The Tektronix P5200 is an active high-voltage differential oscilloscope probe designed to allow safe, floating measurements of circuit components. With a bandwidth of 50 MHz, attenuation settings of 50x and 500x, and a maximum differential voltage range of up to ±1300 V, it is ideal for analyzing power electronics, motor drives, and industrial electrical systems. The probe features standard safety banana plug inputs and an integrated warning system with audible and visual overrange indicators.

Specifications

General

ParameterValue
InterfacesBNC (male) output with specialized power jack
Warm-up Period20 minutes
Standard Warranty1 year

Physical

ParameterValue
Power Supply9 V DC via external AC adapter or powered from a Tektronix 1103 power supply
Operating Temperature0 to +50 °C

Electrical & Signal Performance

ParameterValue
Bandwidth50 MHz
Rise Time< 7 ns
Attenuation Ratios50x / 500x
DC Attenuation Accuracy±3%
Maximum Differential Voltage (50x)±130 V (DC + peak AC)
Maximum Differential Voltage (500x)±1300 V (DC + peak AC)
Maximum Common-Mode Voltage±1300 V (DC + peak AC)
Maximum Input Voltage-to-Earth (RMS)1000 V RMS
Maximum Input Voltage-to-Earth (CAT Rating)1000 V CAT II / 600 V CAT III
Input Resistance (Differential)8 MΩ
Input Capacitance (Differential)4 pF
Input Resistance (Each Input to Ground)4 MΩ
Input Capacitance (Each Input to Ground)8 pF
Common-Mode Rejection Ratio (CMRR) at 50/60 Hz-80 dB
Common-Mode Rejection Ratio (CMRR) at 100 kHz-50 dB
Common-Mode Rejection Ratio (CMRR) at 1 MHz-30 dB
Output Voltage Range (Into 1 Megohm)±2.6 V

Connectivity & Interface

ParameterValue
Output Connector TypeBNC (male)
Oscilloscope Termination Requirement1 MΩ
Input Lead Connection InterfaceIntegrated banana plugs (shrouded)

Power Requirements

ParameterValue
Power Source OptionsExternal AC Adapter (9 V DC output) or Tektronix 1103 Power Supply
AC Adapter Output9 VDC, 200 mA

Safety & Compliance

ParameterValue
Overvoltage Category Rating1000 V CAT II / 600 V CAT III
Safety StandardsUL3111-1, CSA1010.1, EN61010-1, IEC61010-2-031
Overrange IndicatorAudible buzzer and red LED

Environmental Specifications

ParameterValue
Operating Temperature Range0 °C to +50 °C
Storage Temperature Range-40 °C to +75 °C
Operating Humidity Range (Relative Humidity)Up to 90% RH at or below 30 °C
Operating Altitude Limit2000 m
Storage Altitude Limit15240 m

Physical Characteristics

ParameterValue
Input Lead Length30 cm
Output Cable Length1.5 m

Accessories & Calibration

ParameterValue
Calibration Interval Requirement1 year

Performance

ParameterValue
Propagation Delay18 ns (typical)
Output Impedance50 Ω (nominal)
Output Offset< ±10 mV (typical)

Power

ParameterValue
Battery Life3 hours (typical with alkaline battery)
Current Consumption35 mA (typical)

Environmental & Compliance

ParameterValue
Pollution Degree2

Electrical Characteristics

ParameterValue
Output Noise< 2.5 mV RMS (typical, referred to output)
Overrange Threshold> ±130 V (at 50x attenuation), > ±1300 V (at 500x attenuation)

Environmental

ParameterValue
Storage Humidity RangeUp to 90% RH (non-condensing) at +40 °C

Technology

Active High-Voltage Differential Probe

Applications

Floating measurements, power supply analysis, motor control, power semiconductor evaluation