The Yokogawa WT1600 is a high-performance, high-precision digital sampling power analyzer capable of measuring up to six phases on a single unit. It serves a wide range of applications, from testing extremely small currents in energy-saving devices to evaluating large-sized inverter-driven loads and motors.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | GP-IB (standard), RS-232 (standard), Ethernet (optional /C10), VGA output (optional /V1) |
| Display | 8.4-inch color TFT LCD, 640 × 480 pixels |
| Warm-up Time | Approx. 30 minutes |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 100 to 240 VAC, 50/60 Hz, maximum 150 VA |
| Operating Temperature | 5 to 40 °C |
| Dimensions (W×H×D) | 426 × 177 × 430 mm |
| Weight | Approx. 15 kg |
Input Channels & Configuration
| Parameter | Value |
|---|---|
| Number of Input Elements | Up to 6 channels (phases) |
| Element Configuration Types | 5 A element and 30 A element |
| Phase Configurations | 1P2W, 1P3W, 3P3W, 3P3W (3V3A), 3P4W |
| Voltage Input Coupling | AC or DC |
| Current Input Coupling | AC or DC |
| Input Terminals Type | Safety terminal bushings for voltage; Binding posts for current |
Measurement Ranges & Input Characteristics
| Parameter | Value |
|---|---|
| Direct Voltage Range | 1.5 V, 3 V, 6 V, 15 V, 30 V, 60 V, 150 V, 300 V, 600 V, 1000 V |
| Direct Current Range (5 A Element) | 10 mA, 20 mA, 50 mA, 100 mA, 200 mA, 500 mA, 1 A, 2 A, 5 A |
| Direct Current Range (30 A Element) | 1 A, 2 A, 5 A, 10 A, 20 A, 30 A |
| External Current Sensor Input Range | 50 mV, 100 mV, 200 mV, 500 mV, 1 V, 2 V, 5 V, 10 V |
| Voltage Input Impedance | Approx. 10 MΩ, 12 pF |
| Current Input Impedance (5 A Element) | Approx. 100 mΩ + 0.05 µH |
| Current Input Impedance (30 A Element) | Approx. 6 mΩ + 0.05 µH |
| External Current Sensor Input Impedance | Approx. 1 MΩ, 40 pF |
| Crest Factor Range | Up to 3 (at rated input); up to 6 (at 1/2 rated input) |
| Maximum Input Voltage | 1000 Vrms, 1500 Vpeak |
| Maximum Input Current (5 A Element) | 10 Arms, 15 Apeak |
| Maximum Input Current (30 A Element) | 40 Arms, 100 Apeak |
Measurement Performance & Accuracy
| Parameter | Value |
|---|---|
| Frequency Range | DC, 0.5 Hz to 1 MHz |
| A/D Converter Resolution | 16 bits |
| Sampling Rate | Approx. 200 kS/s |
| Basic Voltage Measurement Accuracy | ±(0.1% of reading + 0.05% of range) at 50/60 Hz |
| Basic Current Measurement Accuracy | ±(0.1% of reading + 0.05% of range) at 50/60 Hz |
| Basic Active Power Accuracy | ±(0.1% of reading + 0.05% of range) at 50/60 Hz |
| Voltage Accuracy (DC) | ±(0.1% of reading + 0.1% of range) |
| Current Accuracy (DC) | 5 A element: ±(0.1% of reading + 0.1% of range); 30 A element: ±(0.1% of reading + 0.1% of range) + 10 mA |
| Active Power Accuracy (DC) | 5 A element: ±(0.1% of reading + 0.1% of range); 30 A element: ±(0.1% of reading + 0.1% of range) + (10 mA × voltage reading) |
| Maximum Common Mode Voltage | 1000 Vrms |
| Influence of Common Mode Voltage | ±0.01% of range or less (at 1000 Vrms, 50/60 Hz) |
Power & Harmonic Analysis Functions
| Parameter | Value |
|---|---|
| Measurement Parameters | U, I, P, S, Q, λ (PF), φ (phase angle), f (frequency), Upk+, Upk-, Ipk+, Ipk-, Wh, Wh+, Wh-, Ah, Ah+, Ah-, varh, η (efficiency) |
| Harmonic Measurement Bandwidth | PLL synchronization (10 Hz to 1.2 kHz) or external clock |
| Harmonic Order Limit | Up to 100th order (frequency dependent) |
| Harmonic Parameters Measured | Harmonic voltage, current, active power, phase angle, content, THD |
| Total Harmonic Distortion Modes | THD of fundamental (IEC) or THD of total (DIN) |
| Integration Modes | Standard, Continuous, Real-time Standard, Real-time Continuous |
| Efficiency Calculation Capacity | Up to 4 efficiency equations can be defined |
| Flicker Measurement | IEC61000-3-3 compliant (optional /FL) |
DC Source & Stimulus Capabilities
| Parameter | Value |
|---|---|
| Motor Evaluation Function | Analog or pulse input for speed and torque; calculates motor and total efficiency (optional /MTR) |
| Speed Sensor Input Type | Analog voltage or pulse |
| Torque Sensor Input Type | Analog voltage or pulse |
Data Logging, Memory & Visualization
| Parameter | Value |
|---|---|
| Display Modes | Numerical, Waveform, Trend, Vector, Bar graph, Split screen |
| Display Update Interval | 0.05 s, 0.1 s, 0.2 s, 0.5 s, 1 s, 2 s, 5 s |
| Maximum Logging Rate | Up to 50 ms (20 times per second) into internal memory |
| Internal Storage Capacity | Approx. 11 MB |
| External Storage | PC Card slot (Type II PCMCIA), Floppy disk drive |
System Interfaces & Connectivity
| Parameter | Value |
|---|---|
| Standard Ports | VGA output (15-pin D-sub) |
| GP-IB Interface Option | Standard (IEEE-488 compliant) |
| Serial Interface Option | Standard (RS-232) |
| Ethernet Interface Option | 100BASE-TX / 10BASE-T (option /C10) |
| Built-in Printer Option | Thermal line dot method, 112 mm paper width (option /B5) |
| Analog Outputs (DAC) | 30 channels, ±5 V FS output voltage (option /DA30) |
| External Synchronization Input | For synchronizing two WT1600 units |
Power Supply & Safety Compliance
| Parameter | Value |
|---|---|
| Mains Input Voltage Range | 100 to 240 VAC |
| Mains Input Frequency | 50/60 Hz |
| Maximum Power Consumption | 150 VA |
| Safety Standards | EN61010-1, EN61010-2-031, CAT II 1000V / CAT III 600V |
| EMC Directive Standards | EN61326, EN55011 Class A |
| Withstanding Voltage | 1500 VAC for 1 minute (between power terminal and casing) |
| Insulation Resistance | 50 MΩ or more (at 500 VDC, between power terminal and casing) |
Physical & Environmental Specifications
| Parameter | Value |
|---|---|
| Operating Humidity Range | 20% to 80% RH (no condensation) |
| Storage Temperature Range | -25 to 60 °C |
| Storage Humidity Range | 20% to 80% RH (no condensation) |
| Installation Altitude | 2000 m or less |
| Cooling System Type | Forced air cooling (internal fan) |
| Rack-Mount Compatibility | Supports 19-inch rack-mount (using optional EIA/JIS kits) |
Scaling Function
| Parameter | Value |
|---|---|
| VT (PT) Ratio Range | 0.0001 to 99999.9999 |
| CT Ratio Range | 0.0001 to 99999.9999 |
| SF (Scaling Factor) Range | 0.0001 to 99999.9999 |
Filters
| Parameter | Value |
|---|---|
| Line Filter Cutoff Frequencies | 5.5 kHz, 50 kHz, or OFF |
| Frequency Filter Cutoff Frequencies | 500 Hz or OFF |
Averaging
| Parameter | Value |
|---|---|
| Averaging Types | Exponential average (attenuation constant 2 to 64), Moving average (average count 8 to 64) |
Measurement Ranges
| Parameter | Value |
|---|---|
| Phase Difference Display Range | -180.00° to 180.00° |
Integration
| Parameter | Value |
|---|---|
| Integration Time Range | 0000h00m00s to 10000h00m00s |
| Integration Time Accuracy | ±0.02% of reading |
Interfaces & Outputs
| Parameter | Value |
|---|---|
| Video Output Port | D-sub 15-pin VGA (option /V1) |
| D/A Output Channels | 30 channels (option /DA) |
| Built-in Printer Paper Width | 112 mm (option /B5) |
Storage
| Parameter | Value |
|---|---|
| PCMCIA Card Slot | Type II (standard) |
| Built-in Floppy Disk Drive | 3.5-inch 1.44 MB (standard) |
Performance
| Parameter | Value |
|---|---|
| Phase Angle Accuracy | ±0.15° (at 50/60 Hz) |
Technology
Digital Sampling Power Analysis
Applications
Efficiency testing of inverters, motors, hybrid/electric vehicles, renewable energy systems, and power quality analysis