PM 6306/02n

FlukeImpedance

The Fluke PM 6306/02n is a high-precision benchtop RCL meter featuring a test frequency range of 50 Hz to 1 MHz and a basic measurement accuracy of 0.1%. Designed for comprehensive passive component testing, this model is equipped with a standard GPIB (IEEE-488) interface, allowing for seamless integration into automated test environments. It features 4-wire Kelvin front-panel test posts, continuously adjustable AC/DC test signals, and built-in contact checking to ensure optimal connection and measurement reliability.

Specifications

General

ParameterValue
InterfacesGPIB (IEEE-488)
DisplayBacklit LCD with dual 5-digit parameter readouts and a 3-digit frequency/level readout
Warm-Up Time30 minutes
Recommended Calibration Interval1 year

Physical

ParameterValue
Operating Temperature0 to 50 °C
Dimensions (W×H×D)315 mm × 105 mm × 405 mm

Measurement Functions & Parameters

ParameterValue
Primary Measurement ParametersL, C, R, Z
Secondary Measurement ParametersD, Q, Phase Angle (θ), Rs, Rp, Cp, Cs, Lp, Ls
Equivalent Circuit ModesSeries, Parallel (Auto/Manual selection)
Automatic Parameter SelectionSupported
Test Signal Frequency Range50 Hz to 1 MHz
Test Frequency Resolution3-digit resolution (over 8,000 steps)
Test Frequency Accuracy±0.01%
AC Test Signal Voltage Range50 mV to 2 V rms
AC Test Signal Voltage Resolution10 mV
Output Impedance100 Ω
Internal DC Bias Voltage Range0 to 10 V (continuously adjustable in 10 mV steps)
External DC Bias Voltage Input Range0 to 40 V
External DC Bias Current Input RangeUp to 250 mA

Measurement Range & Accuracy

ParameterValue
Basic Impedance Accuracy0.1%
Inductance (L) Measurement Range0.00 µH to 31.8 kH
Capacitance (C) Measurement Range0.00 pF to 318 mF
Resistance / Impedance (R/Z) Measurement Range0.0000 Ω to 200 MΩ
Dissipation Factor (D) Measurement Range0.0002 to 500
Quality Factor (Q) Measurement Range0.002 to 500
Phase Angle Range-90.0° to +90.0°
Measurement Speed ModesFast, Medium, Slow (up to 10 measurements/second)
Averaging Function Range2 to 99 measurements
Zero Trim FunctionAutomatic open and short circuit calibration

Display & User Interface

ParameterValue
Primary Parameter Display Range5 digits (up to 99999 counts)
Secondary Parameter Display Range5 digits (up to 99999 counts)
Deviation / Percent Deviation ModeSupported (shows differences from a stored reference value)
Internal Setup Memory Capacity9 front panel setups
Actual Component Test Voltage/Current ReadbackSupported (real-time display of V and I levels applied to the DUT)

Interfaces & Connectivity

ParameterValue
Guard TerminalStandard front panel chassis guard terminal

Power Specifications

ParameterValue
Line Frequency Range50 Hz to 60 Hz
Power Consumption20 VA

Physical Characteristics

ParameterValue
Form FactorBenchtop with tilt stand

Environmental & Safety Compliance

ParameterValue
Storage Temperature Range-40 to +70 °C
Relative Humidity Operating LimitUp to 90% RH (non-condensing)
Safety Standards ComplianceIEC 1010-1 / EN 61010-1 Class I
EMC / RFI StandardsEN 55011 / EN 50082-1

Performance

ParameterValue
Measurement Principle4-wire auto-balancing bridge
Fast Measurement Time250 ms (at test frequencies ≥ 1 kHz)
Medium Measurement Time450 ms (at test frequencies ≥ 1 kHz)
Slow Measurement Time900 ms (at test frequencies ≥ 1 kHz)

Test Signal

ParameterValue
AC Test Current Range10 µA to 20 mA
AC Test Signal Voltage Accuracy±(5% + 10 mV) for 200 mV to 2 V, ±(10% + 5 mV) for 50 mV to 200 mV

Interfaces

ParameterValue
GPIB InterfaceIEEE-488.2 (standard on PM 6306/02 and PM 6306/02n)

Power Supply

ParameterValue
Power Supply Voltage Settings100 V / 120 V / 220 V / 240 V AC (±10%)
Mains Fuse RatingT 200 mA / 250 V (for 220 V / 240 V), T 400 mA / 250 V (for 100 V / 120 V)

Environmental

ParameterValue
Operating Altitude LimitUp to 2000 m

DC Bias

ParameterValue
Internal DC Bias Voltage Resolution10 mV (0.00 V to 2.00 V)
Internal DC Bias Voltage Accuracy±(3% + 10 mV)
External DC Bias Voltage Maximum Limit40 V DC

Measurement Control

ParameterValue
Trigger ModesContinuous (Internal), Manual, External (via interface)

Measurement Ranges

ParameterValue
DC Resistance (Rdc) Measurement Range0.0001 Ω to 20 MΩ

Binning

ParameterValue
Binning Tolerance Limits-99.9% to +99.9%

Technology

Analog-to-digital converter with phase-coherent detection

Applications