The Fluke 199C ScopeMeter is a high-performance handheld oscilloscope designed for demanding applications. Featuring a 200 MHz bandwidth, 2.5 GS/s real-time sampling rate, and a high-resolution color display, it combines the analytical power of a benchtop oscilloscope with the rugged portability of a handheld field tool.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | Optically isolated RS-232 / USB (via optional PM9080 or OC4USB adapter cable) |
| Display | 144 mm (5.7 in) Color Active Matrix TFT LCD, 320 x 240 pixels |
| Calibration Interval | 1 year |
| Warranty | 3 years (instrument), 1 year (accessories) |
Physical
| Parameter | Value |
|---|---|
| Power Supply | Rechargeable Ni-MH battery pack (BP190) or external mains adapter/charger BC190 |
| Operating Temperature | 0 °C to 50 °C |
| Dimensions (W×H×D) | 256 mm × 169 mm × 64 mm |
| Weight | 2.0 kg |
Oscilloscope Specifications
| Parameter | Value |
|---|---|
| Bandwidth | 200 MHz |
| Number of Channels | 2 channels plus isolated DMM input |
| Max. Real-time Sampling Rate | 2.5 GS/s (single channel), 1.25 GS/s (dual channels) |
| Rise Time | 1.75 ns |
| Record Length | 3,000 points per channel (Scope mode), 27,500 points per channel (ScopeRecord mode) |
| Input Sensitivity | 2 mV/div to 100 V/div |
| Vertical Resolution | 8-bit |
| Input Coupling | AC, DC |
| Input Impedance | 1 MΩ ±1% // 15 pF ±2 pF |
| Max. Scope Input Voltage | 1000 V CAT II / 600 V CAT III (independently floating isolated inputs) |
| Vertical Accuracy | ±(1.5% of reading + 0.05 range/div) |
| Timebase Range | 5 ns/div to 5 s/div |
| Timebase Accuracy | ±(0.01% + 1 pixel) |
| Trigger Modes | Connect-and-View, Free Run, Single Shot, Edge, Delay, Video, Selectable Pulse Width, External |
| Trigger Sources | Input A, Input B, External (via DMM input) |
| Glitch Capture | 50 ns |
| Frequency Spectrum Analysis | FFT (Fast Fourier Transform) standard |
| Digital Persistence | Yes, with envelope, decay, and persistence modes |
| Waveform Capture and Replay | Automatic capture and replay of last 100 screens |
| Vpwm Function | Yes (for motor drive and frequency inverter applications) |
| Waveform Pass/Fail Testing | Yes, using waveform reference |
| Automatic Measurements | 30 automatic measurements |
Digital Multimeter (DMM) Specifications
| Parameter | Value |
|---|---|
| DMM Channels | 1 (isolated from oscilloscope inputs) |
| DMM Resolution | 5,000 counts |
| DC Voltage Ranges | 500 mV, 5 V, 50 V, 500 V, 1000 V |
| DC Voltage Accuracy | ±(0.5% + 5 counts) |
| AC Voltage Ranges | 500 mV, 5 V, 50 V, 500 V, 1000 V |
| AC Voltage Accuracy | ±(1.0% + 10 counts) |
| Resistance Ranges | 500 Ω, 5 kΩ, 50 kΩ, 500 kΩ, 5 MΩ, 30 MΩ |
| Resistance Accuracy | ±(0.6% + 5 counts) |
| Continuity Threshold | Audible alert at < 50 Ω (±30 Ω) |
| Diode Test | Up to 2.8 V |
| Capacitance Ranges | 50 nF, 500 nF, 5 µF, 50 µF, 500 µF |
| Current Measurement | Via optional current clamp or shunt |
Recorder Mode
| Parameter | Value |
|---|---|
| TrendPlot Record Time | Up to 22 days |
| ScopeRecord Capacity | 27,500 points per input |
| ScopeRecord Sampling Rate | 20 MS/s (maximum in roll mode) |
| ScopeRecord Timebase | 5 ms/div to 2 min/div |
| Recorder Functions | TrendPlot (chart recorder) and ScopeRecord (roll mode) |
Display and User Interface
| Parameter | Value |
|---|---|
| Display Type | Color Active Matrix TFT LCD |
| Display Size | 144 mm (5.7 in) diagonal |
| Display Resolution | 320 x 240 pixels |
| Backlight | Cold Cathode Fluorescent (CCFL) backlight |
| Contrast Adjustment | Brightness adjustment (no contrast adjustment) |
Connectivity and Interfaces
| Parameter | Value |
|---|---|
| Isolated Interface Port | Yes, optical isolation |
| PC Software Support | FlukeView® ScopeMeter software |
Power Specifications
| Parameter | Value |
|---|---|
| Battery Type | Ni-MH rechargeable battery pack (BP190) |
| Battery Operating Time | Up to 4 hours |
| Battery Charging Time | 4 hours (when instrument is off) |
| Mains Adapter | BC190 Charger/Adapter included |
| Mains Input Voltage | 115 V or 230 V (dependent on model version) |
Environmental and Safety
| Parameter | Value |
|---|---|
| Storage Temperature | -20 °C to 60 °C |
| Safety Category Rating | 1000 V CAT II, 600 V CAT III (IEC 1010-1) |
| Pollution Degree | 2 |
| IP Rating | IP51 (drip proof and dust protected) |
| Shock and Vibration | MIL-PRF-28800F Class 2 |
Physical Characteristics
| Parameter | Value |
|---|---|
| Enclosure Material | Rugged, shock-proof with integrated protective holster |
Triggering
| Parameter | Value |
|---|---|
| Video Triggering | NTSC, PAL, SECAM, and non-standard video (up to 2800 lines) |
| Pulse Width Triggering Range | 15 ns to 40 s |
| Trigger Delay Range | Up to 12 divisions pre-trigger, up to 650 divisions post-trigger |
| Connect-and-View Triggering | Continuous hands-off auto-triggering on fast, slow, and complex signals |
Measurement & Analysis
| Parameter | Value |
|---|---|
| Waveform Math Functions | Addition, Subtraction, Multiplication, FFT |
| Frequency Counter Resolution | 5 digits |
Power & Charger
| Parameter | Value |
|---|---|
| Mains Adapter Output | 17.8 V DC, 840 mA (via BC190) |
Environmental
| Parameter | Value |
|---|---|
| Operating Altitude | 3000 m (10000 ft) |
| Storage Altitude | 12000 m (40000 ft) |
| Operating Humidity | 0% to 95% (0 °C to 30 °C) / 0% to 75% (30 °C to 40 °C) / 0% to 45% (40 °C to 50 °C) |
Safety & Compliance
| Parameter | Value |
|---|---|
| Maximum Floating Voltage | 1000 V rms CAT II / 600 V rms CAT III (between any terminal and earth ground) |
| Safety Standards | EN 61010-1 2nd Edition |
| Electromagnetic Compatibility (EMC) | EN 61326-1 |
Memory & Storage
| Parameter | Value |
|---|---|
| Setup Memories | 15 screen and setup memories |
| Waveform Memories | 15 screen and setup memories |
| Recording Memories | 2 records (each stores 100 replay screens, ScopeRecord, or TrendPlot) |
Measurement Modes & Functions
| Parameter | Value |
|---|---|
| Zoom Function | Up to 16x (in Scope mode) / up to 100x (in ScopeRecord mode) |
Technology
Digital Storage Oscilloscope (DSO)
Applications
Industrial electronic troubleshooting, field service, electrical testing, and motor drive maintenance