The Tektronix P7350 is a high-bandwidth, low-loading 5.0 GHz active differential probe designed for use with high-performance oscilloscopes featuring the TekConnect interface. It is ideal for debugging and validating high-speed differential signals in communication, semiconductor, and computer bus architectures.

Specifications

General

ParameterValue
InterfacesTekConnect
Warranty1 year
Recommended Calibration Interval1 year

Physical

ParameterValue
Power SupplyPowered directly by host oscilloscope through TekConnect interface
Operating Temperature0 °C to +40 °C

Electrical Performance

ParameterValue
Bandwidth5.0 GHz
Rise Time (10% to 90%)<= 100 ps
Rise Time (20% to 80%)< 70 ps (typical)
Attenuation Ratio5:1 (5X)
DC Attenuation Accuracy±2%
Differential Input Resistance100 kΩ
Common-Mode Input Resistance50 kΩ
Differential Input Capacitance< 0.3 pF (typical 0.27 pF)

Voltage Limits & Dynamic Range

ParameterValue
Differential Dynamic Range±1.25 V (2.5 Vp-p)
Common-Mode Operating Voltage Range±6.25 V
Maximum Non-Destructive Input Voltage (DC + peak AC)±15 V
DC Offset Range±1.25 V

Common Mode Rejection Ratio (CMRR)

ParameterValue
CMRR at DC> 60 dB
CMRR at 1 MHz> 55 dB (typical)
CMRR at 100 MHz> 45 dB (typical)
CMRR at 500 MHz> 38 dB (typical)
CMRR at 1 GHz> 30 dB (typical)
CMRR at 5 GHz> 15 dB (typical)

Host Interface & Connectivity

ParameterValue
Output Impedance50 Ω
Probe Interface TypeTekConnect
Supported Oscilloscope FamiliesTDS6000, TDS7000, CSA7000, TDS8000, CSA8000, DPO7000, DPO/DSA70000 series with TekConnect interface
Auto-Scaling SupportYes, automatically communicates attenuation factor and scaling to host oscilloscope

Physical & Mechanical

ParameterValue
Cable Length1.3 m (51 in)
Connector TypeTekConnect

Environmental & Compliance

ParameterValue
Non-Operating Temperature Range-55 °C to +75 °C
Operating Humidity Range0% to 90% RH (at or below +30 °C), 0% to 75% RH (above +30 °C to +40 °C)
Non-Operating Humidity Range0% to 90% RH (at or below +30 °C), 0% to 75% RH (above +30 °C to +75 °C)
Operating AltitudeUp to 3,000 m (9,843 ft)
Non-Operating AltitudeUp to 15,240 m (50,000 ft)

Compliance

ParameterValue
Safety CertificationsUL61010B-1, CAN/CSA-C22.2 No. 1010.1, EN 61010-1

Performance

ParameterValue
Input-Referred Noise< 4 mV RMS

Technology

Active Differential

Applications