The QuadTech 7600 is a high-precision LCR meter belonging to the legacy QuadTech 7000 Series. Designed for rigorous component and materials testing, it covers a wide frequency range from 10 Hz to 2 MHz with 289,900 programmable frequencies. The instrument delivers a basic measurement accuracy of 0.05% and can display any two of its 14 measured impedance parameters simultaneously. With standard RS-232, IEEE-488 (GPIB), parallel printer, and PLC handler interfaces, the QuadTech 7600 is highly versatile for automated test environments, QA inspections, and laboratory calibration.

Specifications

General

ParameterValue
InterfacesRS-232, IEEE-488 (GPIB), Parallel Printer, and PLC Handler
DisplayHigh-resolution LCD with 7-digit display

Physical

ParameterValue
Power Supply90 - 250 V AC, 47 - 63 Hz
Operating Temperature0°C to +50°C
Dimensions (W×H×D)432 mm x 178 mm x 356 mm
Weight6.8 kg

Measurement Parameters & Functions

ParameterValue
Primary & Secondary ParametersCs, Cp, Ls, Lp, Rs, Rp, |Z|, |Y|, G, B, X, D, Q, θ (14 parameters total, 2 displayed simultaneously)
Parameter Presentation FormatsDirect Value, Deviation, or % Deviation from a nominal value
Sequence Test ModeAbility to automatically perform 6 unique tests in sequence
Equivalent Circuit ModesSeries, Parallel

Measurement Speed & Accuracy

ParameterValue
Basic Measurement Accuracy±0.05%
Maximum Measurement SpeedUp to 40 measurements/second
Programmable Averaging & DelayProgrammable average and delay times

Test Signal Characteristics

ParameterValue
Test Frequency Range10 Hz to 2 MHz
Number of Programmable Frequencies289,900
AC Test Voltage Range20 mV to 5 V (programmable)
Level ControlConstant voltage maintained to the Device Under Test (DUT)

DC Bias Source

ParameterValue
Internal DC Bias Voltage2 V
External DC Bias Voltage Range0 V to 200 V

Calibration & Compensation

ParameterValue
Zeroing / CompensationAutomatic Open/Short Zeroing
Calibration ReferenceOn-site calibration with NIST traceable calibration kit

Display & User Interface

ParameterValue
Display Resolution7 digits
Setup StorageInternal storage and recall of test setups
Security LockoutKeypad Lockout prevents changes to the test setup

Sweep & Graphic Analysis

ParameterValue
Sorting Bins14 Hi/Lo sorting bins for Pass/Fail testing

Hardware Interfaces & Connectivity

ParameterValue
Test Port ConfigurationFour-Terminal Pair front panel BNC configuration

Test Signal

ParameterValue
AC Test Current Range250 µA to 100 mA (frequency dependent)

Environmental

ParameterValue
Storage Temperature-40°C to +71°C
Operating Humidity< 85% RH (non-condensing)

Measurement Control

ParameterValue
Trigger SourceContinuous (Internal), Manual, External (BNC or Interface)

Measurement Presentation

ParameterValue
Graphical Sweep ParametersFrequency, Voltage, Current

Measurement Ranges

ParameterValue
Capacitance Measurement Range0.00001 pF to 9.9999 F
Inductance Measurement Range0.00001 µH to 99.999 kH
Resistance / Impedance Measurement Range0.00001 Ω to 99.999 MΩ
Admittance Measurement Range0.00001 µS to 9.9999 kS
Dissipation Factor Range (D)0.00001 to 9.9999
Quality Factor Range (Q)0.00001 to 9999.9
Phase Angle Range (θ)-180.00° to +180.00°

Performance

ParameterValue
Frequency Accuracy±0.01%
AC Voltage Resolution5 mV (for 20 mV to 1 V), 20 mV (for 1 V to 5 V)
Trigger Delay Range0 to 9999 ms

Technology

LCR Bridge / Impedance Analyzer

Applications

Precision electronic component characterization, impedance measurement, and automated quality control testing