- Home
- Categories
- Test equipment
- Testers
- Hipots
- Guardian 6000
The QuadTech Guardian 6000 is an advanced, multi-function electrical safety analyzer designed for comprehensive safety testing, including AC/DC hipot, insulation resistance, and ground bond tests.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | RS-232, Remote I/O (PLC), GPIB (optional) |
| Display | 320 x 240 backlit LCD |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 115 / 230 V AC ±10% (switchable), 50/60 Hz |
| Operating Temperature | 0 to 40 |
| Dimensions (W×H×D) | 430 × 132 × 450 mm |
| Weight | 15 kg |
| Rack Mounting | Optional 19-inch rack mount kit |
AC Dielectric Withstand (AC Hipot)
| Parameter | Value |
|---|---|
| Output Voltage Range | 0.050 kV to 5.000 kV AC |
| Voltage Resolution | 1 V |
| Voltage Accuracy | ±(1% of setting + 5 V) |
| Output Frequency Range | 50 Hz / 60 Hz (selectable) |
| Output Waveform | Sine wave |
| Max Rated Current | 40 mA AC |
| Current Measurement Range | 0.001 mA to 40.00 mA AC |
| Current Resolution | 1 µA |
| Current Accuracy | ±(1% of reading + 5 µA) |
| Arc Detection Threshold Range | 1 mA to 20 mA |
DC Dielectric Withstand (DC Hipot)
| Parameter | Value |
|---|---|
| Auto-Discharge Device Function | Yes |
Insulation Resistance (IR)
| Parameter | Value |
|---|---|
| Resistance Measurement Range | 0.1 MΩ to 50 GΩ |
Ground Bond / Continuity (GB)
| Parameter | Value |
|---|---|
| Test Current Range | 3.0 A to 30.0 A AC |
| Test Frequency | 50 Hz / 60 Hz (selectable) |
| Resistance Accuracy | ±(1% of reading + 3 mΩ) |
| Voltage Limit Range | Open circuit voltage < 15 V AC |
Test Control & Timing
| Parameter | Value |
|---|---|
| Ramp-Up Time Range | 0.1 s to 999 s |
| Dwell Time (Test Time) Range | 0.1 s to 999 s (or continuous) |
| Ramp-Down Time Range | 0.1 s to 999 s |
| Timing Resolution | 0.1 s |
| Timing Accuracy | ±(0.1% + 0.05 s) |
Safety & Protection Features
| Parameter | Value |
|---|---|
| Ground Fault Interrupt (GFI) Trip Current | 0.5 mA threshold |
| Safety Interlock Type | Hardware interlock via rear panel terminal |
| Fast Cut-Off Response Time | < 0.4 ms |
| Visual Fail/Pass Indicators | Green (Pass) and Red (Fail) LEDs |
| Audible Alarm Output | Yes, volume adjustable |
System, Memory & Programming
| Parameter | Value |
|---|---|
| Step Link / Auto-Sequence Capability | Yes |
| Front Panel Controls | Tactile numeric keypad and function keys |
| Key Lock / Security Password Protection | Yes |
Connectivity & Interfaces
| Parameter | Value |
|---|---|
| Barcode Scanner Port | Supports RS-232 barcode scanner input |
Measurement Features
| Parameter | Value |
|---|---|
| Open Short Check (OSC) | Yes |
| High Voltage Contact Check (HFCC) | High Frequency Contact Check (HFCC) |
Leakage Current Mode
| Parameter | Value |
|---|---|
| Leakage Current Test Types | Earth, Enclosure, Patient, Patient Auxiliary |
| Built-in Measuring Devices (MD) | 8 selectable networks representing human body impedance per medical/general standards |
AC Hipot Test
| Parameter | Value |
|---|---|
| AC Hipot Voltage Range | 50 V - 5000 V AC |
| AC Hipot Power Rating | 200 VA |
DC Hipot Test
| Parameter | Value |
|---|---|
| DC Hipot Voltage Range | 50 V - 6000 V DC |
| DC Hipot Power Rating | 60 W |
Insulation Resistance Test
| Parameter | Value |
|---|---|
| Insulation Resistance Voltage Range | 50 V - 1000 V DC |
| Insulation Resistance Measurement Range | 100 kΩ - 50 GΩ |
Ground Bond Test
| Parameter | Value |
|---|---|
| Ground Bond Test Current Range | 3.0 A - 30.0 A AC |
| Ground Bond Resistance Range | 0.1 mΩ - 510 mΩ |
Functional Run Test
| Parameter | Value |
|---|---|
| Functional Run Test Voltage Range | 0 V - 277 V AC |
| Functional Run Test Current Range | 0.1 A - 15.0 A AC |
Line Leakage Test
| Parameter | Value |
|---|---|
| Line Leakage Current Range | 0.1 µA - 10.0 mA |
Interfaces & Control
| Parameter | Value |
|---|---|
| Remote Control Signals | Start, Stop, Interlock, Pass, Fail, Under Test |
Calibration & Maintenance
| Parameter | Value |
|---|---|
| Calibration Traceability | NIST traceable |
Technology
Hipot Testing & Ground Bond Analysis
Applications
Electrical Safety Compliance Testing
Looking for the QuadTech Guardian 6000?
Tell us what you need. We're building a network of equipment dealers and will get back to you if we can help source it — no obligation.
Selling this kind of equipment? Sell with us