The Tektronix TCPA400 is a high-performance, dedicated current probe amplifier designed specifically for use with the TCP404XL AC/DC current probe. It forms a complete current measurement system capable of measuring up to 500 A continuous RMS and 750 A peak current, supporting a bandwidth of DC to 2 MHz.

Specifications

General

ParameterValue
InterfacesTEKPROBE Interface (BNC output with scale factor coding pins)
DisplayNone (LED indicator lights only)
Oscilloscope CompatibilityOscilloscopes with TEKPROBE interface or any 50 Ω BNC input
CoolingInternal fan

Physical

ParameterValue
Power Supply90 V to 264 V AC, 47 Hz to 440 Hz, max 50 W
Operating Temperature0 to 50 °C
Dimensions (W×H×D)130 mm × 156 mm × 292 mm
Weight1.75 kg
Form FactorBenchtop (rackmountable with optional kit)

System Performance with Compatible Probes

ParameterValue
Compatible Probe ModelTCP404XL
BandwidthDC to 2 MHz
Rise Time≤ 175 ns
Maximum Continuous Current (RMS)500 A
Maximum Peak Pulse Current750 A
DC Gain Accuracy±3% (warranted), ±1% (typical)
Display Noise (RMS)≤ 250 mA
Maximum Bare-Wire VoltageInsulated wire only
Amp-Second Product50,000 A-µs

Amplifier Characteristics

ParameterValue
Output Impedance50 Ω
Output TerminationRequires 50 Ω oscilloscope input
Signal CouplingAC, DC
Deflection Ranges1 A/V, 10 A/V

Connectivity & Interfaces

ParameterValue
Probe Input Connector TypeTEKPROBE current probe interface (proprietary multi-pin)
Signal Output Connector Type50 Ω BNC

Control & Calibration Functions

ParameterValue
Degauss and Autobalance ControlSingle-button manual control with integrated auto-routine
Manual Balance AdjustmentButtons
Overload Indicator LEDYes (red LED)
Degauss Needed Indicator LEDYes (yellow LED)
Probe Open Indicator LEDYes (red LED)
Selected Range Indicator LEDsLEDs for 1 A/V and 10 A/V

Power Specifications

ParameterValue
AC Input Voltage Range90 V to 264 V AC
AC Input Frequency47 Hz to 440 Hz
Power Consumption50 W maximum

Environmental Specifications

ParameterValue
Non-Operating Temperature Range-40 °C to +75 °C
Operating Humidity Range5% to 95% RH (to +30 °C), 5% to 85% RH (+30 °C to +50 °C)
Non-Operating Humidity Range5% to 95% RH (to +30 °C), 5% to 85% RH (+30 °C to +75 °C)
Operating AltitudeUp to 2,000 m
Non-Operating AltitudeUp to 12,192 m

Compliance & Certifications

ParameterValue
Safety CertificationsUL61010-1, CSA C22.2 No. 1010.1, EN61010-1
Electromagnetic Compatibility (EMC)EN61326 Class A, FCC Part 15 Subpart B Class A

Performance

ParameterValue
Warm-Up Time20 minutes
Degauss and Autobalance Cycle Time5 seconds (typical)
DC Gain Drift≤ 0.1%/°C (typically 0.05%/°C)
Output Offset Drift≤ 100 μV/°C (typically 20 μV/°C)

Outputs

ParameterValue
Full-Scale Output Voltage1 V (into 50 Ω)

Compliance & Safety

ParameterValue
Overvoltage CategoryCategory II
Pollution Degree2

Power

ParameterValue
Fuse Rating3.15 A, 250 V, Time-Lag (T), 5x20 mm

Accessories

ParameterValue
Standard Accessories50 Ω BNC cable (012-0124-00), 50 Ω feedthrough termination (011-0049-02), Instruction manual, Certificate of Traceable Calibration
Optional AccessoriesTravel case (016-1924-00), Rackmount kit (016-1922-00)

Technology

AC/DC Current Measurement

Applications

Power electronics testing, motor drive analysis, power supply design