A/D Converter)8-bit (up to 11-bit with Enhanced Resolution/ERES)
DC Gain Accuracy±1.5% of full scale
Vertical Sensitivity Range (50 Ω)2 mV/div - 1 V/div
Vertical Sensitivity Range (1 MΩ)2 mV/div - 10 V/div
Maximum Input Voltage (50 Ω)5 Vrms
Maximum Input Voltage (1 MΩ)400 V max (DC + Peak AC ≤ 10 kHz)
Hardware Bandwidth Limiters20 MHz, 200 MHz

Horizontal System & Acquisition

ParameterValue
Maximum Real-time Sample Rate (Single-Channel)10 GS/s (interleaved)
Maximum Real-time Sample Rate (Multi-Channel)5 GS/s (all 4 channels)
Standard Record Length (Memory Depth)12.5 Mpts/ch (all 4 channels), 25 Mpts/ch (interleaved 2 channels)
Time Base Range200 ps/div - 1000 s/div
Time Base Accuracy±10 ppm

Trigger System

ParameterValue
Trigger SourcesChannels 1-4, External, Line, Ext/10
Trigger ModesAuto, Normal, Single, Stop
Trigger CouplingAC, DC, HFRej, LFRej
Standard Trigger TypesEdge, Width, Glitch, Period, Timeout, Slew Rate, Interval, Dropout, Pattern, State, Window, TV/Video
Advanced and Protocol Trigger TypesI2C, SPI, UART, RS-232, LIN, CAN, FlexRay (some capability optional)
External Trigger Input Impedance1 MΩ || 16 pF
External Trigger Max Input Voltage (1 MΩ)400 V max (DC + Peak AC ≤ 10 kHz)
Trigger Level Range±4 divisions from center (internal)
External Trigger Input Range±400 mV (Ext), ±4 V (Ext/10)
Trigger Jitter< 2.5 ps rms (typical)

Acquisition Modes & Signal Processing

ParameterValue
Acquisition ModesReal-time, Roll, Random Repetitive (RIS), WaveStream (Fast Viewing)
Waveform Math OperatorsDual operators; Sum, Difference, Product, Ratio, FFT, Absolute Value, Derivative, Integral, Square, Square Root, Reciprocal, Rescale, and more
Advanced Waveform FunctionsCompletely customizable math and measurements using Excel, MATLAB, or VBScripts

Measurements & Analysis

ParameterValue
Automatic Voltage MeasurementsAmplitude, Base, Maximum, Minimum, Mean, Peak-to-Peak, RMS, Top, Custom
Automatic Time & Frequency MeasurementsDelay, Duty Cycle, Fall Time, Frequency, Period, Phase, Rise Time, Width, Custom
Search EngineWaveScan Advanced Search and Analysis

Display & User Interface

ParameterValue
User Interface ControlDedicated Front Panel Cursor Knobs, Front Panel Zoom Controls

Environmental & Compliance

ParameterValue
Storage Temperature Range-20 °C to +60 °C
Operating Humidity Range5% to 80% RH (non-condensing) at ≤ 30 °C
Operating AltitudeUp to 3,048 m (10,000 ft) at ≤ 25 °C

PC Platform

ParameterValue
ProcessorIntel Core 2 Duo (3.0 GHz or better)
System Memory2 GB standard
Storage160 GB or larger removable hard drive

Vertical System

ParameterValue
Probe InterfaceTeledyne LeCroy ProBus
Channel-to-Channel Isolation> 30 dB at rated bandwidth (typical)

Acquisition System

ParameterValue
Random Interleaved Sampling (RIS) Rate200 GS/s
Enhanced Resolution (ERES)Up to 11 bits (with standard software)

Auxiliary Outputs

ParameterValue
Calibrator Signal Output1 kHz, 1 V square wave into ≥ 100 kΩ

Power Requirements

ParameterValue
Power Consumption (Typical)340 VA
Power Consumption (Maximum)440 VA

Interfaces

ParameterValue
LXI ComplianceLXI Class C

Compliance

ParameterValue
Environmental Safety ComplianceConforms to EN 61010-1, EN 61326-1

Technology

DSO (Digital Storage Oscilloscope)

Applications

Electronic Design, Embedded Systems Validation, Jitter/Timing Analysis, Serial Bus Debugging, Power Diagnostics

Looking for the LeCroy (Teledyne LeCroy) WaveRunner 64MXi-A?