The Tektronix THDP0100 is a high-voltage active differential oscilloscope probe designed for safe, high-resolution measurements up to ±6000 V. Equipped with the TekVPI interface, it integrates seamlessly with compatible Tektronix oscilloscopes, providing automatic scaling, unit detection, and direct power without requiring an external power supply.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | TekVPI |
| Display | None (LED indicators for attenuation range and status) |
| Overrange Indicator | Audible alarm and visual LED |
| Host Compatibility | Tektronix oscilloscopes with TekVPI interface |
Physical
| Parameter | Value |
|---|---|
| Power Supply | Powered directly via TekVPI interface |
| Operating Temperature | 0 °C to +50 °C |
| Dimensions (W×H×D) | 207 mm × 38 mm × 22 mm (probe head) |
| Weight | 0.362 |
Electrical Performance
| Parameter | Value |
|---|---|
| Bandwidth | 100 MHz |
| Rise Time (10% to 90%) | < 3.5 ns |
| Attenuation Ratios | 100x / 1000x |
| DC Attenuation Accuracy | ±2% |
| Input Resistance (Differential Mode) | 40 MΩ |
| Input Resistance (Common Mode) | 20 MΩ |
| Input Capacitance (Differential Mode) | < 2.5 pF |
| Input Capacitance (Common Mode) | < 5 pF |
| Propagation Delay | 18.5 ns |
Voltage Range and Signal Limits
| Parameter | Value |
|---|---|
| Differential Dynamic Range (1000x) | ±6000 V |
| Differential Dynamic Range (100x) | ±600 V |
| Common Mode Operating Range | ±6000 V |
| Maximum Non-Destructive Input Voltage | 2300 V CAT I / 1000 V CAT III |
Common Mode Rejection Ratio (CMRR)
| Parameter | Value |
|---|---|
| CMRR at DC | > 80 dB |
| CMRR at 100 kHz | > 60 dB |
| CMRR at 3.2 MHz | > 30 dB |
| CMRR at 100 MHz | > 26 dB |
Connectivity and Interfaces
| Parameter | Value |
|---|---|
| Input Connection Type | Shrouded banana plugs |
| Input Lead Length | 1.5 m |
| Output Cable Length | 1.2 m |
Environmental and Safety Compliance
| Parameter | Value |
|---|---|
| Storage Temperature Range | -40 °C to +75 °C |
| Operating Humidity Range | 5% to 85% RH (at or below +30 °C, non-condensing) |
| Operating Altitude | Up to 3000 m |
| Non-Operating Altitude | Up to 15240 m |
| Safety Standards | UL61010-031, CAN/CSA-C22.2 No. 61010-031, EN61010-031, IEC61010-031 |
Performance
| Parameter | Value |
|---|---|
| Bandwidth Limit Filter | 5 MHz (selectable) |
| Noise (100X) | < 30 mVrms (referred to input) |
| Noise (1000X) | < 300 mVrms (referred to input) |
Inputs & Outputs
| Parameter | Value |
|---|---|
| Output Interface / Connector | TekVPI |
User Interface
| Parameter | Value |
|---|---|
| Overrange Indicator Type | Audible and visual (LED) |
Environmental
| Parameter | Value |
|---|---|
| Pollution Degree | 2 |
Technology
Active Differential
Applications
High Voltage Power Analysis, Power Supply Design, Motor Drive Testing
Looking for the Tektronix THDP0100?
Tell us what you need. We're building a network of equipment dealers and will get back to you if we can help source it — no obligation.
Selling this kind of equipment? Sell with us