The Fluke 5520A Multi-Product Calibrator is a high-performance calibrator designed to address a wide-ranging workload, sourcing direct voltage and current, alternating voltage and current, resistance, capacitance, temperature, power, and phase. With available oscilloscope calibration and power quality options, it provides comprehensive calibration capability for up to 6.5-digit multimeters, power quality analyzers, process tools, and handheld instruments.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | GPIB (IEEE-488), RS-232 |
Physical
| Parameter | Value |
|---|---|
| Operating Temperature | 0 °C to 50 °C |
| Dimensions (W×H×D) | 432 mm × 178 mm × 473 mm |
DC Voltage Function
| Parameter | Value |
|---|---|
| DC Voltage Range Limits | 0 to ±1020 V |
| DC Voltage Best 1-Year Uncertainty | ±11 ppm of setting |
DC Current Function
| Parameter | Value |
|---|---|
| DC Current Range Limits | 0 to ±20.5 A |
AC Voltage Function
| Parameter | Value |
|---|---|
| AC Voltage Range Limits | 1 mV to 1020 V |
| AC Voltage Frequency Range | 10 Hz to 500 kHz |
| AC Voltage Waveforms | Sine, square, triangle, truncated sine |
AC Current Function
| Parameter | Value |
|---|---|
| AC Current Range Limits | 29 µA to 20.5 A |
| AC Current Frequency Range | 10 Hz to 30 kHz |
Resistance Function
| Parameter | Value |
|---|---|
| Resistance Range | 0 to 1100 MΩ |
| Resistance Output Mode | Continuously variable |
Capacitance Function
| Parameter | Value |
|---|---|
| Capacitance Range | 0.19 nF to 110 mF |
| Capacitance Output Mode | Continuously variable |
Power and Phase Function
| Parameter | Value |
|---|---|
| Phase Resolution | 0.01° |
| Dual Output Capabilities | Two simultaneous voltage outputs or voltage and current |
Frequency and Duty Cycle
| Parameter | Value |
|---|---|
| Frequency Output Range | 0.01 Hz to 2.0 MHz |
Oscilloscope Calibration Option
| Parameter | Value |
|---|---|
| Oscilloscope Calibration Bandwidth Options | Up to 300 MHz, 600 MHz, or 1.1 GHz (by option) |
Power Quality Option
| Parameter | Value |
|---|---|
| Power Quality Option Model | 5520A-PQ |
Pressure and Humidity Measurement
| Parameter | Value |
|---|---|
| Pressure Module Compatibility | Fluke 700 Series |
General & Environmental Specifications
| Parameter | Value |
|---|---|
| Calibration Software Compatibility | MET/CAL Plus |
| Warm-up Time | 30 minutes |
| Recommended Calibration Interval | 1 year (12 months) |
| Storage Temperature Range | -20 °C to 70 °C |
| Operating Altitude Limit | 3,050 m (10,000 ft) |
| Safety Compliance | Designed to comply with IEC 1010-1 |
Performance
| Parameter | Value |
|---|---|
| Frequency Output Accuracy | ± 2.5 ppm |
Power
| Parameter | Value |
|---|---|
| Power Consumption | 600 VA |
| Line Voltage Settings | 100 V / 120 V / 220 V / 240 V (± 10%) |
| Line Frequency Range | 47 Hz to 63 Hz |
Temperature
| Parameter | Value |
|---|---|
| Thermocouple Simulation Resolution | 0.01 °C / 0.01 °F |
| RTD Simulation Resolution | 0.01 °C / 0.01 °F |
Safety & Isolation
| Parameter | Value |
|---|---|
| Maximum Guard-to-Earth Voltage | 20 V peak |
Technology
Analog/Digital Signal Generation & Synthesis
Applications
Calibration and Metrology