This comparison details the differences and similarities between the legacy industry-standard Keysight Technologies (Agilent HP) 34401A and its direct successor, the Keysight Technologies (Agilent HP) 34461A digital multimeter. Both are highly precise 6.5-digit instruments designed for benchtop and automated test system applications, sharing identical best one-year DC voltage accuracy specifications. However, the Keysight Technologies (Agilent HP) 34461A introduces a modern color graphical interface, standard USB and LAN connectivity, expanded current and sensor measurement ranges, and significantly larger internal memory capacity.

Best 1-Year DCV Accuracy
0.0035% of reading + 0.0005% of range
Resolution
6.5 Digits
34401A Interface
RS-232, GPIB Standard (USB Optional)
34461A Interface
USB 2.0, LAN Standard (GPIB Optional)
34401A Max Memory
512 readings
34461A Max Memory
10,000 readings
Keysight Technologies (Agilent HP) 34401AKeysight Technologies (Agilent HP) 34461A

Specifications

Specification
Keysight Technologies (Agilent HP)34401A
Keysight Technologies (Agilent HP)34461A
General
InterfacesRS-232, GPIB (Standard); USB (Optional via adapter)USB 2.0 (Host and Device), LAN (LXI Core), GPIB (optional)
DisplayVacuum Fluorescent Display (VFD)4.3-inch color TFT-LCD, 480 x 272 resolution, supporting trend charts, histograms, and bar charts
Measurement FunctionsDCV, ACV, DCI, ACI, 2-wire Resistance, 4-wire Resistance, Frequency, Period, Continuity, Diode Test—
Real-Time Clock—Yes (with battery backup)
SCPI Version—1999.0
Physical
Power Supply100V / 120V / 220V / 240V ±10%, 45 Hz - 66 Hz or 360 Hz - 440 Hz, 25 VA max100 V / 120 V / 220 V / 240 V (±10%), 50 Hz / 60 Hz / 400 Hz (±10%), 55 VA maximum
Operating Temperature(°C)0 to 550 °C to 55 °C
Dimensions (W×H×D)(mm)254 × 103.6 × 374261.2 mm × 103.8 mm × 303.2 mm (with bumpers and handle)
Weight(kg)3.63.76 kg
DC Voltage Specifications
Ranges100.0000 mV, 1.000000 V, 10.00000 V, 100.0000 V, 1000.000 V—
Maximum Resolution100 nV (on 100 mV range)—
Input Resistance>10 GΩ (100 mV, 1 V, 10 V ranges); 10 MΩ ±1% (100 V, 1000 V ranges)—
Input Bias Current< 30 pA at 25 °C—
Maximum Input Protection1000 VDC—
Common Mode Rejection Ratio (CMRR)140 dB (for 1 kΩ unbalance in LO lead, DC)—
Normal Mode Rejection Ratio (NMRR)60 dB (at 50/60 Hz ±0.1%, for 1 NPLC or greater)—
1-Year Accuracy±(0.0035% of reading + 0.0005% of range) on 10 V range—
DC Current Specifications
Shunt Resistance5 Ω for 10 mA / 100 mA; 0.1 Ω for 1 A / 3 A—
Burden Voltage< 0.15 V at 10 mA, < 1.5 V at 100 mA, < 0.3 V at 1 A, < 1 V at 3 A—
Resistance Specifications
Measurement MethodSelectable 4-wire or 2-wire ohms (Current source referenced to LO input)—
Maximum Lead Resistance (4-Wire)10% of range per lead for 100 Ω and 1 kΩ ranges; 1 kΩ per lead on all other ranges—
AC Voltage Specifications
Frequency Range3 Hz to 300 kHz—
Measurement TechniqueAC-coupled True RMS (measures AC component only)—
Maximum Crest Factor5:1 at full scale—
Input Impedance1 MΩ ±2% in parallel with 100 pF—
Frequency and Period Specifications
Frequency Ranges3 Hz to 300 kHz—
Period Ranges3.33 µs to 333 ms—
Input Signal Level100 mV to 750 V rms—
Continuity and Diode Test
Continuity ThresholdAdjustable from 1 Ω to 1000 Ω—
Diode Test Current1 mA—
Diode Voltage Range0 to 1.2 V—
Measurement Characteristics
A/D Converter TypeMulti-Slope III—
Integration Times (NPLC)0.02, 0.2, 1, 10, 100 Power Line Cycles—
Math FunctionsNull, Min/Max/Average, dB, dBm, Limit Test—
Reading Memory / Storage512 readings—
System and Operating Speeds
Maximum Readings per Second (Over Interface)1000 readings/second in ASCII format across GPIB—
Maximum Readings per Second (To Memory)1000 readings/second—
Triggering Characteristics
Trigger SourcesInternal, External, Bus (Software)—
Trigger Delay Range0 to 3600 seconds (10 µs step size)—
Interfaces and Programming
Supported Command LanguagesSCPI (Standard Commands for Programmable Instruments), Agilent 3478A, Fluke 8840A/8842A—
Environmental and Compliance
Storage Temperature Range-40 °C to 70 °C—
Warm-up Time1 hour—
Calibration Cycle1 year (recommended)—
Physical Specifications
Rack Mount Size2U half-rack—
Front and Rear Input TerminalsStandard, switch-selectable—
Performance
Overrange Capability20% on all ranges (except 1000 Vdc and 750 Vac)—
Max Reading Rate to USB Flash Drive—80 readings/s
4-Wire Resistance Max Lead Resistance—10% of range per lead (100 Ω, 1 kΩ ranges), 1 kΩ per lead (other ranges)
Power
Power Line Frequency45 Hz to 66 Hz, and 360 Hz to 440 Hz (automatically sensed at power-on)—
Power Consumption25 VA peak (10 W average)75 VA maximum
AC Power Voltage Options100V, 120V, 220V, 240V (±10%)—
AC Line Frequency Range—45 Hz to 66 Hz, 360 Hz to 440 Hz
Environmental
Shock and VibrationMIL-T-28800E Type III, Class 5 (operating)—
Acoustic Noise—35 dBA (nominal)
Vibration & Shock—MIL-PRF-28800F, Class 3
Compliance
Safety ComplianceIEC 1010-1, EN 61010-1, UL 1244, CSA 1010—
AC Characteristics
AC Filter Bandwidths3 Hz (slow), 20 Hz (medium), 200 Hz (fast)—
Performance Characteristics
A/D Linearity0.0002% of reading + 0.0001% of range—
System Speeds
Reading Speed (4½ Digits, 0.02 NPLC)1000 readings/s—
Reading Speed (5½ Digits, 1 NPLC)60 readings/s (60 Hz line) / 50 readings/s (50 Hz line)—
Reading Speed (6½ Digits, 10 NPLC)6 readings/s (60 Hz line) / 5 readings/s (50 Hz line)—
Measurement Specifications
Display Resolution—6.5 digits
DC Voltage Ranges—100 mV, 1 V, 10 V, 100 V, 1000 V
DC Voltage Best 1-Year Accuracy—0.0035% of reading + 0.0005% of range
AC Voltage Ranges—100 mV, 1 V, 10 V, 100 V, 750 V
AC Voltage Frequency Range—3 Hz to 300 kHz
AC Voltage Best 1-Year Accuracy—0.06% of reading + 0.03% of range (at 10 Hz - 20 kHz)
DC Current Ranges—100 µA, 1 mA, 10 mA, 100 mA, 1 A, 3 A, 10 A
DC Current Best 1-Year Accuracy—0.05% of reading + 0.005% of range (on 100 mA range)
AC Current Ranges—100 µA, 1 mA, 10 mA, 100 mA, 1 A, 3 A, 10 A
AC Current Frequency Range—3 Hz to 10 kHz
AC Current Best 1-Year Accuracy—0.10% of reading + 0.04% of range
Resistance Ranges—100 Ω, 1 kΩ, 10 kΩ, 100 kΩ, 1 MΩ, 10 MΩ, 100 MΩ (2-wire and 4-wire)
Resistance Best 1-Year Accuracy—0.010% of reading + 0.001% of range (on 10 kΩ range)
Diode Test Threshold / Source Current—5.0 V maximum test voltage, 1 mA source current
Continuity Threshold—1 Ω to 1000 Ω (adjustable)
Frequency and Period Range—3 Hz to 300 kHz (Period: 333 ms to 3.3 µs)
Frequency Best 1-Year Accuracy—0.01% of reading
Supported Temperature Sensors—RTD (PT100), Thermistor (2.2 kΩ, 5 kΩ, 10 kΩ)
System Performance & Digitizing
Maximum Reading Rate (to Internal Memory)—1,000 readings/s
Maximum Reading Rate (to Bus)—1,000 readings/s
Internal Memory Capacity (Readings)—10,000 readings
Aperture / Integration Time Range—0.02 NPLC, 0.2 NPLC, 1 NPLC, 10 NPLC, 100 NPLC
Input & Noise Characteristics
DC Voltage Input Impedance—10 MΩ ±1% (all ranges) or > 10 GΩ (100 mV, 1 V, 10 V ranges)
Input Bias Current—< 30 pA at 25 °C
Maximum Input Voltage (Terminal to Terminal)—1000 V (HI to LO)
Maximum Input Voltage (Terminal to Earth)—500 Vpk (LO to Earth)
DC Common Mode Rejection Ratio (CMRR)—140 dB (for 1 kΩ LO lead unbalance)
AC Common Mode Rejection Ratio (CMRR)—70 dB (at 50/60 Hz, for 1 kΩ LO lead unbalance)
Normal Mode Rejection Ratio (NMRR)—60 dB (at 50/60 Hz, for 1 NPLC or greater)
AC Crest Factor Limit—3:1 at full scale
Triggering & Mathematical Functions
Trigger Sources—Internal, External, Manual, Software
Trigger Delay Range—0 s to 3600 s
Pre-trigger Capability—Up to 10,000 readings
External Trigger Input Level / Pulse Width—TTL-compatible, minimum 1 µs pulse width
Voltmeter Complete (VM Comp) Output—TTL-compatible, ≈ 2 µs pulse width
Math Functions—Null, Min/Max/Average, Standard Deviation, Limit Test, dB, dBm
Connectivity & System Interfaces
USB Port (Host/Device)—USB 2.0 (Host on front panel, Device on rear panel)
LAN Interface / LXI Compliance—LAN (10/100 Base-T, LXI Core 2011 compliant)
GPIB Interface (Standard/Optional)—Optional (via option 3446GPBU)
Command Emulation Modes—Keysight 34401A SCPI
SCPI Language Compatibility—Standard SCPI-1999
Physical & Environmental Specifications
Storage Temperature Range—-40 °C to 70 °C
Operating Humidity Limit—Up to 80% RH at 40 °C (non-condensing)
Maximum Operating Altitude—3,000 meters (approx. 10,000 feet)
Warm-up Time—60 minutes
Compliance & Safety
Standard Calibration Interval—1 year
Electromagnetic Compatibility (EMC) Standards—EN 61326-1, IEC 61326-1, ICES-001
Standard Warranty Period—3 years
Measurement Ranges
Capacitance Ranges—1 nF, 10 nF, 100 nF, 1 µF, 10 µF, 100 µF
Maximum Diode Test Voltage—5 V
Continuity Test Current—1 mA
Measurement Accuracy
Capacitance Best 1-Year Accuracy—1.0% of reading + 0.3% of range
AC Measurement
AC Low-Frequency Filter Settings—3 Hz, 20 Hz, 200 Hz
Measurement Options
Resistance Measurement Configuration—2-wire and 4-wire
Safety & Compliance
Safety Category Rating—CAT I 1000 V, CAT II 300 V
Pollution Degree—Pollution Degree 2
Math & Analytics
Statistical Calculations—Min, Max, Average, Standard Deviation, Peak-to-Peak
Limits Testing—Pass/Fail limits with high/low visual indicators
Display Options
Histogram Display Mode—Supported
Trend Chart Display Mode—Supported
Display
Display Size—4.3 in (109 mm) diagonal
Display Resolution (Pixels)—480 × 272
Storage
Instrument State Storage—Supports saving any number of states to internal flash or external USB drive
Triggering
Voltmeter Complete Pulse Width—≈ 2 µs
External Trigger Latency—< 15 µs
External Trigger Jitter—< 1 µs

Display and Graphical Capabilities

The Keysight Technologies (Agilent HP) 34401A utilizes a traditional Vacuum Fluorescent Display (VFD) to output text and numeric data. In contrast, the Keysight Technologies (Agilent HP) 34461A features a 4.3-inch color TFT-LCD screen with a resolution of 480 x 272 pixels. This color display enables advanced graphical representations of measurement data directly on the instrument front panel, including real-time trend charts, histograms, and bar charts, facilitating quicker diagnostic analysis without an external computer.

Measurement Functions and Expanded Ranges

While both multimeters cover standard DMM functions like DCV, ACV, DCI, ACI, 2-wire/4-wire resistance, frequency, period, continuity, and diode testing, their specific ranges differ.

The Keysight Technologies (Agilent HP) 34461A extends current measurements down to a 100 µA range and up to 10 A for both AC and DC, whereas the Keysight Technologies (Agilent HP) 34401A spans up to 3 A. Additionally, the 34461A includes native support for capacitance measurements (from 1 nF to 100 µF) and temperature sensor measurements using RTDs (PT100) or thermistors (2.2 kΩ, 5 kΩ, and 10 kΩ), functions not standard on the 34401A.

Memory and Connectivity Interfaces

Data logging and communication capabilities differ significantly between the generations. The Keysight Technologies (Agilent HP) 34401A contains an internal memory capable of holding up to 512 readings and features standard RS-232 and GPIB interfaces, with USB only available via an optional external adapter.

The Keysight Technologies (Agilent HP) 34461A provides an internal memory capacity of 10,000 readings and supports pre-triggering up to 10,000 readings. It integrates USB 2.0 (Host on the front panel, Device on the rear) and LAN (LXI Core) interfaces as standard, while GPIB is available as an option.

Frequently asked questions

Can the 34461A emulate the programming commands of the older 34401A?
Yes, the Keysight Technologies (Agilent HP) 34461A features a 34401A SCPI command emulation mode, allowing it to serve as a direct replacement in legacy automated test systems.
How do the power consumption specifications compare?
The Keysight Technologies (Agilent HP) 34401A has a maximum power consumption of 25 VA peak (10 W average). The Keysight Technologies (Agilent HP) 34461A has a higher power consumption rated at 75 VA maximum.