The Chroma 63102A is a dual-channel, programmable DC electronic load module designed for use with the Chroma 6310A series mainframes. Each independent channel provides up to 100W of power, 80V of voltage, and 20A of current, making it highly suitable for testing multi-output AC/DC power supplies, DC/DC converters, chargers, and portable electronic components.

Specifications

General

ParameterValue
InterfacesVia mainframe (GPIB, USB, RS-232, RS-485, LAN optional interfaces)
DisplayNone (uses mainframe display or host software control)

Physical

ParameterValue
Power SupplyPowered directly by mainframe slot
Operating Temperature0 to 40 °C
Dimensions (W×H×D)82 × 172 × 409 mm
Weight4.8 kg

Input Ratings and Operating Ranges

ParameterValue
Number of Channels2 (isolated)
Maximum Input Power (per channel)100 W
Maximum Input Voltage (per channel)80 V
Maximum Input Current (per channel)20 A
Minimum Operating Voltage at Maximum Current1.6 V @ 20 A

Operating Modes & Accuracy

ParameterValue
Constant Current (CC) Low Range0 to 2 A
Constant Current (CC) High Range0 to 20 A
Constant Current (CC) Programming Resolution (Low Range)0.5 mA
Constant Current (CC) Programming Resolution (High Range)5 mA
Constant Current (CC) Programming Accuracy (Low Range)±(0.1% of setting + 0.1% of range)
Constant Current (CC) Programming Accuracy (High Range)±(0.1% of setting + 0.1% of range)
Constant Voltage (CV) Low Range0 to 16 V
Constant Voltage (CV) High Range0 to 80 V
Constant Voltage (CV) Programming Resolution (Low Range)4 mV
Constant Voltage (CV) Programming Resolution (High Range)20 mV
Constant Voltage (CV) Programming Accuracy±(0.05% of setting + 0.1% of range)
Constant Power (CP) Low Range0 to 10 W
Constant Power (CP) High Range0 to 100 W

Measurement and Readback Accuracy

ParameterValue
Voltage Readback Accuracy±(0.05% of reading + 0.05% of range)
Current Readback Accuracy±(0.1% of reading + 0.1% of range)

Physical and Mechanical Specifications

ParameterValue
Form FactorSingle-slot plug-in module for 6310A mainframe series
Compatible MainframesChroma 6312A (2-slot), Chroma 6314A (4-slot)

Protection

ParameterValue
Over Temperature Protection (OTP)Yes
Over Current Protection (OCP) Range0.05 to 2.05 A (Low Range), 0.1 to 20.5 A (High Range)
Over Power Protection (OPP) Range1 to 102.5 W

Dynamic Mode

ParameterValue
Dynamic Mode Frequency Range0.1 Hz to 20 kHz
Dynamic Mode Duty Cycle1% to 99% (1% resolution)
Slew Rate (Low Range)0.0008 to 0.08 A/μs
Slew Rate (High Range)0.008 to 0.8 A/μs
Slew Rate Resolution (Low Range)0.0008 A/μs
Slew Rate Resolution (High Range)0.008 A/μs
Dynamic Timer 1 & 2 Range0.025 ms to 9.999 s

Constant Resistance (CR) Mode

ParameterValue
Constant Resistance (CR) Low Range0.04 Ω to 2 Ω
Constant Resistance (CR) Medium Range1.5 Ω to 75 Ω
Constant Resistance (CR) High Range60 Ω to 3 kΩ

Measurement

ParameterValue
Voltage Readback Resolution (Low Range)4 mV
Voltage Readback Resolution (High Range)20 mV
Current Readback Resolution (Low Range)0.5 mA
Current Readback Resolution (High Range)5 mA

Technology

Programmable DC Electronic Load Module

Applications

Multi-output power supply testing, DC/DC converter evaluation, battery charger testing