The Tektronix TAP2500 is a 2.5 GHz active oscilloscope probe designed for high-speed signal acquisition with minimal device loading. Powered directly through the TekVPI interface, it provides outstanding electrical and physical performance, making it ideal for the design, validation, and debugging of high-speed digital systems.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | TekVPI |
Physical
| Parameter | Value |
|---|---|
| Power Supply | Powered directly by host oscilloscope via TekVPI interface |
| Operating Temperature | 0 °C to +50 °C |
Electrical Performance Parameters
| Parameter | Value |
|---|---|
| Bandwidth (Probe Only) | ≥ 2.5 GHz |
| Rise Time (10% to 90%, Probe Only) | < 140 ps |
| Attenuation Ratio | 10:1 |
| DC Input Resistance | 40 kΩ |
| Input Capacitance | < 0.8 pF |
| Input Dynamic Range | ±4 V |
| Input Offset Range | ±10 V |
| Maximum Non-Destructive Input Voltage | ±30 V (DC + Peak AC) |
| Propagation Delay | 5.3 ns (nominal) |
Physical & Mechanical Characteristics
| Parameter | Value |
|---|---|
| Cable Length | 1.3 m |
Host Compatibility & Power
| Parameter | Value |
|---|---|
| Host Interface Connection Type | TekVPI |
| Power Requirements | Powered directly by host oscilloscope via TekVPI interface |
Environmental & Safety
| Parameter | Value |
|---|---|
| Operating Temperature Range | 0 °C to +50 °C |
| Non-Operating Temperature Range | -40 °C to +71 °C |
| Operating Altitude | Up to 3,000 m (9,843 ft) |
| Non-Operating Altitude | Up to 12,192 m (40,000 ft) |
| Safety Standard Certifications | UL61010-031, CAN/CSA-C22.2 No. 61010-031, IEC61010-031 |
Accessories & Ordering Information
| Parameter | Value |
|---|---|
| Standard Warranty Period | 1 Year |
Physical Characteristics
| Parameter | Value |
|---|---|
| Probe Head Dimensions (W×H×D) | 5.1 mm × 7.6 mm × 57.2 mm |
Electrical Characteristics
| Parameter | Value |
|---|---|
| Output Termination | 50 Ω |
| Input Coupling | DC |
Performance
| Parameter | Value |
|---|---|
| System Bandwidth | 2.5 GHz (with ≥2.5 GHz TekVPI oscilloscopes) |
Compliance & Safety
| Parameter | Value |
|---|---|
| Pollution Degree | 2 |
Technology
Active Oscilloscope Probe
Applications
High-speed digital system design, validation, and debugging