The Fluke 5500A is a versatile multi-product calibrator designed to cover a broad spectrum of medium-accuracy electrical measurement devices. It functions as a single, compact standard capable of sourcing direct and alternating voltage and current, resistance, capacitance, and simulated power. With dedicated options for oscilloscope calibration and comprehensive temperature simulation (thermocouples and RTDs), it effectively replaces multiple standalone calibration instruments.

Specifications

General

ParameterValue
InterfacesRS-232; IEEE-488 (GPIB)

Physical

ParameterValue
Power Supply100 / 120 / 220 / 240 V AC ±10%, 47 to 63 Hz
Operating Temperature0 to +50 °C
Dimensions (W×H×D)432 × 178 × 473 mm
Weight20 kg

General Specifications

ParameterValue
Frequency Range0.01 Hz to 2.0 MHz
Output CapabilitiesDC/AC voltage, DC/AC current, resistance, capacitance, simulated power (phase control), thermocouples, RTDs
Waveform GenerationMultiple waveforms and harmonics

DC Voltage Specifications

ParameterValue
Output Ranges0 to ±1020 V

Temperature Calibration

ParameterValue
Supported TypesThermocouples and RTDs

Power and Energy (Simulated)

ParameterValue
DC/AC Power CapabilitiesDual DC or AC outputs (two voltages, or voltage + current) with phase control
DC Power Range109 µW to 11 kW
AC Power Range109 µW to 11 kW

Oscilloscope Calibration Options

ParameterValue
Leveled Sine Wave BandwidthUp to 600 MHz (with Oscilloscope Calibration option)
Supported SignalsLeveled sine wave, fast edge, time mark, amplitude signals

Safety and Compliance

ParameterValue
Calibration TraceabilityTraceable to nationally recognized standards (certificate included)

Environmental

ParameterValue
Storage Temperature-20 to +70 °C
Non-Operating AltitudeUp to 12,200 m (40,000 ft)
Shock and VibrationMIL-T-28800E, Type III, Class 5

Performance

ParameterValue
Warm-up Time30 minutes (to meet full specifications)

Performance Specifications

ParameterValue
DC Voltage Output Range0 to ±1020 V
AC Voltage Output Range1 mV to 1020 V
DC Current Output Range0 to ±11 A
AC Current Output Range29 µA to 11 A
Resistance Output Range0 to 329.999 MΩ
Capacitance Output Range330 pF to 1.0999 mF
Phase Range0 to ±179.99°
Phase Resolution0.01°

Options & Upgrades

ParameterValue
Oscilloscope Calibration300 MHz (Option SC300) or 600 MHz (Option SC600)

Power

ParameterValue
Maximum Power Consumption600 VA

Compliance

ParameterValue
Safety StandardsIEC 1010-1 (1990), ANSI/ISA S82.01-1994, CAN/CSA-C22.2 No. 1010.1-92

Technology

Multi-Product Calibration

Applications

Calibration of handheld and bench multimeters, oscilloscopes, wattmeters, electronic thermometers, data loggers, chart recorders, power harmonics analyzers, process calibrators, and current clamps.