The Yokogawa WT310 is a compact, high-performance single-phase digital power meter designed for accurate power consumption measurement in R&D, production line testing, and QA applications. Featuring high basic accuracy of 0.1%, a wide bandwidth from DC and 0.5 Hz to 100 kHz, and versatile integration capabilities, it delivers reliable measurement of power, voltage, current, power factor, and harmonics.

Specifications

General

ParameterValue
InterfacesUSB standard; GPIB, RS-232C, Ethernet, Analog D/A output (option-dependent)
Display7-segment LED display

Physical

ParameterValue
Power Supply100 V AC to 240 V AC, 50/60 Hz, max 50 VA
Operating Temperature5 °C to 40 °C
Dimensions (W×H×D)213 × 88 × 379 mm
Weight3.0 kg

Measurement Inputs and Ranges

ParameterValue
Number of Measurement Channels1 input element (1-phase 2-wire)
Voltage Input Ranges15 V, 30 V, 60 V, 150 V, 300 V, 600 V (Crest Factor 3)
Crest Factor Rating3 or 6
Voltage Input ImpedanceApprox. 2 MΩ in parallel with approx. 13 pF
Frequency Measurement Range0.5 Hz to 100 kHz

Accuracy and Performance

ParameterValue
Basic Active Power Accuracy±(0.1% of reading + 0.1% of range) at 50/60 Hz
Basic Voltage Accuracy±(0.1% of reading + 0.1% of range) at 50/60 Hz
Basic Current Accuracy±(0.1% of reading + 0.1% of range) at 50/60 Hz
Frequency BandwidthDC, 0.5 Hz to 100 kHz
A/D Converter Resolution16-bit
Data Update Interval100 ms, 250 ms, 500 ms, 1 s, 2 s, 5 s
Line Filter Cutoff Frequency500 Hz

Measurement Parameters and Functions

ParameterValue
Active Power Measurement (P)Supported
Apparent Power Measurement (S)Supported
Reactive Power Measurement (Q)Supported
Power Factor Measurement (PF)Supported
Phase Angle MeasurementSupported
Peak Hold FunctionVoltage and Current (+/- peak hold)
Integration FunctionAh (Ampere-hour) and Wh (Watt-hour) integration
Harmonic Measurement OptionUp to 50th order (with /G5 option installed)

Display and Control

ParameterValue
Display Type7-segment LED display
Simultaneous Display Parameters4 display areas

Interfaces and Connectivity

ParameterValue
USB InterfaceUSB Type-B (standard)
RS-232C InterfaceOptional (Suffix -C2)
GPIB InterfaceOptional (Suffix -C1)
Ethernet InterfaceOptional (Suffix -C7)
Analog D/A OutputOptional, 4-channel output (Option /DA4)

Power Supply

ParameterValue
Supply Voltage Range100 V AC to 240 V AC
Supply Frequency50 Hz / 60 Hz
Power ConsumptionMax 50 VA

Environmental and Safety

ParameterValue
Operating Temperature Range5 °C to 40 °C
Operating Humidity Range20% to 80% RH (non-condensing)
Storage Temperature Range-25 °C to 60 °C
Maximum Operating Altitude2000 m
Safety StandardsEN61010-1, EN61010-2-030
EMC StandardsEN61326-1 Class A, EN55011 Class A Group 1
Measurement CategoryCAT II 600 V

Options and Accessories

ParameterValue
Harmonic Measurement Option Code/G5
Ethernet Option Code-C7
GPIB Option Code-C1
RS-232 Option Code-C2
D/A Output Option Code/DA4

Input & Measurement

ParameterValue
Direct Current Input Ranges5 mA, 10 mA, 20 mA, 50 mA, 100 mA, 200 mA, 0.5 A, 1 A, 2 A, 5 A, 10 A, 20 A
Current Input ImpedanceApprox. 6 mΩ + 0.07 µH (0.5 A to 20 A range), approx. 500 mΩ (5 mA to 200 mA range)
External Sensor Input ImpedanceApprox. 100 kΩ

Input Limits

ParameterValue
Maximum Input Voltage1000 Vrms, 1500 Vpeak
Maximum Input Current30 Arms, 100 Apeak

General Specifications

ParameterValue
Warm-up TimeApprox. 30 minutes

Harmonic Measurement

ParameterValue
Maximum Harmonic Order50th order (with /G5 option)

Functions & Features

ParameterValue
Auto-Range IntegrationSupported (automatically switches range during active power/current integration)

Integration

ParameterValue
Integration Time Range0000:00:00 to 10000:00:00 (hours:minutes:seconds)

Software & Communication

ParameterValue
Bundled PC ApplicationWTViewerFreePlus

Maintenance

ParameterValue
Recommended Calibration Interval1 year

Technology

Digital Power Analyzer

Applications

Power consumption measurement, efficiency evaluation, standby power testing, and harmonic analysis