The Stanford Research Systems SR620 is a universal time interval and frequency counter offering 25 ps single-shot timing resolution and frequency measurements up to 1.3 GHz. Designed for laboratory and ATE environments, it features built-in statistical analysis, Allan variance calculations, and live graphical outputs to external oscilloscopes for histograms and strip charts.

Specifications

General

ParameterValue
InterfacesGPIB, RS-232
Display16-digit alphanumeric LED

Physical

ParameterValue
Power Supply100 / 120 / 220 / 240 VAC, 50/60 Hz, 50 W
Operating Temperature0 to +50
Dimensions (W×H×D)356 x 89 x 356
Weight4.5

Measurement Modes

ParameterValue
Available MeasurementsTime interval, frequency, pulse-width, rise and fall time, period, phase, events

Time Interval Specifications

ParameterValue
Single-Shot Resolution25 ps rms

Frequency and Period Specifications

ParameterValue
Frequency Range0.001 Hz to 1.3 GHz
Frequency ResolutionUp to 11 digits (1 s gate)

Phase Specifications

ParameterValue
Phase Resolution0.001°

Timebase / Reference Oscillator

ParameterValue
Standard Oscillator Aging Rate1 × 10^-6 / year
Optional Ovenized Oscillator Aging Rate5 × 10^-10 / day

Statistical and Math Functions

ParameterValue
Sample Size Range1 to 1,000,000
Calculated StatisticsMean, standard deviation, Allan variance, minimum, maximum
Histogram GenerationUp to 250 bins
Strip Chart DisplayMean or jitter (standard deviation / Allan variance), up to 250 points

Display Characteristics

ParameterValue
Graphical OutputVia X-Y scope inputs or HP-GL compatible plotter/printer

Auxiliary Outputs

ParameterValue
Internal Reference Output1 kHz, 50% duty cycle square wave

Auxiliary Inputs

ParameterValue
DVM Inputs2 rear-panel channels
DVM Input Voltage Range±20 VDC
DVM Input Accuracy0.3%

Calibration

ParameterValue
Auto-Calibration Duration~2 minutes

Measurement Ranges

ParameterValue
Time Interval Range-1000 s to +1000 s
Phase Measurement Range-180° to +180°

Inputs A & B

ParameterValue
Input Impedance (Inputs A & B)50 Ω or 1 MΩ
Input Coupling (Inputs A & B)AC or DC
Trigger Level Range±5.00 V
Trigger Level Resolution10 mV
Input Bandwidth300 MHz
Input Sensitivity50 mVrms
Typical Input Noise350 µVrms

Reference / Timebase

ParameterValue
Timebase Frequency10.000 MHz

Outputs

ParameterValue
Internal Reference Output Level1 Vpp into 50 Ω
Analog Outputs (D/A) ChannelsTwo rear-panel outputs (X-axis, Y-axis)

DVM Inputs

ParameterValue
DVM Input Impedance1 MΩ

Performance

ParameterValue
Internal Time Jitter25 ps rms (typical)

Inputs & Outputs

ParameterValue
External Arming InputTTL level

Power

ParameterValue
Power Consumption50 W

Applications

Laboratory, ATE, Atomic clock drift measurement, PLL jitter measurement