The Sorensen XPH Series by Ametek consists of compact, highly efficient benchtop DC power supplies offering 175W to 420W of output power. Designed with a hybrid mixed-mode regulation system that combines switch-mode pre-regulation with linear post-regulation, the XPH Series delivers performance comparable to pure linear designs, including excellent line and load regulation, low output noise, and fast transient response. Available in single, dual, and triple output configurations, these economical power supplies are highly suited for engineering benches, educational laboratories, and system integration.
Specifications
General
| Parameter | Value |
|---|---|
| Display | Dual 3-digit LED displays per output (voltage and current) |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 115V AC / 230V AC ± 10% (switchable), 50Hz / 60Hz |
| Operating Temperature | 5°C to 40°C |
| Dimensions (W×H×D) | 140 x 160 x 320 mm (175/180W models), 210 x 160 x 395 mm (420W model), 260 x 160 x 320 mm (Dual/Triple models) |
| Weight | 4.4 - 10.2 kg (model dependent) |
| Output Terminals | 4 mm safety binding posts on 19 mm (0.75 in) spacing |
Output Ratings
| Parameter | Value |
|---|---|
| XPH 35-5 Voltage Range | 0 to 35 V |
| XPH 35-5 Current Range | 0 to 5 A |
| XPH 35-5 Maximum Output Power | 175 W |
| XPH 18-10 Voltage Range | 0 to 18 V |
| XPH 18-10 Current Range | 0 to 10 A |
| XPH 18-10 Maximum Output Power | 180 W |
| XPH 35-4D Output 1 & 2 Voltage Range | 0 to 35 V |
| XPH 35-4D Output 1 & 2 Current Range | 0 to 4 A |
| XPH 35-4D Maximum Output Power | 280 W |
| XPH 35-4T Output 1 & 2 Voltage Range | 0 to 35 V |
| XPH 35-4T Output 1 & 2 Current Range | 0 to 4 A |
| XPH 35-4T Auxiliary Output Voltage | 3.3 V or 5.0 V switchable |
| XPH 35-4T Auxiliary Output Current | 5 A max |
| XPH 35-4T Maximum Output Power | 305 W |
| XPH 42-10 Voltage Range | 0 to 42 V |
| XPH 42-10 Current Range | 0 to 10 A |
| XPH 42-10 Maximum Output Power | 420 W |
| XPH 75-2D Output 1 & 2 Voltage Range | 0 to 75 V |
| XPH 75-2D Output 1 & 2 Current Range | 0 to 2 A |
| XPH 75-2D Maximum Output Power | 300 W |
| Number of Channels | Single (XPH 35-5, 18-10, 42-10), Dual (XPH 35-4D, 75-2D), Triple (XPH 35-4T) |
| Output Configuration Modes | Independent, Series, or Parallel (Dual/Triple models) |
Line & Load Regulation
| Parameter | Value |
|---|---|
| Load Regulation - Voltage | <0.01% + 2mV (using remote sense on applicable models) |
| Load Regulation - Current | <0.01% + 1mA |
| Line Regulation - Voltage | <0.01% + 2mV for a 10% line change |
| Line Regulation - Current | <0.01% + 1mA for a 10% line change |
Ripple & Noise
| Parameter | Value |
|---|---|
| Ripple & Noise - Voltage (RMS) | <1mV RMS |
| Ripple & Noise - Current (RMS) | <1mA RMS (<2mA RMS on 10A models) |
Programming & Readback Resolution
| Parameter | Value |
|---|---|
| Meter Resolution (XPH 35-5) | 100mV, 10mA |
| Meter Resolution (XPH 18-10) | 100mV, 100mA |
| Meter Resolution (XPH 35-4D) | 100mV, 10mA |
| Meter Resolution (XPH 35-4T) | 100mV, 10mA |
| Meter Resolution (XPH 42-10) | 100mV, 100mA |
| Meter Resolution (XPH 75-2D) | 100mV, 10mA |
Programming & Readback Accuracy
| Parameter | Value |
|---|---|
| Meter Accuracy - Voltage | ±(0.3% of reading + 1 digit) |
| Meter Accuracy - Current | ±(0.6% of reading + 1 digit) |
Transient Response
| Parameter | Value |
|---|---|
| Transient Recovery Time | <250µs to within 50mV of setting after a 50% to 100% load change |
User Interface & Controls
| Parameter | Value |
|---|---|
| Operating Modes | Constant Voltage (CV) or Constant Current (CC) with automatic crossover |
| Mode Indication | Individual LED indicators for CV and CC modes |
| Voltage Control Panel Controls | Coarse and fine rotary analog controls |
| Current Control Panel Controls | Single rotary analog control |
| Output ON/OFF Switches | Individual switch per output (excluding auxiliary output on XPH 35-4T) |
| Remote Sense Terminals | Available on XPH 18-10 and XPH 42-10 (switchable on front panel) |
Environmental Specifications
| Parameter | Value |
|---|---|
| Relative Humidity (Operating) | 20% to 80% RH (non-condensing) |
| Storage Temperature Range | -20°C to 60°C |
| Relative Humidity (Storage) | 10% to 90% RH (non-condensing) |
| Cooling System (Power levels < 420W) | Silent convection cooling (fanless) |
| Cooling System (XPH 42-10) | Low-noise temperature-controlled fan cooling |
Physical Characteristics
| Parameter | Value |
|---|---|
| XPH 35-5 Weight | 3.0 kg |
| XPH 18-10 Weight | 3.0 kg |
| XPH 35-4D Weight | 4.3 kg |
| XPH 35-4T Weight | 4.3 kg |
| XPH 75-2D Weight | 4.3 kg |
| XPH 42-10 Weight | 5.0 kg |
| XPH 35-5 Dimensions (W x H x D) | 140 x 160 x 320 mm |
| XPH 18-10 Dimensions (W x H x D) | 140 x 160 x 320 mm |
| XPH 35-4D Dimensions (W x H x D) | 260 x 160 x 320 mm |
| XPH 35-4T Dimensions (W x H x D) | 260 x 160 x 320 mm |
| XPH 75-2D Dimensions (W x H x D) | 260 x 160 x 320 mm |
| XPH 42-10 Dimensions (W x H x D) | 210 x 160 x 395 mm |
Power Requirements
| Parameter | Value |
|---|---|
| AC Input Voltage | 115 VAC or 230 VAC ±10% (switch selectable) |
| AC Input Frequency | 47 Hz to 63 Hz |
| Power Consumption (XPH 35-5) | 320 VA max |
| Power Consumption (XPH 18-10) | 320 VA max |
| Power Consumption (XPH 35-4D) | 640 VA max |
| Power Consumption (XPH 35-4T) | 650 VA max |
| Power Consumption (XPH 42-10) | 800 VA max |
| Power Consumption (XPH 75-2D) | 640 VA max |
Environmental
| Parameter | Value |
|---|---|
| Operating Altitude | Up to 2000 m |
| Pollution Degree | 2 |
Regulatory
| Parameter | Value |
|---|---|
| Safety Compliance | IEC 61010-1 |
| EMC Compliance | EN 61326 |
Output Specifications
| Parameter | Value |
|---|---|
| Maximum Output Floating Voltage | 300 VDC |
| XPH 35-4D & XPH 35-4T Series Mode Output | Up to 70 V at 4 A |
| XPH 35-4D & XPH 35-4T Parallel Mode Output | Up to 35 V at 8 A |
| XPH 75-2D Series Mode Output | Up to 150 V at 2 A |
| XPH 75-2D Parallel Mode Output | Up to 75 V at 4 A |
Output Protection
| Parameter | Value |
|---|---|
| Reverse Diode Protection | Diode clamped, maximum current 3A |
Performance
| Parameter | Value |
|---|---|
| Voltage Temperature Coefficient | Typically < 100 ppm/°C |
| Current Temperature Coefficient | Typically < 150 ppm/°C |
| Output Impedance (CV Mode) | Typically < 5 mΩ at 1 kHz |
Technology
Mixed-mode regulation (switch-mode pre-regulation and linear post-regulation)
Applications
Benchtop testing, laboratory development, education, and system maintenance