The Tektronix A621 is a passive, high-current AC current probe designed for use with oscilloscopes and digital multimeters. It measures AC currents from 100 mA to 2000 A peak over a frequency range of 5 Hz to 50 kHz. Featuring a large clamp-on jaw, it accommodates conductors up to 54 mm in diameter, making it ideal for industrial and power utility applications.

Specifications

General

ParameterValue
InterfacesBNC (integrated cable)

Physical

ParameterValue
Power SupplyPassive (no battery or external power required)
Operating Temperature-10 °C to +55 °C
Dimensions (W×H×D)97 mm × 231 mm × 36 mm
Weight0.65 kg
Maximum Busbar Size50 mm x 5 mm (2.0 in. x 0.2 in.)

Electrical & Measurement Specifications

ParameterValue
Frequency Range5 Hz to 50 kHz (-3 dB)
Maximum Peak AC Current2000 A
Maximum Continuous AC Current1000 A RMS
AC Current Range100 mA to 2000 A peak
Output Sensitivity Settings1 mV/A, 10 mV/A, 100 mV/A

Safety & Isolation Ratings

ParameterValue
Maximum Working Voltage600 V RMS
Measurement CategoryCAT III 600 V

Physical & Mechanical Specifications

ParameterValue
Maximum Conductor Diameter (Jaw Opening)54 mm (2.13 in.)
Cable Length1.5 m (5 ft.)
Output Connector TypeBNC

Power Source & Management

ParameterValue
Power RequirementsPassive (none)

Compatibility & Ordering

ParameterValue
Recommended Instrument Input Impedance1 MΩ

Environmental

ParameterValue
Storage Temperature-40 °C to +70 °C
Operating Altitude2000 m
Storage Altitude12000 m

Performance

ParameterValue
Accuracy (100 mV/A Range)3% of reading + 50 mA (48 Hz to 65 Hz)
Accuracy (10 mV/A Range)2% of reading + 0.5 A (48 Hz to 65 Hz)
Accuracy (1 mV/A Range)2% of reading (100 A to 1000 A), 3% of reading (1000 A to 2000 A) (48 Hz to 65 Hz)
Phase Shift (100 mV/A Range)≤ 2° at 50/60 Hz (for currents 0.1 A to 10 A)
Phase Shift (10 mV/A Range)≤ 1° at 50/60 Hz (for currents 1 A to 100 A)
Phase Shift (1 mV/A Range)≤ 1° at 50/60 Hz (for currents 10 A to 2000 A)
Conductor Position Influence≤ 0.5% of reading (at 50/60 Hz)
Adjacent Conductor Influence≤ 15 mA/A (at 50 Hz)

Compliance

ParameterValue
Pollution Degree2

Compliance & Safety

ParameterValue
Safety StandardsUL 61010-1, CAN/CSA-C22.2 No. 61010-1, EN 61010-2-032, IEC 61010-2-032

Technology

Passive electromagnetic transformer

Applications

AC current measurement for oscilloscopes and multimeters