The Tektronix RSA306 is a portable, USB-powered real-time spectrum analyzer offering a frequency range of 9 kHz to 6.2 GHz with a 40 MHz real-time bandwidth. Powered through a USB 3.0 interface and utilizing the comprehensive Tektronix SignalVu-PC software on an external PC, the RSA306 delivers full-featured spectrum analysis, streaming capture, and high-performance DPX spectrogram analysis in a compact, ruggedized package designed to meet MIL-STD-28800 Class 2 environmental standards.

Specifications

General

ParameterValue
InterfacesUSB 3.0 Type Micro-B with screw-lock, SMA female trigger and reference connectors
DisplayNone (uses external PC screen via SignalVu-PC software)

Physical

ParameterValue
Power SupplyUSB 3.0 bus powered (5.0 V, 900 mA maximum, ≤4.5 W)
Operating Temperature-10 °C to +55 °C
Dimensions (W×H×D)127.0 mm × 30.5 mm × 190.5 mm
Weight0.59 kg
Form FactorPortable, USB-powered RF digitizer / receiver

Frequency

ParameterValue
Frequency Range9 kHz to 6.2 GHz
Real-Time Capture Bandwidth40 MHz
Center Frequency Resolution (Block IQ)1 Hz
Center Frequency Resolution (Streamed ADC)500 kHz
Frequency Reference Initial Accuracy±3.0 ppm (at 23 °C)
Frequency Reference Temperature Stability±14 ppm (-10 °C to 55 °C)
Frequency Reference Aging Rate±1.0 ppm per year

Amplitude & Sensitivity

ParameterValue
Measurement Range+20 dBm to -160 dBm
RF Input Impedance50 Ω
RF Input VSWR≤ 1.8:1 (10 MHz to 6200 MHz, reference level ≥ -15 dBm, typical)
Maximum Safe RF Input DC Voltage±40 VDC
Maximum Safe RF Input (Reference Level ≥ -10 dBm)+23 dBm (continuous or peak)
Displayed Average Noise Level (5 GHz)-150 dBm/Hz (typical)
Absolute Amplitude Accuracy±1.25 dB (9 kHz to 6.2 GHz, 18 °C to 28 °C)
Frequency Response Flatness±1.0 dB (over 40 MHz RF Bandwidth, 18 °C to 28 °C)
Input Attenuation Range0 dB to 51 dB
Input Attenuator Step Size1 dB

Real-Time & Acquisition

ParameterValue
Minimum Event Duration for 100% POI100 μs
A/D Converter Resolution14 bits
A/D Converter Sampling Rate112 MS/s
IF Bandwidth40 MHz
Maximum Block Acquisition Time1 second
Block Baseband Acquisition Bandwidths≤ 40 / (2^N) MHz, 0 Hz Digital IF, N ≥ 0
Block Baseband Acquisition Sample Rates≤ 56 / (2^N) MS/s, 32-bit float complex samples, N ≥ 0
Spurious-Free Dynamic Range (SFDR)50 dBc (typical)

Spectral Purity

ParameterValue
Phase Noise @ 10 kHz Offset-86 dBc/Hz (-89 dBc/Hz typical) at 1 GHz Center Frequency
Phase Noise @ 100 kHz Offset-89 dBc/Hz (-92 dBc/Hz typical) at 1 GHz Center Frequency
Phase Noise @ 1 MHz Offset-115 dBc/Hz (-120 dBc/Hz typical) at 1 GHz Center Frequency
Spurious Responses (Operating Range 18 °C to 28 °C)≤ -50 dBc (typical)

Triggering

ParameterValue
Trigger / Sync Input Impedance10 kΩ
IF Power Trigger Threshold Range0 dB to -50 dB from reference level
Trigger ModesFree run, Triggered
Trigger SourcesRF Input (IF power), Trigger/Sync Input (External SMA)

Software & Measurements

ParameterValue
Included SoftwareTektronix SignalVu-PC™ RF Signal Analysis Software
Display ModesReal-time Spectrum, Spectrogram/Waterfall, DPX Spectrum
Optional Measurement ApplicationsAM/FM/PM/Direct Audio Analysis, Settling Time, Advanced Signal Analysis (Pulse), General modulation analysis, OFDM, WLAN 802.11ac
Programming Interface SupportSignalVu-PC programmatic interface, Windows API, MATLAB driver for Instrument Control Toolbox

Inputs & Outputs

ParameterValue
RF Input ConnectorN-type female
Trigger / Sync Input ConnectorSMA female
Reference Frequency Input ConnectorSMA female (10 MHz reference input)
PC Connection PortUSB 3.0 Micro-B (with screw-lock holes)

Environmental & Compliance

ParameterValue
Environmental Standard ComplianceMIL-PRF-28800F Class 2 (operational and non-operational shock, vibration)
Storage Temperature Range-51 °C to +71 °C
Operating Humidity Range5% to 95% relative humidity (non-condensing)
Maximum Operating AltitudeUp to 9,144 meters (30,000 feet)

RF Input Performance

ParameterValue
Local Oscillator Feedthrough to Input< -75 dBm (with 0 dB input attenuation, -30 dBm reference level)

General Characteristics

ParameterValue
Warm-up Time20 minutes
Calibration Interval1 year

External Reference Input

ParameterValue
External Reference Input Frequency10 MHz
External Reference Input Level Range-10 dBm to +10 dBm
External Reference Input Frequency QualityMust be within ±0.5 ppm of 10 MHz

Host PC Requirements

ParameterValue
Host Processor RequirementIntel Core i7 4th Gen (or AMD equivalent) with AVX2 support
Host RAM Requirement8 GB minimum (16 GB recommended for streaming)
Host Interface RequirementUSB 3.0 port
Host OS SupportWindows 7, 8, 8.1, 10 (64-bit)

Compliance

ParameterValue
Safety ComplianceUL 61010-1, CAN/CSA-C22.2 No. 61010-1, EN 61010-1, IEC 61010-1
EMC ComplianceEN 61326-1, EN 61326-2-1, AS/NZS 2064, ICES-003, FCC 47 CFR Part 15 Subpart B Class A

Power

ParameterValue
USB Interface Power ConsumptionUp to 4.5 W (5 V, 900 mA)

Environmental

ParameterValue
Random Vibration (Operating)0.015 g²/Hz (5 Hz to 500 Hz) per MIL-PRF-28800F Class 2
Random Vibration (Non-Operating)0.030 g²/Hz (5 Hz to 500 Hz) per MIL-PRF-28800F Class 2

Technology

Real-Time Spectrum Analysis (RTSA)

Applications

RF signal analysis, interference hunting, academic labs, field testing, and design troubleshooting

Looking for the Tektronix RSA306?

Tell us what you need. We're building a network of equipment dealers and will get back to you if we can help source it — no obligation.

Selling this kind of equipment? Sell with us