- Home
- Categories
- Test equipment
- Meters
- Impedance
- PM 6306/02n
The Fluke PM 6306/02n is a high-precision benchtop RCL meter featuring a test frequency range of 50 Hz to 1 MHz and a basic measurement accuracy of 0.1%. Designed for comprehensive passive component testing, this model is equipped with a standard GPIB (IEEE-488) interface, allowing for seamless integration into automated test environments. It features 4-wire Kelvin front-panel test posts, continuously adjustable AC/DC test signals, and built-in contact checking to ensure optimal connection and measurement reliability.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | GPIB (IEEE-488) |
| Display | Backlit LCD with dual 5-digit parameter readouts and a 3-digit frequency/level readout |
| Warm-Up Time | 30 minutes |
| Recommended Calibration Interval | 1 year |
Physical
| Parameter | Value |
|---|---|
| Operating Temperature | 0 to 50 °C |
| Dimensions (W×H×D) | 315 mm × 105 mm × 405 mm |
Measurement Functions & Parameters
| Parameter | Value |
|---|---|
| Primary Measurement Parameters | L, C, R, Z |
| Secondary Measurement Parameters | D, Q, Phase Angle (θ), Rs, Rp, Cp, Cs, Lp, Ls |
| Equivalent Circuit Modes | Series, Parallel (Auto/Manual selection) |
| Automatic Parameter Selection | Supported |
| Test Signal Frequency Range | 50 Hz to 1 MHz |
| Test Frequency Resolution | 3-digit resolution (over 8,000 steps) |
| Test Frequency Accuracy | ±0.01% |
| AC Test Signal Voltage Range | 50 mV to 2 V rms |
| AC Test Signal Voltage Resolution | 10 mV |
| Output Impedance | 100 Ω |
| Internal DC Bias Voltage Range | 0 to 10 V (continuously adjustable in 10 mV steps) |
| External DC Bias Voltage Input Range | 0 to 40 V |
| External DC Bias Current Input Range | Up to 250 mA |
Measurement Range & Accuracy
| Parameter | Value |
|---|---|
| Basic Impedance Accuracy | 0.1% |
| Inductance (L) Measurement Range | 0.00 µH to 31.8 kH |
| Capacitance (C) Measurement Range | 0.00 pF to 318 mF |
| Resistance / Impedance (R/Z) Measurement Range | 0.0000 Ω to 200 MΩ |
| Dissipation Factor (D) Measurement Range | 0.0002 to 500 |
| Quality Factor (Q) Measurement Range | 0.002 to 500 |
| Phase Angle Range | -90.0° to +90.0° |
| Measurement Speed Modes | Fast, Medium, Slow (up to 10 measurements/second) |
| Averaging Function Range | 2 to 99 measurements |
| Zero Trim Function | Automatic open and short circuit calibration |
Display & User Interface
| Parameter | Value |
|---|---|
| Primary Parameter Display Range | 5 digits (up to 99999 counts) |
| Secondary Parameter Display Range | 5 digits (up to 99999 counts) |
| Deviation / Percent Deviation Mode | Supported (shows differences from a stored reference value) |
| Internal Setup Memory Capacity | 9 front panel setups |
| Actual Component Test Voltage/Current Readback | Supported (real-time display of V and I levels applied to the DUT) |
Interfaces & Connectivity
| Parameter | Value |
|---|---|
| Guard Terminal | Standard front panel chassis guard terminal |
Power Specifications
| Parameter | Value |
|---|---|
| Line Frequency Range | 50 Hz to 60 Hz |
| Power Consumption | 20 VA |
Physical Characteristics
| Parameter | Value |
|---|---|
| Form Factor | Benchtop with tilt stand |
Environmental & Safety Compliance
| Parameter | Value |
|---|---|
| Storage Temperature Range | -40 to +70 °C |
| Relative Humidity Operating Limit | Up to 90% RH (non-condensing) |
| Safety Standards Compliance | IEC 1010-1 / EN 61010-1 Class I |
| EMC / RFI Standards | EN 55011 / EN 50082-1 |
Performance
| Parameter | Value |
|---|---|
| Measurement Principle | 4-wire auto-balancing bridge |
| Fast Measurement Time | 250 ms (at test frequencies ≥ 1 kHz) |
| Medium Measurement Time | 450 ms (at test frequencies ≥ 1 kHz) |
| Slow Measurement Time | 900 ms (at test frequencies ≥ 1 kHz) |
Test Signal
| Parameter | Value |
|---|---|
| AC Test Current Range | 10 µA to 20 mA |
| AC Test Signal Voltage Accuracy | ±(5% + 10 mV) for 200 mV to 2 V, ±(10% + 5 mV) for 50 mV to 200 mV |
Interfaces
| Parameter | Value |
|---|---|
| GPIB Interface | IEEE-488.2 (standard on PM 6306/02 and PM 6306/02n) |
Power Supply
| Parameter | Value |
|---|---|
| Power Supply Voltage Settings | 100 V / 120 V / 220 V / 240 V AC (±10%) |
| Mains Fuse Rating | T 200 mA / 250 V (for 220 V / 240 V), T 400 mA / 250 V (for 100 V / 120 V) |
Environmental
| Parameter | Value |
|---|---|
| Operating Altitude Limit | Up to 2000 m |
DC Bias
| Parameter | Value |
|---|---|
| Internal DC Bias Voltage Resolution | 10 mV (0.00 V to 2.00 V) |
| Internal DC Bias Voltage Accuracy | ±(3% + 10 mV) |
| External DC Bias Voltage Maximum Limit | 40 V DC |
Measurement Control
| Parameter | Value |
|---|---|
| Trigger Modes | Continuous (Internal), Manual, External (via interface) |
Measurement Ranges
| Parameter | Value |
|---|---|
| DC Resistance (Rdc) Measurement Range | 0.0001 Ω to 20 MΩ |
Binning
| Parameter | Value |
|---|---|
| Binning Tolerance Limits | -99.9% to +99.9% |
Technology
Analog-to-digital converter with phase-coherent detection
Applications
Precision passive component impedance testing, quality control, and production line automation