The Chroma 61603 is a highly reliable, programmable AC power source designed for high-precision testing and simulation applications. Rated at 1500 VA for AC output and 750 W for DC output, this single-phase power source provides clean, low-distortion power across a frequency range of 15 Hz to 1000 Hz. Standard interfaces include RS-232 and analog programming, making it highly suitable for commercial, defense, and aerospace test systems.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | RS-232, Analog Programming (standard); GPIB, USB, LAN (optional) |
| Display | LCD Display |
| Protection Features | Over Voltage Protection (OVP), Over Current Protection (OCP), Over Power Protection (OPP), Over Temperature Protection (OTP), Fan Fail |
| Cooling System | Forced air cooling by temperature-controlled fans |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 90 - 250 VAC, single-phase, 47 - 63 Hz |
| Operating Temperature | 0 to 40 °C |
| Dimensions (W×H×D) | 425 x 177 x 567 mm |
| Weight | 25 |
AC Output Specifications
| Parameter | Value |
|---|---|
| Maximum AC Power Rating | 1500 VA |
| Output Phase Configuration | 1 Phase (Single Phase) |
| AC Voltage Ranges | 150 V / 300 V |
| AC Voltage Resolution | 0.1 V |
| AC Voltage Accuracy | 0.2% of reading + 0.2% F.S. |
| Maximum RMS Current (Low Range) | 12 A (0 - 150 V) |
| Maximum RMS Current (High Range) | 6 A (0 - 300 V) |
| Maximum Peak Current (Low Range) | 72 A |
| Maximum Peak Current (High Range) | 36 A |
| Crest Factor | 6 |
| Frequency Range | 15 Hz - 1000 Hz, DC |
| Frequency Resolution | 0.01 Hz (15.00 - 99.99 Hz), 0.1 Hz (100.0 - 999.9 Hz) |
| Frequency Accuracy | 0.15% |
| Total Harmonic Distortion (THD) | < 0.3% (50/60 Hz, resistive load), < 1% (15 Hz - 1000 Hz) |
| Voltage Line Regulation | 0.1% |
| Voltage Load Regulation | 0.2% |
DC Output Specifications
| Parameter | Value |
|---|---|
| Maximum DC Power Rating | 750 W |
| DC Voltage Ranges | 212 V / 424 V |
| DC Voltage Resolution | 0.1 V |
| DC Voltage Accuracy | 0.2% of reading + 0.2% F.S. |
| Maximum DC Current (Low Range) | 6 A (0 - 212 V) |
| Maximum DC Current (High Range) | 3 A (0 - 424 V) |
| DC Voltage Line Regulation | 0.1% |
| DC Voltage Load Regulation | 0.2% |
Measurement & Readback
| Parameter | Value |
|---|---|
| Voltage Measurement Range (RMS) | 0 - 150 V / 0 - 300 V |
| Voltage Measurement Accuracy | 0.1% of reading + 0.1% F.S. |
| Current Measurement Accuracy (RMS) | 0.2% of reading + 0.2% F.S. |
| Active Power (W) Measurement Accuracy | 0.4% of reading + 0.4% F.S. |
| Frequency Measurement Accuracy | 0.1 Hz |
| Power Factor Measurement Range | 0.000 - 1.000 |
Input Power Requirements
| Parameter | Value |
|---|---|
| Input Phase Configuration | Single Phase |
| Input Voltage Range | 90 - 250 VAC |
| Input Frequency Range | 47 - 63 Hz |
| Input Power Factor | 0.97 (min) @ 110 VAC, full load; 0.95 (min) @ 220 VAC, full load |
Physical & Environmental Specifications
| Parameter | Value |
|---|---|
| Rack Height Units | 4U |
| Storage Temperature Range | -40 to 85 °C |
| Operating Humidity Range | 0 to 90% RH (non-condensing) |
Measurements
| Parameter | Value |
|---|---|
| Current Measurement Ranges | 0 - 2 A / 0 - 20 A |
| Current Measurement Resolution | 1 mA (Low Range) / 10 mA (High Range) |
| Active Power (W) Measurement Ranges | 0 - 150 W / 0 - 1500 W |
| Active Power (W) Measurement Resolution | 0.1 W (Low Range) / 1 W (High Range) |
| Peak Current Measurement Range | 0 - 6 A / 0 - 60 A |
| Crest Factor Measurement Range | 1.00 to 9.99 |
| Crest Factor Measurement Resolution | 0.01 |
| Power Factor Measurement Resolution | 0.001 |
| Peak Current Measurement Accuracy | ±(1.5% of rdg + 0.5% of range) for < 100 Hz; ±(3% of rdg + 1% of range) for 100 Hz to 1000 Hz |
Performance
| Parameter | Value |
|---|---|
| Voltage Temperature Coefficient | ±0.02% per °C |
| Remote Sense Compensation | 5 V |
Output Phase & Control
| Parameter | Value |
|---|---|
| Phase Angle Setting Range | 0° to 359.9° |
| Phase Angle Setting Resolution | 0.1° |
Output Characteristics
| Parameter | Value |
|---|---|
| Transient Recovery Time | < 150 µs (for output voltage to recover within 10% of rating under 100% load change) |
System Features
| Parameter | Value |
|---|---|
| Parallel Operation Capability | Up to 5 units (master-slave configuration) |
| Setup Memory Locations | 9 user-defined setups |
Inputs & Outputs
| Parameter | Value |
|---|---|
| Analog Programming Input Voltage | 0 to 5 V or 0 to 10 V (user-selectable for external amplitude control) |
Compliance & Safety
| Parameter | Value |
|---|---|
| Safety Standards | EN 61010-1 |
| EMC Standards | EN 61326-1, EN 61000-3-2, EN 61000-3-3 |
Technology
Linear / Switching Hybrid
Applications
AC/DC Power Simulation and Testing