The Tektronix TCP0030 is a high-performance, easy-to-use AC/DC current probe designed for use with and direct connection to oscilloscopes featuring the TekVPI interface. It provides a wide bandwidth of 120 MHz with selectable 5 A and 30 A measurement ranges, offering exceptional low-current measurement capability and accuracy down to 1 mA.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | TekVPI |
| Probe Type | Active AC/DC Current Probe |
| Host Compatibility | Tektronix oscilloscopes with TekVPI™ interface |
Physical
| Parameter | Value |
|---|---|
| Power Supply | Powered directly by host oscilloscope via TekVPI interface |
| Operating Temperature | 0 °C to +50 °C |
| Weight | 0.227 kg |
| Probe Head Dimensions (L×W×H) | 200 mm × 16 mm × 32 mm |
Electrical Performance Specs
| Parameter | Value |
|---|---|
| Bandwidth (-3 dB) | DC to >120 MHz |
| Rise Time (10% to 90%) | ≤2.92 ns |
| Attenuation Ratio / Gain Settings | 5 A and 30 A ranges |
| Maximum Input Current (RMS) | 30 A |
| Maximum Input Current (Peak Pulse) | 50 A |
| DC Accuracy (typical) | ±1% |
| DC Accuracy (guaranteed) | ±3% |
| Amp-Second Product | 500 A*μs (on 30 A range) |
| Maximum Input Voltage (Bare Wire) | 150 V CAT II |
| Maximum Input Voltage (Insulated Wire) | 300 V CAT II |
| Noise Floor (typical) | ≤2.5 mArms |
| Insertion Impedance at 1 MHz | 0.08 Ω |
| Insertion Impedance at 10 MHz | 0.15 Ω |
| Insertion Impedance at 50 MHz | 0.27 Ω |
| Insertion Impedance at 100 MHz | 0.7 Ω |
| Insertion Impedance at 120 MHz | 1.0 Ω |
Physical and Connectivity Specifications
| Parameter | Value |
|---|---|
| Output Connector / Host Interface | TekVPI |
| Cable Length | 2.0 m |
| Probe Tip / Max Conductor Diameter | 5.0 mm |
Power and Control
| Parameter | Value |
|---|---|
| De-gauss / Balance Capability | Degauss/Autozero button on probe compensation box |
| Status / Error Indicators | LEDs for range and probe status on compensation box |
Environmental and Safety Compliance
| Parameter | Value |
|---|---|
| Storage Temperature Range | -40 °C to +75 °C |
| Operating Humidity Range | 5% to 95% RH (at up to +30 °C), 5% to 85% RH (+30 °C to +50 °C) |
| Maximum Operating Altitude | 3,000 m |
| Maximum Storage Altitude | 12,192 m |
| Safety Standard Certifications | UL61010-1, UL61010-2-032, CAN/CSA-C22.2 No. 61010-1, CAN/CSA-C22.2 No. 61010-2-032, EN61010-1, EN61010-2-032 |
| EMC Compliance Standards | EN61326-1, Class A; EN61326-2-1 |
Performance
| Parameter | Value |
|---|---|
| Coupling | AC, DC |
| Sensitivity | 1 mA (on oscilloscopes supporting 1 mV/div setting) |
| Propagation Delay | 17 ns (typical) |
Technology
AC/DC Current Probe (Hall Effect and Transformer)
Applications
Current measurement