The Tektronix P6248 is a 1.5 GHz active differential oscilloscope probe designed for high-bandwidth time and frequency domain measurements. Optimized for use with disk drive read channel electronics, RAMBUS digital IC designs, and high-speed communication interfaces like Gigabit Ethernet, FireWire, and Fibre Channel, it features a miniature probe head and versatile accessories for precise surface-mount device probing.

Specifications

General

ParameterValue
InterfacesTEKPROBE BNC Interface

Physical

ParameterValue
Power SupplyPowered directly via TEKPROBE BNC interface or 1103 TEKPROBE power supply
Operating Temperature0 to +50 °C

Electrical Performance

ParameterValue
Guaranteed Bandwidth1.5 GHz
Typical Bandwidth (1X Mode)1.7 GHz (probe only, ≤27 °C)
Typical Bandwidth (10X Mode)1.85 GHz (probe only, ≤27 °C)
Rise Time<265 ps
DC Attenuation Accuracy±2%
Linearity±2%
Output Noise (Referred to Input)<50 nV/√Hz
Harmonic Distortion≤1.5%
Passband Ripple<±0.25 dB
Propagation Delay6.1 ns (typical)

Impedance and Loading

ParameterValue
Input Resistance (Differential Mode)200 kΩ
Input Resistance (Common Mode)100 kΩ
Input Capacitance (Differential Mode)<1 pF
Input Capacitance (Common Mode)<2 pF

Signal Range and Voltage Limits

ParameterValue
Common Mode Input Voltage Range±7.0 V
Differential Input Voltage Range (1X)±850 mV
Differential Input Voltage Range (10X)±8.5 V
Maximum Nondestructive Input Voltage±25 V (DC + peak AC)
Electrostatic Immunity (ESD)>15 kV

Common-Mode Rejection Ratio (CMRR)

ParameterValue
CMRR at 1 MHz>60 dB (1000:1)
CMRR at 100 MHz>45 dB (180:1)
CMRR at 500 MHz>38 dB (80:1)
CMRR at 1 GHz>30 dB (30:1)

Physical Characteristics

ParameterValue
Probe Input ConnectorTwo standard 0.025 inch / 0.63 mm (0.1 inch center) female square pin receptacles
Cable Length1.2 m

Environmental and Compliance

ParameterValue
Non-Operating Temperature Range-55 °C to +75 °C

Connectivity

ParameterValue
Host Interface CompatibilityTekProbe BNC (connects directly to compatible Tektronix oscilloscopes, or requires 1103 power supply)

Environmental

ParameterValue
Operating Humidity5% to 95% RH (at or below +30 °C), 5% to 75% RH (above +30 °C to +50 °C)
Non-Operating Humidity5% to 95% RH (at or below +30 °C), 5% to 75% RH (above +30 °C to +75 °C)
Operating Altitude4,500 m
Non-Operating Altitude15,240 m

Electrical Characteristics

ParameterValue
Input CouplingDC
System Attenuation Settings1X, 10X
Output Termination Requirement50 Ω
Maximum Output Voltage Range±850 mV

Compliance

ParameterValue
Safety CertificationsUL3111-1, CAN/CSA C22.2 No. 1010.1, EN61010-1, IEC61010-1

Power

ParameterValue
Power RequirementsPowered directly by host instrument via TekProbe BNC interface or through 1103 power supply

Technology

Active Differential Probe

Applications

Disk drive read channel verification, RAMBUS design, Gigabit Ethernet, IEEE-1394 Firewire, and Fibre Channel testing