The Keysight 34461A is a 6.5 digit Truevolt digital multimeter (DMM), designed as the direct replacement for the industry-standard Keysight/Agilent 34401A. It offers enhanced usability with a 4.3-inch color graphical display, built-in trend/histogram charts, and advanced Truevolt technology for superior measurement accuracy and reduced noise.

Specifications

General

ParameterValue
InterfacesUSB 2.0 (Host and Device), LAN (LXI Core), GPIB (optional)
Display4.3-inch color TFT-LCD, 480 x 272 resolution, supporting trend charts, histograms, and bar charts
Real-Time ClockYes (with battery backup)
SCPI Version1999.0

Physical

ParameterValue
Power Supply100 V / 120 V / 220 V / 240 V (±10%), 50 Hz / 60 Hz / 400 Hz (±10%), 55 VA maximum
Operating Temperature0 °C to 55 °C
Dimensions (W×H×D)261.2 mm × 103.8 mm × 303.2 mm (with bumpers and handle)
Weight3.76 kg

Measurement Specifications

ParameterValue
Display Resolution6.5 digits
DC Voltage Ranges100 mV, 1 V, 10 V, 100 V, 1000 V
DC Voltage Best 1-Year Accuracy0.0035% of reading + 0.0005% of range
AC Voltage Ranges100 mV, 1 V, 10 V, 100 V, 750 V
AC Voltage Frequency Range3 Hz to 300 kHz
AC Voltage Best 1-Year Accuracy0.06% of reading + 0.03% of range (at 10 Hz - 20 kHz)
DC Current Ranges100 µA, 1 mA, 10 mA, 100 mA, 1 A, 3 A, 10 A
DC Current Best 1-Year Accuracy0.05% of reading + 0.005% of range (on 100 mA range)
AC Current Ranges100 µA, 1 mA, 10 mA, 100 mA, 1 A, 3 A, 10 A
AC Current Frequency Range3 Hz to 10 kHz
AC Current Best 1-Year Accuracy0.10% of reading + 0.04% of range
Resistance Ranges100 Ω, 1 kΩ, 10 kΩ, 100 kΩ, 1 MΩ, 10 MΩ, 100 MΩ (2-wire and 4-wire)
Resistance Best 1-Year Accuracy0.010% of reading + 0.001% of range (on 10 kΩ range)
Diode Test Threshold / Source Current5.0 V maximum test voltage, 1 mA source current
Continuity Threshold1 Ω to 1000 Ω (adjustable)
Frequency and Period Range3 Hz to 300 kHz (Period: 333 ms to 3.3 µs)
Frequency Best 1-Year Accuracy0.01% of reading
Supported Temperature SensorsRTD (PT100), Thermistor (2.2 kΩ, 5 kΩ, 10 kΩ)

System Performance & Digitizing

ParameterValue
Maximum Reading Rate (to Internal Memory)1,000 readings/s
Maximum Reading Rate (to Bus)1,000 readings/s
Internal Memory Capacity (Readings)10,000 readings
Aperture / Integration Time Range0.02 NPLC, 0.2 NPLC, 1 NPLC, 10 NPLC, 100 NPLC

Input & Noise Characteristics

ParameterValue
DC Voltage Input Impedance10 MΩ ±1% (all ranges) or > 10 GΩ (100 mV, 1 V, 10 V ranges)
Input Bias Current< 30 pA at 25 °C
Maximum Input Voltage (Terminal to Terminal)1000 V (HI to LO)
Maximum Input Voltage (Terminal to Earth)500 Vpk (LO to Earth)
DC Common Mode Rejection Ratio (CMRR)140 dB (for 1 kΩ LO lead unbalance)
AC Common Mode Rejection Ratio (CMRR)70 dB (at 50/60 Hz, for 1 kΩ LO lead unbalance)
Normal Mode Rejection Ratio (NMRR)60 dB (at 50/60 Hz, for 1 NPLC or greater)
AC Crest Factor Limit3:1 at full scale

Triggering & Mathematical Functions

ParameterValue
Trigger SourcesInternal, External, Manual, Software
Trigger Delay Range0 s to 3600 s
Pre-trigger CapabilityUp to 10,000 readings
External Trigger Input Level / Pulse WidthTTL-compatible, minimum 1 µs pulse width
Voltmeter Complete (VM Comp) OutputTTL-compatible, ≈ 2 µs pulse width
Math FunctionsNull, Min/Max/Average, Standard Deviation, Limit Test, dB, dBm

Connectivity & System Interfaces

ParameterValue
USB Port (Host/Device)USB 2.0 (Host on front panel, Device on rear panel)
LAN Interface / LXI ComplianceLAN (10/100 Base-T, LXI Core 2011 compliant)
GPIB Interface (Standard/Optional)Optional (via option 3446GPBU)
Command Emulation ModesKeysight 34401A SCPI
SCPI Language CompatibilityStandard SCPI-1999

Physical & Environmental Specifications

ParameterValue
Storage Temperature Range-40 °C to 70 °C
Operating Humidity LimitUp to 80% RH at 40 °C (non-condensing)
Maximum Operating Altitude3,000 meters (approx. 10,000 feet)
Warm-up Time60 minutes

Compliance & Safety

ParameterValue
Standard Calibration Interval1 year
Electromagnetic Compatibility (EMC) StandardsEN 61326-1, IEC 61326-1, ICES-001
Standard Warranty Period3 years

Measurement Ranges

ParameterValue
Capacitance Ranges1 nF, 10 nF, 100 nF, 1 µF, 10 µF, 100 µF
Maximum Diode Test Voltage5 V
Continuity Test Current1 mA

Measurement Accuracy

ParameterValue
Capacitance Best 1-Year Accuracy1.0% of reading + 0.3% of range

AC Measurement

ParameterValue
AC Low-Frequency Filter Settings3 Hz, 20 Hz, 200 Hz

Measurement Options

ParameterValue
Resistance Measurement Configuration2-wire and 4-wire

Power

ParameterValue
AC Line Frequency Range45 Hz to 66 Hz, 360 Hz to 440 Hz
Power Consumption75 VA maximum

Safety & Compliance

ParameterValue
Safety Category RatingCAT I 1000 V, CAT II 300 V
Pollution DegreePollution Degree 2

Math & Analytics

ParameterValue
Statistical CalculationsMin, Max, Average, Standard Deviation, Peak-to-Peak
Limits TestingPass/Fail limits with high/low visual indicators

Display Options

ParameterValue
Histogram Display ModeSupported
Trend Chart Display ModeSupported

Display

ParameterValue
Display Size4.3 in (109 mm) diagonal
Display Resolution (Pixels)480 × 272

Environmental

ParameterValue
Acoustic Noise35 dBA (nominal)
Vibration & ShockMIL-PRF-28800F, Class 3

Storage

ParameterValue
Instrument State StorageSupports saving any number of states to internal flash or external USB drive

Triggering

ParameterValue
Voltmeter Complete Pulse Width≈ 2 µs
External Trigger Latency< 15 µs
External Trigger Jitter< 1 µs

Performance

ParameterValue
Max Reading Rate to USB Flash Drive80 readings/s
4-Wire Resistance Max Lead Resistance10% of range per lead (100 Ω, 1 kΩ ranges), 1 kΩ per lead (other ranges)

Technology

Truevolt technology

Applications