XPH Series - 175 to 420W

orensen (Ametek)Sorensen

The Sorensen XPH Series by Ametek consists of compact, highly efficient benchtop DC power supplies offering 175W to 420W of output power. Designed with a hybrid mixed-mode regulation system that combines switch-mode pre-regulation with linear post-regulation, the XPH Series delivers performance comparable to pure linear designs, including excellent line and load regulation, low output noise, and fast transient response. Available in single, dual, and triple output configurations, these economical power supplies are highly suited for engineering benches, educational laboratories, and system integration.

Specifications

General

ParameterValue
DisplayDual 3-digit LED displays per output (voltage and current)

Physical

ParameterValue
Power Supply115V AC / 230V AC ± 10% (switchable), 50Hz / 60Hz
Operating Temperature5°C to 40°C
Dimensions (W×H×D)140 x 160 x 320 mm (175/180W models), 210 x 160 x 395 mm (420W model), 260 x 160 x 320 mm (Dual/Triple models)
Weight4.4 - 10.2 kg (model dependent)
Output Terminals4 mm safety binding posts on 19 mm (0.75 in) spacing

Output Ratings

ParameterValue
XPH 35-5 Voltage Range0 to 35 V
XPH 35-5 Current Range0 to 5 A
XPH 35-5 Maximum Output Power175 W
XPH 18-10 Voltage Range0 to 18 V
XPH 18-10 Current Range0 to 10 A
XPH 18-10 Maximum Output Power180 W
XPH 35-4D Output 1 & 2 Voltage Range0 to 35 V
XPH 35-4D Output 1 & 2 Current Range0 to 4 A
XPH 35-4D Maximum Output Power280 W
XPH 35-4T Output 1 & 2 Voltage Range0 to 35 V
XPH 35-4T Output 1 & 2 Current Range0 to 4 A
XPH 35-4T Auxiliary Output Voltage3.3 V or 5.0 V switchable
XPH 35-4T Auxiliary Output Current5 A max
XPH 35-4T Maximum Output Power305 W
XPH 42-10 Voltage Range0 to 42 V
XPH 42-10 Current Range0 to 10 A
XPH 42-10 Maximum Output Power420 W
XPH 75-2D Output 1 & 2 Voltage Range0 to 75 V
XPH 75-2D Output 1 & 2 Current Range0 to 2 A
XPH 75-2D Maximum Output Power300 W
Number of ChannelsSingle (XPH 35-5, 18-10, 42-10), Dual (XPH 35-4D, 75-2D), Triple (XPH 35-4T)
Output Configuration ModesIndependent, Series, or Parallel (Dual/Triple models)

Line & Load Regulation

ParameterValue
Load Regulation - Voltage<0.01% + 2mV (using remote sense on applicable models)
Load Regulation - Current<0.01% + 1mA
Line Regulation - Voltage<0.01% + 2mV for a 10% line change
Line Regulation - Current<0.01% + 1mA for a 10% line change

Ripple & Noise

ParameterValue
Ripple & Noise - Voltage (RMS)<1mV RMS
Ripple & Noise - Current (RMS)<1mA RMS (<2mA RMS on 10A models)

Programming & Readback Resolution

ParameterValue
Meter Resolution (XPH 35-5)100mV, 10mA
Meter Resolution (XPH 18-10)100mV, 100mA
Meter Resolution (XPH 35-4D)100mV, 10mA
Meter Resolution (XPH 35-4T)100mV, 10mA
Meter Resolution (XPH 42-10)100mV, 100mA
Meter Resolution (XPH 75-2D)100mV, 10mA

Programming & Readback Accuracy

ParameterValue
Meter Accuracy - Voltage±(0.3% of reading + 1 digit)
Meter Accuracy - Current±(0.6% of reading + 1 digit)

Transient Response

ParameterValue
Transient Recovery Time<250µs to within 50mV of setting after a 50% to 100% load change

User Interface & Controls

ParameterValue
Operating ModesConstant Voltage (CV) or Constant Current (CC) with automatic crossover
Mode IndicationIndividual LED indicators for CV and CC modes
Voltage Control Panel ControlsCoarse and fine rotary analog controls
Current Control Panel ControlsSingle rotary analog control
Output ON/OFF SwitchesIndividual switch per output (excluding auxiliary output on XPH 35-4T)
Remote Sense TerminalsAvailable on XPH 18-10 and XPH 42-10 (switchable on front panel)

Environmental Specifications

ParameterValue
Relative Humidity (Operating)20% to 80% RH (non-condensing)
Storage Temperature Range-20°C to 60°C
Relative Humidity (Storage)10% to 90% RH (non-condensing)
Cooling System (Power levels < 420W)Silent convection cooling (fanless)
Cooling System (XPH 42-10)Low-noise temperature-controlled fan cooling

Physical Characteristics

ParameterValue
XPH 35-5 Weight3.0 kg
XPH 18-10 Weight3.0 kg
XPH 35-4D Weight4.3 kg
XPH 35-4T Weight4.3 kg
XPH 75-2D Weight4.3 kg
XPH 42-10 Weight5.0 kg
XPH 35-5 Dimensions (W x H x D)140 x 160 x 320 mm
XPH 18-10 Dimensions (W x H x D)140 x 160 x 320 mm
XPH 35-4D Dimensions (W x H x D)260 x 160 x 320 mm
XPH 35-4T Dimensions (W x H x D)260 x 160 x 320 mm
XPH 75-2D Dimensions (W x H x D)260 x 160 x 320 mm
XPH 42-10 Dimensions (W x H x D)210 x 160 x 395 mm

Power Requirements

ParameterValue
AC Input Voltage115 VAC or 230 VAC ±10% (switch selectable)
AC Input Frequency47 Hz to 63 Hz
Power Consumption (XPH 35-5)320 VA max
Power Consumption (XPH 18-10)320 VA max
Power Consumption (XPH 35-4D)640 VA max
Power Consumption (XPH 35-4T)650 VA max
Power Consumption (XPH 42-10)800 VA max
Power Consumption (XPH 75-2D)640 VA max

Environmental

ParameterValue
Operating AltitudeUp to 2000 m
Pollution Degree2

Regulatory

ParameterValue
Safety ComplianceIEC 61010-1
EMC ComplianceEN 61326

Output Specifications

ParameterValue
Maximum Output Floating Voltage300 VDC
XPH 35-4D & XPH 35-4T Series Mode OutputUp to 70 V at 4 A
XPH 35-4D & XPH 35-4T Parallel Mode OutputUp to 35 V at 8 A
XPH 75-2D Series Mode OutputUp to 150 V at 2 A
XPH 75-2D Parallel Mode OutputUp to 75 V at 4 A

Output Protection

ParameterValue
Reverse Diode ProtectionDiode clamped, maximum current 3A

Performance

ParameterValue
Voltage Temperature CoefficientTypically < 100 ppm/°C
Current Temperature CoefficientTypically < 150 ppm/°C
Output Impedance (CV Mode)Typically < 5 mΩ at 1 kHz

Technology

Mixed-mode regulation (switch-mode pre-regulation and linear post-regulation)

Applications

Benchtop testing, laboratory development, education, and system maintenance