The Yokogawa AQ6370C is a high-performance optical spectrum analyzer (OSA) designed for the 600 to 1700 nm wavelength range. Well-suited for both research and development and manufacturing environments, it is available in standard (AQ6370C-10) and high-performance (AQ6370C-20) editions. Featuring an advanced monochromator with a stray-light suppression ratio that enables a dynamic range of up to 78 dB, the AQ6370C supports both single-mode and multimode optical fibers through a unique free-space optical input.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | GP-IB, RS-232, Ethernet (10/100Base-T), 4 × USB (Type A) |
| Display | 10.4-inch color LCD with SVGA (800 × 600) resolution |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 100 to 240 V AC, 50/60 Hz, maximum power consumption approx. 100 VA |
| Operating Temperature | 5 to 35 |
| Dimensions (W×H×D) | 426 × 221 × 459 |
| Weight | 19.0 |
Wavelength Specifications
| Parameter | Value |
|---|---|
| Wavelength Range | 600 to 1700 nm |
| Wavelength Accuracy (Standard Model AQ6370C-10) | ±0.02 nm (1520 to 1580 nm) |
| Wavelength Accuracy (High-Performance Model AQ6370C-20) | ±0.01 nm (1520 to 1580 nm) |
| Wavelength Linearity | ±0.01 nm (1520 to 1580 nm) / ±0.02 nm (1580 to 1620 nm) |
| Wavelength Repeatability | ±0.005 nm (1 min) |
| Wavelength Span Range | 0.1 nm to 1100 nm, and 0 nm (coarse) |
Optical Power and Amplitude
| Parameter | Value |
|---|---|
| Dynamic Range (at ±0.2 nm from peak) | 70 dB (AQ6370C-10) / 73 dB (AQ6370C-20) |
| Dynamic Range (at ±0.4 nm from peak) | 78 dB typical (AQ6370C-20) |
| Power Measurement Range (Sensitivity) | -90 to +20 dBm |
| Maximum Input Power | +20 dBm |
| Power Level Accuracy | ±0.4 dB (at 1310/1550 nm, -20 dBm input) |
| Power Flatness | ±0.1 dB (1520 to 1580 nm) |
| Polarization Dependency (PDL) | ±0.05 dB (at 1310/1550 nm) |
| Level Linearity | ±0.05 dB (-50 to +10 dBm) |
| Optical Return Loss | 30 dB or more (with flat connector) |
| Level Sensitivity Categories | 7 user-selectable categories |
| Level Sensitivity (1000 to 1300 nm) | -80 dBm |
Resolution and Bandwidth
| Parameter | Value |
|---|---|
| Resolution Bandwidth Settings (FWHM) | 0.02, 0.05, 0.1, 0.2, 0.5, 1.0, 2.0 nm |
| Minimum Resolution Bandwidth | 0.02 nm |
Measurement Speed and Sweep
| Parameter | Value |
|---|---|
| Sweep Time (Minimum / 100nm Span) | 0.2 s (100 nm span) |
| Sampling Points per Sweep | 50,001 data sampling points |
Optical Interfaces
| Parameter | Value |
|---|---|
| Optical Input Connector Type | Free space input (supports various optical connector adapters) |
| Compatible Fiber Types | Single-mode (SM) and Multimode (MM) fibers |
Analysis and Application Software
| Parameter | Value |
|---|---|
| Number of Traces | 7 individual traces |
| Spectral Analysis Functions | 13 built-in spectral analysis functions (WDM, LED, Spectral Width, Filter, FP-LD, DFB-LD, etc.) |
| Trace Calculations | Calculation between traces (e.g. subtraction) |
| Trace Hold Modes | Max/Min hold |
| Macro Programming | Supported (automatic execution of program recorded via key strokes and parameter selections) |
| All-at-Once Trace Filing | Supported (saves all 7 traces simultaneously in one file) |
External Connectivity and Control
| Parameter | Value |
|---|---|
| Internal Storage Capacity | 512 Mbyte (sufficient for >20,000 traces) |
| External Storage Compatibility | USB removable flash memory and external hard disk drives |
| Remote Control Command Set | SCPI compatible and AQ6317 emulation mode |
| Software Driver Support | LabVIEW driver available |
| USB Peripheral Support | USB mouse, keyboard, and storage device interfaces |
Environmental and Regulatory
| Parameter | Value |
|---|---|
| Calibration Interval | 1 year (recommended) |
| Operating Humidity Range | 20 to 80% RH (no condensation) |
| Storage Temperature Range | -10 to 50 °C |
| Warm-up Time | approx. 1 hour |
Performance
| Parameter | Value |
|---|---|
| Safe Maximum Input Power | +20 dBm (total input power) |
| Level Sensitivity (1300 to 1620 nm) | -90 dBm (sensitivity setting: HIGH3) |
| Level Sensitivity (1620 to 1700 nm) | -85 dBm (sensitivity setting: HIGH3) |
| Dynamic Range at ±0.1 nm (High-Performance Model) | 50 dB (at 0.02 nm resolution, high dynamic mode ON) |
Inputs & Outputs
| Parameter | Value |
|---|---|
| Trigger Input Terminal | BNC connector, TTL level |
| Trigger Output Terminal | BNC connector, TTL level |
| Analog Output Terminal | BNC connector, 0 to 5 V (approx.) |
| External Monitor Output | VGA D-Sub 15-pin (female) |
| Analog Output Voltage | 0 to 5 V |
| External Display Output | VGA (D-Sub 15-pin) |
| GP-IB Specification | IEEE 488.1 / 488.2 compliant |
| Ethernet Interface | 10BASE-T/100BASE-TX (RJ-45) |
| USB Interfaces | Type A (USB 2.0) x 4, Type B (USB 1.1) x 1 |
Environmental
| Parameter | Value |
|---|---|
| Maximum Operating Altitude | 2000 m |
Safety & Compliance
| Parameter | Value |
|---|---|
| Laser Safety Class | Class 1 (IEC 60825-1:2007) |
Wavelength
| Parameter | Value |
|---|---|
| Wavelength Display Resolution | 0.001 nm |
| Resolution Bandwidth Accuracy | ±10% (0.05 to 2.0 nm settings, 1520 to 1580 nm) |
Level
| Parameter | Value |
|---|---|
| Level Display Resolution | 0.01 dB |
Sweep / Trigger
| Parameter | Value |
|---|---|
| Trigger Input Level | TTL level |
| Trigger Output Level | TTL level |
Power
| Parameter | Value |
|---|---|
| Power Consumption | 150 VA or less |
Technology
Optical Spectrum Analyzer (Monochromator-based)
Applications
Spectral measurements of active/passive optical components, optical transceivers, and DWDM transmission equipment
Looking for the Yokogawa AQ6370C?
Tell us what you need. We're building a network of equipment dealers and will get back to you if we can help source it — no obligation.